The mitek mini G2 suture anchor: Biomechanical analysis of use in the hand

Barbara D. Buch, Peter Innis, Michael A. McClinton, Yoshihisa Kotani

Research output: Contribution to journalArticlepeer-review

33 Scopus citations

Abstract

We inserted 36 Mitek Mini G2 suture anchors (Mitek Surgical Products, Westwood, MA) into various sites of 14 fresh cadaver hands. A single-strand 0-Ethibond suture (Ethicon, Sommerville, NJ) was threaded through each anchor, inserted into bone, and loaded to failure. In all cases, either the anchor pulled out (11 of 36) or the suture broke (25 of 36). No anchors fractured. We also measured the breaking strength of six types of suture material (without anchors) commonly used by hand surgeons; all failed at much lower loads than those generated in our anchor pullout model. We conclude that the Mitek Mini G2 suture anchor possesses sufficient pullout strength for tendon and ligament fixation in the hand, including the distal phalanx.

Original languageEnglish (US)
Pages (from-to)877-881
Number of pages5
JournalJournal of Hand Surgery
Volume20
Issue number5
DOIs
StatePublished - Sep 1995
Externally publishedYes

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'The mitek mini G2 suture anchor: Biomechanical analysis of use in the hand'. Together they form a unique fingerprint.

Cite this