SU‐GG‐I‐114: Detector Characterization of a New Flat‐Panel Imager: Prospects for Image Quality Improvement in Fluoroscopy and Cone‐Beam CT

N. Mail, D. Moseley, Jeff Siewerdsen, D. Jaffray

Research output: Contribution to journalArticle

Abstract

Purpose: To characterize the imaging performance of a new flat‐panel‐imager (FPI) with respect to metrics such as detector gain, linearity, lag, modulation transfer function (MTF), noise‐power spectrum (NPS), and detective quantum efficiency (DQE) and to evaluate its potential application in fluoroscopy and cone‐beam CT for radiation therapy guidance. Method and Materials: The detector examined was a Trixell (Pixium 4343RF) indirect‐ FPI with a 2881×2880 array of 0.143×0.143 mm2 pixels, 43×43 cm2 FOV and a 0.06 mm thick CsI:Tl x‐ray converter. The FPI was operated at a frame rate 3 fps. Gain, linearity and NPS were calculated using gain‐corrected flood images. MTF was measured using an edge‐spread function method. DQE was calculated from the measured MTF and NPS. Image lag was characterized as a function of incident exposure. NPS, MTF, DQE and lag were compared with a FPI design (Perkin Elmer RID1640) currently employed in image‐guided radiotherapy. Results: The dark current stabilizes after 30 minutes. The detector has high gain and linearity with R2∼1. The 50% MTF was achieved at 1.51 and 0.91 lp/mm at 120 kVp for Trixel and Perkin Elmer (PE) FPI, respectively. The spatial resolution was limited by the focal spot size. The NPS(f) is found lower than the PE at 120 kVp for the same pixel saturation. The DQE is calculated 55 and 34% at 1.25 lp/mm for Trixel and PE, respectively. The first frame lag is 7 times lower than Perkin Elmer for the same pixel saturation at 120 kVp. Radiographic images of a head phantom show high contrast and spatial resolution. Conclusion: Imaging performance metrics (in particular, the high linearity, low lag, and high DQE(f)) suggest a significant improvement for the Trixell FPI and strongly support potential application of this detector for fluoroscopy and cone‐beam CT. Research sponsored by Elekta.

Original languageEnglish (US)
Pages (from-to)2668
Number of pages1
JournalMedical Physics
Volume35
Issue number6
DOIs
StatePublished - 2008
Externally publishedYes

Fingerprint

Fluoroscopy
Quality Improvement
Radiotherapy
Head
X-Rays
Research

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐GG‐I‐114 : Detector Characterization of a New Flat‐Panel Imager: Prospects for Image Quality Improvement in Fluoroscopy and Cone‐Beam CT. / Mail, N.; Moseley, D.; Siewerdsen, Jeff; Jaffray, D.

In: Medical Physics, Vol. 35, No. 6, 2008, p. 2668.

Research output: Contribution to journalArticle

@article{9b46d7b680ad4d5a9c41d228d9c3c4dd,
title = "SU‐GG‐I‐114: Detector Characterization of a New Flat‐Panel Imager: Prospects for Image Quality Improvement in Fluoroscopy and Cone‐Beam CT",
abstract = "Purpose: To characterize the imaging performance of a new flat‐panel‐imager (FPI) with respect to metrics such as detector gain, linearity, lag, modulation transfer function (MTF), noise‐power spectrum (NPS), and detective quantum efficiency (DQE) and to evaluate its potential application in fluoroscopy and cone‐beam CT for radiation therapy guidance. Method and Materials: The detector examined was a Trixell (Pixium 4343RF) indirect‐ FPI with a 2881×2880 array of 0.143×0.143 mm2 pixels, 43×43 cm2 FOV and a 0.06 mm thick CsI:Tl x‐ray converter. The FPI was operated at a frame rate 3 fps. Gain, linearity and NPS were calculated using gain‐corrected flood images. MTF was measured using an edge‐spread function method. DQE was calculated from the measured MTF and NPS. Image lag was characterized as a function of incident exposure. NPS, MTF, DQE and lag were compared with a FPI design (Perkin Elmer RID1640) currently employed in image‐guided radiotherapy. Results: The dark current stabilizes after 30 minutes. The detector has high gain and linearity with R2∼1. The 50{\%} MTF was achieved at 1.51 and 0.91 lp/mm at 120 kVp for Trixel and Perkin Elmer (PE) FPI, respectively. The spatial resolution was limited by the focal spot size. The NPS(f) is found lower than the PE at 120 kVp for the same pixel saturation. The DQE is calculated 55 and 34{\%} at 1.25 lp/mm for Trixel and PE, respectively. The first frame lag is 7 times lower than Perkin Elmer for the same pixel saturation at 120 kVp. Radiographic images of a head phantom show high contrast and spatial resolution. Conclusion: Imaging performance metrics (in particular, the high linearity, low lag, and high DQE(f)) suggest a significant improvement for the Trixell FPI and strongly support potential application of this detector for fluoroscopy and cone‐beam CT. Research sponsored by Elekta.",
author = "N. Mail and D. Moseley and Jeff Siewerdsen and D. Jaffray",
year = "2008",
doi = "10.1118/1.2961512",
language = "English (US)",
volume = "35",
pages = "2668",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐GG‐I‐114

T2 - Detector Characterization of a New Flat‐Panel Imager: Prospects for Image Quality Improvement in Fluoroscopy and Cone‐Beam CT

AU - Mail, N.

