The cause of the artifact in 4-slice helical computed tomography

Katsuyuki Taguchi, Hiroshi Aradate, Yasuo Saito, Ilan Zmora, Kyung S. Han, Michael D. Silver

Research output: Contribution to journalArticle

Abstract

The causes of the image artifacts in a 4-slice helical computed tomography have been discussed as follows: (1) changeover in pairs of data used in z interpolation, (2) sampling interval in z, and (3) the cone angle. This study analyzes the first two causes of the artifact and describes how the current algorithm [K. Taguchi and H. Aradate, Radiology 205P, 390 (1997); 205P, 618 (1997); Med. Phys. 25, 550-561 (1998); H. Hu, ibid. 26, 5-18 (1999); S. Schaller et al., IEEE Trans. Med. Imaging 19, 822-834 (2000); K. Taguchi, Ph.D. thesis, University of Tsukuba, 2002] solves the problem. An interpolated sinogram for a slice at the edge of a ball phantom shows discontinuity caused by the changeover. If we extend the streak artifact in the reconstructed image, it crosses the focus orbit at the corresponding projection angle. Applying z filtering can reduce such causes by its feathering effect and mixing data obtained by different cone angles; the best results are provided when z filtering is applied to densely sampled helical data.

Original languageEnglish (US)
Pages (from-to)2033-2037
Number of pages5
JournalMedical Physics
Volume31
Issue number7
DOIs
StatePublished - Jul 2004
Externally publishedYes

Fingerprint

Spiral Computed Tomography
Artifacts
Orbit
Radiology

Keywords

  • Computed tomography
  • Helical scan
  • Image artifacts
  • Multi-slice CT

ASJC Scopus subject areas

  • Biophysics

Cite this

Taguchi, K., Aradate, H., Saito, Y., Zmora, I., Han, K. S., & Silver, M. D. (2004). The cause of the artifact in 4-slice helical computed tomography. Medical Physics, 31(7), 2033-2037. https://doi.org/10.1118/1.1763005

The cause of the artifact in 4-slice helical computed tomography. / Taguchi, Katsuyuki; Aradate, Hiroshi; Saito, Yasuo; Zmora, Ilan; Han, Kyung S.; Silver, Michael D.

In: Medical Physics, Vol. 31, No. 7, 07.2004, p. 2033-2037.

Research output: Contribution to journalArticle

Taguchi, K, Aradate, H, Saito, Y, Zmora, I, Han, KS & Silver, MD 2004, 'The cause of the artifact in 4-slice helical computed tomography', Medical Physics, vol. 31, no. 7, pp. 2033-2037. https://doi.org/10.1118/1.1763005
Taguchi, Katsuyuki ; Aradate, Hiroshi ; Saito, Yasuo ; Zmora, Ilan ; Han, Kyung S. ; Silver, Michael D. / The cause of the artifact in 4-slice helical computed tomography. In: Medical Physics. 2004 ; Vol. 31, No. 7. pp. 2033-2037.
@article{52df7136d10242caadcc9074da44f960,
title = "The cause of the artifact in 4-slice helical computed tomography",
abstract = "The causes of the image artifacts in a 4-slice helical computed tomography have been discussed as follows: (1) changeover in pairs of data used in z interpolation, (2) sampling interval in z, and (3) the cone angle. This study analyzes the first two causes of the artifact and describes how the current algorithm [K. Taguchi and H. Aradate, Radiology 205P, 390 (1997); 205P, 618 (1997); Med. Phys. 25, 550-561 (1998); H. Hu, ibid. 26, 5-18 (1999); S. Schaller et al., IEEE Trans. Med. Imaging 19, 822-834 (2000); K. Taguchi, Ph.D. thesis, University of Tsukuba, 2002] solves the problem. An interpolated sinogram for a slice at the edge of a ball phantom shows discontinuity caused by the changeover. If we extend the streak artifact in the reconstructed image, it crosses the focus orbit at the corresponding projection angle. Applying z filtering can reduce such causes by its feathering effect and mixing data obtained by different cone angles; the best results are provided when z filtering is applied to densely sampled helical data.",
keywords = "Computed tomography, Helical scan, Image artifacts, Multi-slice CT",
author = "Katsuyuki Taguchi and Hiroshi Aradate and Yasuo Saito and Ilan Zmora and Han, {Kyung S.} and Silver, {Michael D.}",
year = "2004",
month = "7",
doi = "10.1118/1.1763005",
language = "English (US)",
volume = "31",
pages = "2033--2037",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "7",

}

TY - JOUR

T1 - The cause of the artifact in 4-slice helical computed tomography

AU - Taguchi, Katsuyuki

AU - Aradate, Hiroshi

AU - Saito, Yasuo

AU - Zmora, Ilan

AU - Han, Kyung S.

AU - Silver, Michael D.

PY - 2004/7

Y1 - 2004/7

N2 - The causes of the image artifacts in a 4-slice helical computed tomography have been discussed as follows: (1) changeover in pairs of data used in z interpolation, (2) sampling interval in z, and (3) the cone angle. This study analyzes the first two causes of the artifact and describes how the current algorithm [K. Taguchi and H. Aradate, Radiology 205P, 390 (1997); 205P, 618 (1997); Med. Phys. 25, 550-561 (1998); H. Hu, ibid. 26, 5-18 (1999); S. Schaller et al., IEEE Trans. Med. Imaging 19, 822-834 (2000); K. Taguchi, Ph.D. thesis, University of Tsukuba, 2002] solves the problem. An interpolated sinogram for a slice at the edge of a ball phantom shows discontinuity caused by the changeover. If we extend the streak artifact in the reconstructed image, it crosses the focus orbit at the corresponding projection angle. Applying z filtering can reduce such causes by its feathering effect and mixing data obtained by different cone angles; the best results are provided when z filtering is applied to densely sampled helical data.

AB - The causes of the image artifacts in a 4-slice helical computed tomography have been discussed as follows: (1) changeover in pairs of data used in z interpolation, (2) sampling interval in z, and (3) the cone angle. This study analyzes the first two causes of the artifact and describes how the current algorithm [K. Taguchi and H. Aradate, Radiology 205P, 390 (1997); 205P, 618 (1997); Med. Phys. 25, 550-561 (1998); H. Hu, ibid. 26, 5-18 (1999); S. Schaller et al., IEEE Trans. Med. Imaging 19, 822-834 (2000); K. Taguchi, Ph.D. thesis, University of Tsukuba, 2002] solves the problem. An interpolated sinogram for a slice at the edge of a ball phantom shows discontinuity caused by the changeover. If we extend the streak artifact in the reconstructed image, it crosses the focus orbit at the corresponding projection angle. Applying z filtering can reduce such causes by its feathering effect and mixing data obtained by different cone angles; the best results are provided when z filtering is applied to densely sampled helical data.

KW - Computed tomography

KW - Helical scan

KW - Image artifacts

KW - Multi-slice CT

UR - http://www.scopus.com/inward/record.url?scp=3142764517&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=3142764517&partnerID=8YFLogxK

U2 - 10.1118/1.1763005

DO - 10.1118/1.1763005

M3 - Article

C2 - 15305455

AN - SCOPUS:3142764517

VL - 31

SP - 2033

EP - 2037

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 7

ER -