Structural failure of a Starr-Edward aortic track valve

R. E. Ringel, A. L. Moulton, J. E. Burns

Research output: Contribution to journalArticle

Abstract

Structural failure of a Model 2400 Starr-Edwards aortic track valve occurred suddenly, 4 years after implantation. At operation, the valve cage was removed from the descending aorta. Examination of the excised prosthesis disclosed minimal cloth wear and no evidence of infective growth; however, three struts were fractures above their insertion into the valve ring. To our knowledge this type of valve malfunction has not been previously noted.

Original languageEnglish (US)
Pages (from-to)81-83
Number of pages3
JournalTexas Heart Institute Journal
Volume10
Issue number1
StatePublished - Jan 1 1983

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint Dive into the research topics of 'Structural failure of a Starr-Edward aortic track valve'. Together they form a unique fingerprint.

  • Cite this