AU - Moseley, D.

AU - Siewerdsen, Jeff

AU - Jaffray, D.

PY - 2008

Y1 - 2008

N2 - Purpose: To characterize the imaging performance of a new flat‐panel‐imager (FPI) with respect to metrics such as detector gain, linearity, lag, modulation transfer function (MTF), noise‐power spectrum (NPS), and detective quantum efficiency (DQE) and to evaluate its potential application in fluoroscopy and cone‐beam CT for radiation therapy guidance. Method and Materials: The detector examined was a Trixell (Pixium 4343RF) indirect‐ FPI with a 2881×2880 array of 0.143×0.143 mm2 pixels, 43×43 cm2 FOV and a 0.06 mm thick CsI:Tl x‐ray converter. The FPI was operated at a frame rate 3 fps. Gain, linearity and NPS were calculated using gain‐corrected flood images. MTF was measured using an edge‐spread function method. DQE was calculated from the measured MTF and NPS. Image lag was characterized as a function of incident exposure. NPS, MTF, DQE and lag were compared with a FPI design (Perkin Elmer RID1640) currently employed in image‐guided radiotherapy. Results: The dark current stabilizes after 30 minutes. The detector has high gain and linearity with R2∼1. The 50% MTF was achieved at 1.51 and 0.91 lp/mm at 120 kVp for Trixel and Perkin Elmer (PE) FPI, respectively. The spatial resolution was limited by the focal spot size. The NPS(f) is found lower than the PE at 120 kVp for the same pixel saturation. The DQE is calculated 55 and 34% at 1.25 lp/mm for Trixel and PE, respectively. The first frame lag is 7 times lower than Perkin Elmer for the same pixel saturation at 120 kVp. Radiographic images of a head phantom show high contrast and spatial resolution. Conclusion: Imaging performance metrics (in particular, the high linearity, low lag, and high DQE(f)) suggest a significant improvement for the Trixell FPI and strongly support potential application of this detector for fluoroscopy and cone‐beam CT. Research sponsored by Elekta.

AB - Purpose: To characterize the imaging performance of a new flat‐panel‐imager (FPI) with respect to metrics such as detector gain, linearity, lag, modulation transfer function (MTF), noise‐power spectrum (NPS), and detective quantum efficiency (DQE) and to evaluate its potential application in fluoroscopy and cone‐beam CT for radiation therapy guidance. Method and Materials: The detector examined was a Trixell (Pixium 4343RF) indirect‐ FPI with a 2881×2880 array of 0.143×0.143 mm2 pixels, 43×43 cm2 FOV and a 0.06 mm thick CsI:Tl x‐ray converter. The FPI was operated at a frame rate 3 fps. Gain, linearity and NPS were calculated using gain‐corrected flood images. MTF was measured using an edge‐spread function method. DQE was calculated from the measured MTF and NPS. Image lag was characterized as a function of incident exposure. NPS, MTF, DQE and lag were compared with a FPI design (Perkin Elmer RID1640) currently employed in image‐guided radiotherapy. Results: The dark current stabilizes after 30 minutes. The detector has high gain and linearity with R2∼1. The 50% MTF was achieved at 1.51 and 0.91 lp/mm at 120 kVp for Trixel and Perkin Elmer (PE) FPI, respectively. The spatial resolution was limited by the focal spot size. The NPS(f) is found lower than the PE at 120 kVp for the same pixel saturation. The DQE is calculated 55 and 34% at 1.25 lp/mm for Trixel and PE, respectively. The first frame lag is 7 times lower than Perkin Elmer for the same pixel saturation at 120 kVp. Radiographic images of a head phantom show high contrast and spatial resolution. Conclusion: Imaging performance metrics (in particular, the high linearity, low lag, and high DQE(f)) suggest a significant improvement for the Trixell FPI and strongly support potential application of this detector for fluoroscopy and cone‐beam CT. Research sponsored by Elekta.

UR - http://www.scopus.com/inward/record.url?scp=85024811487&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024811487&partnerID=8YFLogxK

U2 - 10.1118/1.2961512

DO - 10.1118/1.2961512

M3 - Article

AN - SCOPUS:85024811487

VL - 35

SP - 2668

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -