The need for better methods to determine release criteria for patients administered radioactive material

Richard B. Sparks, Jeffry A. Siegel, Richard L. Wahl

Research output: Contribution to journalArticle

Abstract

In current NRC regulations, three options exist that may be used to determine release criteria for patients administered radioactive materials. Absorbed dose estimates may be based on administered activity, measured dose rate, or on patient-specific calculations. All of these methods proposed by the NRC can lead to overestimation of the dose equivalent to others due to their oversimplified nature. The primary oversimplifications are the use of a point source methodology and using the measured surface entrance dose rate to determine whole body dose. In order to show the inaccuracy of these oversimplifications for 131I, results using Monte Carlo radiation transport analysis with simplified anthropomorphic mathematical phantoms were determined. These results were then compared to actual patient measurements and the results of point source analysis. The measurement data were taken from 49 131I radioimmunotherapy patients. The point source calculations were performed using well established methodologies and using the same assumptions as in the NRC regulations for patient release criteria. Monte Carlo results were obtained by implementing two simplified 70 kg anthropomorphic phantoms and performing radiation transport simulation. The activity in the 'patient' phantom was assumed to be localized in the abdominal region to correspond to the activity localization seen in the radioimmunotherapy patients who were measured. Dose equivalents per unit cumulated activities were determined for 131I using the various methods. The relationship between measured dose equivalent per unit cumulated activity and whole body dose equivalent per unit cumulated activity was also investigated using Monte Carlo analysis. The point source method as implemented by the NRC yields an estimated dose equivalent per unit cumulated activity of 1.6 x 10-8 mSv MBq-1 s-1 at 1 m (2.2 x 10-4 rem mCi-1 h-1 at 1 m), and the Monte Carlo based method yielded a whole body dose equivalent per unit cumulated activity in the target phantom of 6.8 x 10-9 mSv MBq-1 s-1(9.0 x 10-5 rem mCi-1 h-1) for abdominal localization of activity in the source phantom. The measurements of the radioimmunotherapy patients yielded an average result of 1.0 x 10-8 mSv MBq-1 s-1 (1.3 x 10-4 rem Ci-1 h-1). When corrected for the difference between measured surface dose equivalent and whole body dose equivalent as determined by Monte Carlo analysis, these measurements represent a whole body dose equivalent per unit cumulated activity of about 6.2 x 10-9 mSv MBq-1 s-1 (8.1 x 10- 5 rem mCi-1 h-1). Based on these results, the current NRC dose-based methodology for the release of patients administered radioactive materials significantly overestimates the dose equivalent to others from 131I therapy patients.

Original languageEnglish (US)
Pages (from-to)385-388
Number of pages4
JournalHealth Physics
Volume75
Issue number4
StatePublished - 1998
Externally publishedYes

Fingerprint

radioactive materials
Radioactive materials
dosage
Radioimmunotherapy
point source
Radiation
Monte Carlo analysis
point sources
methodology
radiation transport
need
equivalent dose
method
material
Monte Carlo Method
dose

Keywords

  • I
  • Medical radiation
  • Modeling, dose assessment
  • Regulations

ASJC Scopus subject areas

  • Environmental Science (miscellaneous)
  • Health, Toxicology and Mutagenesis
  • Toxicology
  • Physics and Astronomy (miscellaneous)
  • Public Health, Environmental and Occupational Health
  • Environmental Chemistry
  • Environmental Science(all)

Cite this

The need for better methods to determine release criteria for patients administered radioactive material. / Sparks, Richard B.; Siegel, Jeffry A.; Wahl, Richard L.

In: Health Physics, Vol. 75, No. 4, 1998, p. 385-388.

