Stability of reamed and unreamed intramedullary tibial nails: a biomechanical study

A. C. Fairbank, D. Thomas, B. Cunningham, M. Curtis, R. H. Jinnah

Research output: Contribution to journalArticlepeer-review

45 Scopus citations

Abstract

We examined the correlation of bone mineral density and bone strength in the cadaver tibia, and looked in vitro at the relative stability of tibial fractures fixed with either reamed or unreamed libial nails. Bonemineral density correlated well with bone strength (r = 0.946), but paired tibias did not correlate closely. The unreamed nail-bone construct was less stable than the reamed construct in each pair tested (P<0.05), and a comparison of all bones showed it to be less stable at all levels of our testing regimen, including failure (P<0.01). Bending and breakage was seen in four of the smaller unreamed interlocking screws (4 mm).

Original languageEnglish (US)
Pages (from-to)483-485
Number of pages3
JournalInjury
Volume26
Issue number7
DOIs
StatePublished - Sep 1995
Externally publishedYes

ASJC Scopus subject areas

  • Emergency Medicine
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Stability of reamed and unreamed intramedullary tibial nails: a biomechanical study'. Together they form a unique fingerprint.

Cite this