Abstract
We have tested the porosity and fatigue life of five commonly used bone cements: Simplex P, LVC, Zimmer regular, CMW and Palacos R. Tests were conducted with and without centrifugation and with the monomer at room temperature and, except for LVC, at 0°C. We found that the fatigue life of different specimens varied by a factor of nearly 100. It did not depend on porosity alone, but was more influenced by the basic composition of the cement. Simplex P when mixed with the monomer at 0°C and centrifuged for 60 seconds had the highest fatigue life and was still sufficiently liquid to use easily.
Original language | English (US) |
---|---|
Pages (from-to) | 39-42 |
Number of pages | 4 |
Journal | Journal of Bone and Joint Surgery - Series B |
Volume | 71 |
Issue number | 1 |
DOIs | |
State | Published - 1989 |
Externally published | Yes |
ASJC Scopus subject areas
- Surgery
- Orthopedics and Sports Medicine