Research output: Contribution to journalArticle

Sparks, Richard B. ; Siegel, Jeffry A. ; Wahl, Richard L. / The need for better methods to determine release criteria for patients administered radioactive material. In: Health Physics. 1998 ; Vol. 75, No. 4. pp. 385-388.
@article{cb4dbe40070b4587979f45dcbc18080b,
title = "The need for better methods to determine release criteria for patients administered radioactive material",
abstract = "In current NRC regulations, three options exist that may be used to determine release criteria for patients administered radioactive materials. Absorbed dose estimates may be based on administered activity, measured dose rate, or on patient-specific calculations. All of these methods proposed by the NRC can lead to overestimation of the dose equivalent to others due to their oversimplified nature. The primary oversimplifications are the use of a point source methodology and using the measured surface entrance dose rate to determine whole body dose. In order to show the inaccuracy of these oversimplifications for 131I, results using Monte Carlo radiation transport analysis with simplified anthropomorphic mathematical phantoms were determined. These results were then compared to actual patient measurements and the results of point source analysis. The measurement data were taken from 49 131I radioimmunotherapy patients. The point source calculations were performed using well established methodologies and using the same assumptions as in the NRC regulations for patient release criteria. Monte Carlo results were obtained by implementing two simplified 70 kg anthropomorphic phantoms and performing radiation transport simulation. The activity in the 'patient' phantom was assumed to be localized in the abdominal region to correspond to the activity localization seen in the radioimmunotherapy patients who were measured. Dose equivalents per unit cumulated activities were determined for 131I using the various methods. The relationship between measured dose equivalent per unit cumulated activity and whole body dose equivalent per unit cumulated activity was also investigated using Monte Carlo analysis. The point source method as implemented by the NRC yields an estimated dose equivalent per unit cumulated activity of 1.6 x 10-8 mSv MBq-1 s-1 at 1 m (2.2 x 10-4 rem mCi-1 h-1 at 1 m), and the Monte Carlo based method yielded a whole body dose equivalent per unit cumulated activity in the target phantom of 6.8 x 10-9 mSv MBq-1 s-1(9.0 x 10-5 rem mCi-1 h-1) for abdominal localization of activity in the source phantom. The measurements of the radioimmunotherapy patients yielded an average result of 1.0 x 10-8 mSv MBq-1 s-1 (1.3 x 10-4 rem Ci-1 h-1). When corrected for the difference between measured surface dose equivalent and whole body dose equivalent as determined by Monte Carlo analysis, these measurements represent a whole body dose equivalent per unit cumulated activity of about 6.2 x 10-9 mSv MBq-1 s-1 (8.1 x 10- 5 rem mCi-1 h-1). Based on these results, the current NRC dose-based methodology for the release of patients administered radioactive materials significantly overestimates the dose equivalent to others from 131I therapy patients.",
keywords = "I, Medical radiation, Modeling, dose assessment, Regulations",
author = "Sparks, {Richard B.} and Siegel, {Jeffry A.} and Wahl, {Richard L.}",
year = "1998",
language = "English (US)",
volume = "75",
pages = "385--388",
journal = "Health Physics",
issn = "0017-9078",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - The need for better methods to determine release criteria for patients administered radioactive material

AU - Sparks, Richard B.

AU - Siegel, Jeffry A.

AU - Wahl, Richard L.

