Volume Displacement Of Scleral Buckles

John T. Thompson, Ronald G. Michels

Research output: Contribution to journalArticle

Abstract

Indentation of the eye wall by a scleral buckle displaces volume from the vitreous cavity. We developed a mathematical formula to calculate the volume displacement caused by a scleral buckle and verified the accuracy of this mathematical model by performing scleral buckles in 21 cadaver eyes and three eyes undergoing retinal reattachment surgery. A single 5-mm radial sponge of moderate height displaces about 0.2 mL of fluid and a circumferential 2.5-mm-wide band of moderate height displaces about 0.5 mL of fluid. Larger circumferential tires of 7to 10-mm width displace 1.1 to 1.8 mL of fluid, depending on the height and configuration of the scleral buckle. A 7- to 10-mm-wide circumferential scleral buckle with a buckle height of 4 mm may displace up to 45% of the volume of the vitreous cavity. This volume displacement should be considered when injecting expansile gases or pharmacologic agents into the vitreous cavity.

Original languageEnglish (US)
Pages (from-to)1822-1824
Number of pages3
JournalArchives of Ophthalmology
Volume103
Issue number12
DOIs
StatePublished - 1985

Fingerprint

Porifera
Cadaver
Theoretical Models
Gases

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Volume Displacement Of Scleral Buckles. / Thompson, John T.; Michels, Ronald G.

In: Archives of Ophthalmology, Vol. 103, No. 12, 1985, p. 1822-1824.

Research output: Contribution to journalArticle

Thompson, John T. ; Michels, Ronald G. / Volume Displacement Of Scleral Buckles. In: Archives of Ophthalmology. 1985 ; Vol. 103, No. 12. pp. 1822-1824.
@article{4dad2918f9ce46118d2948c0ebc2a6e2,
title = "Volume Displacement Of Scleral Buckles",
abstract = "Indentation of the eye wall by a scleral buckle displaces volume from the vitreous cavity. We developed a mathematical formula to calculate the volume displacement caused by a scleral buckle and verified the accuracy of this mathematical model by performing scleral buckles in 21 cadaver eyes and three eyes undergoing retinal reattachment surgery. A single 5-mm radial sponge of moderate height displaces about 0.2 mL of fluid and a circumferential 2.5-mm-wide band of moderate height displaces about 0.5 mL of fluid. Larger circumferential tires of 7to 10-mm width displace 1.1 to 1.8 mL of fluid, depending on the height and configuration of the scleral buckle. A 7- to 10-mm-wide circumferential scleral buckle with a buckle height of 4 mm may displace up to 45{\%} of the volume of the vitreous cavity. This volume displacement should be considered when injecting expansile gases or pharmacologic agents into the vitreous cavity.",
author = "Thompson, {John T.} and Michels, {Ronald G.}",
year = "1985",
doi = "10.1001/archopht.1985.01050120056019",
language = "English (US)",
volume = "103",
pages = "1822--1824",
journal = "JAMA Ophthalmology",
issn = "2168-6165",
publisher = "American Medical Association",
number = "12",

}

TY - JOUR

T1 - Volume Displacement Of Scleral Buckles

AU - Thompson, John T.

AU - Michels, Ronald G.

PY - 1985

Y1 - 1985

N2 - Indentation of the eye wall by a scleral buckle displaces volume from the vitreous cavity. We developed a mathematical formula to calculate the volume displacement caused by a scleral buckle and verified the accuracy of this mathematical model by performing scleral buckles in 21 cadaver eyes and three eyes undergoing retinal reattachment surgery. A single 5-mm radial sponge of moderate height displaces about 0.2 mL of fluid and a circumferential 2.5-mm-wide band of moderate height displaces about 0.5 mL of fluid. Larger circumferential tires of 7to 10-mm width displace 1.1 to 1.8 mL of fluid, depending on the height and configuration of the scleral buckle. A 7- to 10-mm-wide circumferential scleral buckle with a buckle height of 4 mm may displace up to 45% of the volume of the vitreous cavity. This volume displacement should be considered when injecting expansile gases or pharmacologic agents into the vitreous cavity.

AB - Indentation of the eye wall by a scleral buckle displaces volume from the vitreous cavity. We developed a mathematical formula to calculate the volume displacement caused by a scleral buckle and verified the accuracy of this mathematical model by performing scleral buckles in 21 cadaver eyes and three eyes undergoing retinal reattachment surgery. A single 5-mm radial sponge of moderate height displaces about 0.2 mL of fluid and a circumferential 2.5-mm-wide band of moderate height displaces about 0.5 mL of fluid. Larger circumferential tires of 7to 10-mm width displace 1.1 to 1.8 mL of fluid, depending on the height and configuration of the scleral buckle. A 7- to 10-mm-wide circumferential scleral buckle with a buckle height of 4 mm may displace up to 45% of the volume of the vitreous cavity. This volume displacement should be considered when injecting expansile gases or pharmacologic agents into the vitreous cavity.

UR - http://www.scopus.com/inward/record.url?scp=0022404822&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0022404822&partnerID=8YFLogxK

U2 - 10.1001/archopht.1985.01050120056019

DO - 10.1001/archopht.1985.01050120056019

M3 - Article

C2 - 4074171

AN - SCOPUS:0022404822

VL - 103

SP - 1822

EP - 1824

JO - JAMA Ophthalmology

JF - JAMA Ophthalmology

SN - 2168-6165

IS - 12

ER -