The development of the transvenous automatic defibrillator

M. Mirowski, Morton M. Mower, William S. Staewen, Rollin H. Denniston, Albert I. Mendeloff

Research output: Contribution to journalReview articlepeer-review

Abstract

The transvenous automatic defibrillator is a device which automatically recognizes and treats ventricular fibrillation. The initial clinical prototype resembles an external pacemaker, the electronic components and power supply are external to the body, the sensing and defibrillating elements being contained in a transvenous catheter. The sensors monitor intracardiac R waves and cardiac contraction. Absence of both signals identifies the arrhythmia and triggers the defibrillatory shock, delivered through the right ventricular electrode and another more proximal on the catheter. If needed, the device recycles automatically. It is also adaptable for pacing and for elective intraatrial cardioversion. Results of animal experiments were successful, and tests are currently being done on patients requiring defibrillation during open heart surgery. The clinical indications for the device are numerous and it is hoped that this approach will decrease the present prohibitive mortality from coronary heart disease.

Original languageEnglish (US)
Pages (from-to)94-100
Number of pages7
JournalAnnals of Noninvasive Electrocardiology
Volume5
Issue number1
DOIs
StatePublished - 2000
Externally publishedYes

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Physiology (medical)

Fingerprint

Dive into the research topics of 'The development of the transvenous automatic defibrillator'. Together they form a unique fingerprint.

Cite this