PY - 1998

Y1 - 1998

N2 - In current NRC regulations, three options exist that may be used to determine release criteria for patients administered radioactive materials. Absorbed dose estimates may be based on administered activity, measured dose rate, or on patient-specific calculations. All of these methods proposed by the NRC can lead to overestimation of the dose equivalent to others due to their oversimplified nature. The primary oversimplifications are the use of a point source methodology and using the measured surface entrance dose rate to determine whole body dose. In order to show the inaccuracy of these oversimplifications for 131I, results using Monte Carlo radiation transport analysis with simplified anthropomorphic mathematical phantoms were determined. These results were then compared to actual patient measurements and the results of point source analysis. The measurement data were taken from 49 131I radioimmunotherapy patients. The point source calculations were performed using well established methodologies and using the same assumptions as in the NRC regulations for patient release criteria. Monte Carlo results were obtained by implementing two simplified 70 kg anthropomorphic phantoms and performing radiation transport simulation. The activity in the 'patient' phantom was assumed to be localized in the abdominal region to correspond to the activity localization seen in the radioimmunotherapy patients who were measured. Dose equivalents per unit cumulated activities were determined for 131I using the various methods. The relationship between measured dose equivalent per unit cumulated activity and whole body dose equivalent per unit cumulated activity was also investigated using Monte Carlo analysis. The point source method as implemented by the NRC yields an estimated dose equivalent per unit cumulated activity of 1.6 x 10-8 mSv MBq-1 s-1 at 1 m (2.2 x 10-4 rem mCi-1 h-1 at 1 m), and the Monte Carlo based method yielded a whole body dose equivalent per unit cumulated activity in the target phantom of 6.8 x 10-9 mSv MBq-1 s-1(9.0 x 10-5 rem mCi-1 h-1) for abdominal localization of activity in the source phantom. The measurements of the radioimmunotherapy patients yielded an average result of 1.0 x 10-8 mSv MBq-1 s-1 (1.3 x 10-4 rem Ci-1 h-1). When corrected for the difference between measured surface dose equivalent and whole body dose equivalent as determined by Monte Carlo analysis, these measurements represent a whole body dose equivalent per unit cumulated activity of about 6.2 x 10-9 mSv MBq-1 s-1 (8.1 x 10- 5 rem mCi-1 h-1). Based on these results, the current NRC dose-based methodology for the release of patients administered radioactive materials significantly overestimates the dose equivalent to others from 131I therapy patients.

AB - In current NRC regulations, three options exist that may be used to determine release criteria for patients administered radioactive materials. Absorbed dose estimates may be based on administered activity, measured dose rate, or on patient-specific calculations. All of these methods proposed by the NRC can lead to overestimation of the dose equivalent to others due to their oversimplified nature. The primary oversimplifications are the use of a point source methodology and using the measured surface entrance dose rate to determine whole body dose. In order to show the inaccuracy of these oversimplifications for 131I, results using Monte Carlo radiation transport analysis with simplified anthropomorphic mathematical phantoms were determined. These results were then compared to actual patient measurements and the results of point source analysis. The measurement data were taken from 49 131I radioimmunotherapy patients. The point source calculations were performed using well established methodologies and using the same assumptions as in the NRC regulations for patient release criteria. Monte Carlo results were obtained by implementing two simplified 70 kg anthropomorphic phantoms and performing radiation transport simulation. The activity in the 'patient' phantom was assumed to be localized in the abdominal region to correspond to the activity localization seen in the radioimmunotherapy patients who were measured. Dose equivalents per unit cumulated activities were determined for 131I using the various methods. The relationship between measured dose equivalent per unit cumulated activity and whole body dose equivalent per unit cumulated activity was also investigated using Monte Carlo analysis. The point source method as implemented by the NRC yields an estimated dose equivalent per unit cumulated activity of 1.6 x 10-8 mSv MBq-1 s-1 at 1 m (2.2 x 10-4 rem mCi-1 h-1 at 1 m), and the Monte Carlo based method yielded a whole body dose equivalent per unit cumulated activity in the target phantom of 6.8 x 10-9 mSv MBq-1 s-1(9.0 x 10-5 rem mCi-1 h-1) for abdominal localization of activity in the source phantom. The measurements of the radioimmunotherapy patients yielded an average result of 1.0 x 10-8 mSv MBq-1 s-1 (1.3 x 10-4 rem Ci-1 h-1). When corrected for the difference between measured surface dose equivalent and whole body dose equivalent as determined by Monte Carlo analysis, these measurements represent a whole body dose equivalent per unit cumulated activity of about 6.2 x 10-9 mSv MBq-1 s-1 (8.1 x 10- 5 rem mCi-1 h-1). Based on these results, the current NRC dose-based methodology for the release of patients administered radioactive materials significantly overestimates the dose equivalent to others from 131I therapy patients.

KW - I

KW - Medical radiation

KW - Modeling, dose assessment

KW - Regulations

UR - http://www.scopus.com/inward/record.url?scp=0031707748&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031707748&partnerID=8YFLogxK

M3 - Article

C2 - 9753361

AN - SCOPUS:0031707748

VL - 75

SP - 385

EP - 388

JO - Health Physics

JF - Health Physics

SN - 0017-9078

IS - 4

ER -