Structural stiffness of the hoffmann simple anterior tibial external fixation frame

J. Vossoughi, Y. Youm, M. Bosse, A. R. Burgess, A. Poka

Research output: Contribution to journalArticle

Abstract

Tibial external fixation frames were constructed on aluminum tube simulating tibia bone. A 20-mm gap was left at the fracture site in order to measure the structural stiffness of the frame rather than the aluminum tube. The performance of the frames were experimentally evaluated and quantified using tests which simulated the loading conditions encountered in normal walking. These included axial compression, anteroposterior (AP) bending, lateral bending and torsional loading of the frame. The parameters studied were (a) number of fixation pins, (b) number of connecting rods and (c) location of clamps on the pins. Four constants were evaluated from these tests using various structural configurations of the frames; these resulted in four stiffness coefficients in compression, AP bending, lateral bending and torsion. Stiffnesses of various frames with different geometric configurations were compared by comparing their appropriate stiffness coefficients. Such comparison can set forth a quantitative guideline in selecting a suitable frame configuration for the type of injury and condition of fracture pattern. This type of quantitative analysis can also be useful in modifying the frame during the postoperative bone healing process.

Original languageEnglish (US)
Pages (from-to)127-141
Number of pages15
JournalAnnals of Biomedical Engineering
Volume17
Issue number2
DOIs
StatePublished - Mar 1989
Externally publishedYes

Fingerprint

Stiffness
Bending (deformation)
Bone
Aluminum
Connecting rods
Axial compression
Clamping devices
Torsional stress
Compaction
Chemical analysis

ASJC Scopus subject areas

  • Biomedical Engineering

Cite this

Structural stiffness of the hoffmann simple anterior tibial external fixation frame. / Vossoughi, J.; Youm, Y.; Bosse, M.; Burgess, A. R.; Poka, A.

In: Annals of Biomedical Engineering, Vol. 17, No. 2, 03.1989, p. 127-141.

Research output: Contribution to journalArticle

Vossoughi, J. ; Youm, Y. ; Bosse, M. ; Burgess, A. R. ; Poka, A. / Structural stiffness of the hoffmann simple anterior tibial external fixation frame. In: Annals of Biomedical Engineering. 1989 ; Vol. 17, No. 2. pp. 127-141.
@article{2dd741085f3f48ce885a3e54d195e7ac,
title = "Structural stiffness of the hoffmann simple anterior tibial external fixation frame",
abstract = "Tibial external fixation frames were constructed on aluminum tube simulating tibia bone. A 20-mm gap was left at the fracture site in order to measure the structural stiffness of the frame rather than the aluminum tube. The performance of the frames were experimentally evaluated and quantified using tests which simulated the loading conditions encountered in normal walking. These included axial compression, anteroposterior (AP) bending, lateral bending and torsional loading of the frame. The parameters studied were (a) number of fixation pins, (b) number of connecting rods and (c) location of clamps on the pins. Four constants were evaluated from these tests using various structural configurations of the frames; these resulted in four stiffness coefficients in compression, AP bending, lateral bending and torsion. Stiffnesses of various frames with different geometric configurations were compared by comparing their appropriate stiffness coefficients. Such comparison can set forth a quantitative guideline in selecting a suitable frame configuration for the type of injury and condition of fracture pattern. This type of quantitative analysis can also be useful in modifying the frame during the postoperative bone healing process.",
author = "J. Vossoughi and Y. Youm and M. Bosse and Burgess, {A. R.} and A. Poka",
year = "1989",
month = "3",
doi = "10.1007/BF02368023",
language = "English (US)",
volume = "17",
pages = "127--141",
journal = "Annals of Biomedical Engineering",
issn = "0090-6964",
publisher = "Springer Netherlands",
number = "2",

}

TY - JOUR

T1 - Structural stiffness of the hoffmann simple anterior tibial external fixation frame

AU - Vossoughi, J.

AU - Youm, Y.

AU - Bosse, M.

AU - Burgess, A. R.

AU - Poka, A.

PY - 1989/3

Y1 - 1989/3

N2 - Tibial external fixation frames were constructed on aluminum tube simulating tibia bone. A 20-mm gap was left at the fracture site in order to measure the structural stiffness of the frame rather than the aluminum tube. The performance of the frames were experimentally evaluated and quantified using tests which simulated the loading conditions encountered in normal walking. These included axial compression, anteroposterior (AP) bending, lateral bending and torsional loading of the frame. The parameters studied were (a) number of fixation pins, (b) number of connecting rods and (c) location of clamps on the pins. Four constants were evaluated from these tests using various structural configurations of the frames; these resulted in four stiffness coefficients in compression, AP bending, lateral bending and torsion. Stiffnesses of various frames with different geometric configurations were compared by comparing their appropriate stiffness coefficients. Such comparison can set forth a quantitative guideline in selecting a suitable frame configuration for the type of injury and condition of fracture pattern. This type of quantitative analysis can also be useful in modifying the frame during the postoperative bone healing process.

AB - Tibial external fixation frames were constructed on aluminum tube simulating tibia bone. A 20-mm gap was left at the fracture site in order to measure the structural stiffness of the frame rather than the aluminum tube. The performance of the frames were experimentally evaluated and quantified using tests which simulated the loading conditions encountered in normal walking. These included axial compression, anteroposterior (AP) bending, lateral bending and torsional loading of the frame. The parameters studied were (a) number of fixation pins, (b) number of connecting rods and (c) location of clamps on the pins. Four constants were evaluated from these tests using various structural configurations of the frames; these resulted in four stiffness coefficients in compression, AP bending, lateral bending and torsion. Stiffnesses of various frames with different geometric configurations were compared by comparing their appropriate stiffness coefficients. Such comparison can set forth a quantitative guideline in selecting a suitable frame configuration for the type of injury and condition of fracture pattern. This type of quantitative analysis can also be useful in modifying the frame during the postoperative bone healing process.

UR - http://www.scopus.com/inward/record.url?scp=0024600677&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0024600677&partnerID=8YFLogxK

U2 - 10.1007/BF02368023

DO - 10.1007/BF02368023

M3 - Article

VL - 17

SP - 127

EP - 141

JO - Annals of Biomedical Engineering

JF - Annals of Biomedical Engineering

SN - 0090-6964

IS - 2

ER -