WE‐C‐304A‐06: Computed Radiography Image Exposure Revisited and Case Study

Z. Yang, J. Lobick, Paul G Nagy

Research output: Contribution to journalArticle

Abstract

Purpose: To provide case studies for computed radiographic (CR) system image exposure measurement metrics and drive manufacture innovation. Methods and Materials: Computed radiography image plate exposure indices were monitored over time. For entries in the exposure log file that indicated more than five times of the usual value, time stamps in the entries were matched manually to images stored on PACS to identify possible causes. Results: Large variation in the values of the exposure index was found. Both operator errors and software failures were found for images of high value of exposure index. After corrective actions, the FWHM of the distribution of the exposure index has decreased although the range of the value remains the same. Cases from Pediatric Intensive Care Unit are the most difficult to handle because of the patient size and medical complications. Conflict of Interest (only if applicable): JL is an employee of Agfa HealthCare Corporation.

Original languageEnglish (US)
Pages (from-to)2765-2766
Number of pages2
JournalMedical Physics
Volume36
Issue number6
DOIs
StatePublished - 2009
Externally publishedYes

Fingerprint

Radiography
Conflict of Interest
Pediatric Intensive Care Units
Software
Delivery of Health Care

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

WE‐C‐304A‐06 : Computed Radiography Image Exposure Revisited and Case Study. / Yang, Z.; Lobick, J.; Nagy, Paul G.

In: Medical Physics, Vol. 36, No. 6, 2009, p. 2765-2766.

Research output: Contribution to journalArticle

@article{7ceb8ba601264b0694b7e0845df59cff,
title = "WE‐C‐304A‐06: Computed Radiography Image Exposure Revisited and Case Study",
abstract = "Purpose: To provide case studies for computed radiographic (CR) system image exposure measurement metrics and drive manufacture innovation. Methods and Materials: Computed radiography image plate exposure indices were monitored over time. For entries in the exposure log file that indicated more than five times of the usual value, time stamps in the entries were matched manually to images stored on PACS to identify possible causes. Results: Large variation in the values of the exposure index was found. Both operator errors and software failures were found for images of high value of exposure index. After corrective actions, the FWHM of the distribution of the exposure index has decreased although the range of the value remains the same. Cases from Pediatric Intensive Care Unit are the most difficult to handle because of the patient size and medical complications. Conflict of Interest (only if applicable): JL is an employee of Agfa HealthCare Corporation.",
author = "Z. Yang and J. Lobick and Nagy, {Paul G}",
year = "2009",
doi = "10.1118/1.3182493",
language = "English (US)",
volume = "36",
pages = "2765--2766",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - WE‐C‐304A‐06

T2 - Computed Radiography Image Exposure Revisited and Case Study

AU - Yang, Z.

AU - Lobick, J.

AU - Nagy, Paul G

PY - 2009

Y1 - 2009

N2 - Purpose: To provide case studies for computed radiographic (CR) system image exposure measurement metrics and drive manufacture innovation. Methods and Materials: Computed radiography image plate exposure indices were monitored over time. For entries in the exposure log file that indicated more than five times of the usual value, time stamps in the entries were matched manually to images stored on PACS to identify possible causes. Results: Large variation in the values of the exposure index was found. Both operator errors and software failures were found for images of high value of exposure index. After corrective actions, the FWHM of the distribution of the exposure index has decreased although the range of the value remains the same. Cases from Pediatric Intensive Care Unit are the most difficult to handle because of the patient size and medical complications. Conflict of Interest (only if applicable): JL is an employee of Agfa HealthCare Corporation.

AB - Purpose: To provide case studies for computed radiographic (CR) system image exposure measurement metrics and drive manufacture innovation. Methods and Materials: Computed radiography image plate exposure indices were monitored over time. For entries in the exposure log file that indicated more than five times of the usual value, time stamps in the entries were matched manually to images stored on PACS to identify possible causes. Results: Large variation in the values of the exposure index was found. Both operator errors and software failures were found for images of high value of exposure index. After corrective actions, the FWHM of the distribution of the exposure index has decreased although the range of the value remains the same. Cases from Pediatric Intensive Care Unit are the most difficult to handle because of the patient size and medical complications. Conflict of Interest (only if applicable): JL is an employee of Agfa HealthCare Corporation.

UR - http://www.scopus.com/inward/record.url?scp=85024826687&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024826687&partnerID=8YFLogxK

U2 - 10.1118/1.3182493

DO - 10.1118/1.3182493

M3 - Article

AN - SCOPUS:85024826687

VL - 36

SP - 2765

EP - 2766

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -