Simulation study of the effects of near- and far-field heating during focused ultrasound uterine fibroid ablation using an electronically focused phased array: A theoretical analysis of patient safety

Nicholas Ellens, Kullervo Hynynen

Research output: Contribution to journalArticle

Abstract

Purpose: Assess the feasibility of using large-aperture, flat ultrasonic transducer arrays with 6500 small elements operating at 500 kHz without the use of any mechanical components for the thermal coagulation of uterine fibroids. This study examines the benefits and detriments of using a frequency that is significantly lower than that used in clinical systems (1-1.5 MHz). Methods: Ultrasound simulations were performed using the anatomies of five fibroid patients derived from 3D MRI. Using electronic steering solely, the ultrasound focus from a flat, 6500-element phased array was translated around the volume of the fibroids in various patterns to assess the feasibility of completing full treatments from fixed physical locations. Successive temperature maps were generated by numerically solving the bioheat equation. Using a thermal dose model, the bioeffects of these simulations were quantified and analyzed. Results: The simulations indicate that such an array could be used to perform fibroid treatments to 18-EM43 at an average rate of 90 ± 20 cm3/h without physically moving the transducer array. On average, the maximum near-field thermal dose for each patient was below 4-EM43. Fibroid tissue could be treated as close as 40-mm to the spine without reaching temperatures expected to cause pain or damage. Conclusions: Fibroids were successfully targeted and treated from a single transducer position to acceptable extents and without causing damage in the near- or far-field. Compared to clinical systems, treatment rates were good. The proposed treatment paradigm is a promising alternative to existing systems and warrants further investigation.

Original languageEnglish (US)
Article number072902
JournalMedical Physics
Volume41
Issue number7
DOIs
StatePublished - 2014
Externally publishedYes

Fingerprint

Leiomyoma
Patient Safety
Heating
Transducers
Hot Temperature
Temperature
Therapeutics
Ultrasonics
Anatomy
Spine
Pain

Keywords

  • focused ultrasound
  • HIFU
  • near-field heating
  • phased arrays
  • thermal coagulation
  • uterine fibroids

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging
  • Medicine(all)

Cite this

@article{734ba50cb8f24f05895f731e04556477,
title = "Simulation study of the effects of near- and far-field heating during focused ultrasound uterine fibroid ablation using an electronically focused phased array: A theoretical analysis of patient safety",
abstract = "Purpose: Assess the feasibility of using large-aperture, flat ultrasonic transducer arrays with 6500 small elements operating at 500 kHz without the use of any mechanical components for the thermal coagulation of uterine fibroids. This study examines the benefits and detriments of using a frequency that is significantly lower than that used in clinical systems (1-1.5 MHz). Methods: Ultrasound simulations were performed using the anatomies of five fibroid patients derived from 3D MRI. Using electronic steering solely, the ultrasound focus from a flat, 6500-element phased array was translated around the volume of the fibroids in various patterns to assess the feasibility of completing full treatments from fixed physical locations. Successive temperature maps were generated by numerically solving the bioheat equation. Using a thermal dose model, the bioeffects of these simulations were quantified and analyzed. Results: The simulations indicate that such an array could be used to perform fibroid treatments to 18-EM43 at an average rate of 90 ± 20 cm3/h without physically moving the transducer array. On average, the maximum near-field thermal dose for each patient was below 4-EM43. Fibroid tissue could be treated as close as 40-mm to the spine without reaching temperatures expected to cause pain or damage. Conclusions: Fibroids were successfully targeted and treated from a single transducer position to acceptable extents and without causing damage in the near- or far-field. Compared to clinical systems, treatment rates were good. The proposed treatment paradigm is a promising alternative to existing systems and warrants further investigation.",
keywords = "focused ultrasound, HIFU, near-field heating, phased arrays, thermal coagulation, uterine fibroids",
author = "Nicholas Ellens and Kullervo Hynynen",
year = "2014",
doi = "10.1118/1.4883777",
language = "English (US)",
volume = "41",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "7",

}

TY - JOUR

T1 - Simulation study of the effects of near- and far-field heating during focused ultrasound uterine fibroid ablation using an electronically focused phased array

T2 - A theoretical analysis of patient safety

AU - Ellens, Nicholas

AU - Hynynen, Kullervo

PY - 2014

Y1 - 2014

N2 - Purpose: Assess the feasibility of using large-aperture, flat ultrasonic transducer arrays with 6500 small elements operating at 500 kHz without the use of any mechanical components for the thermal coagulation of uterine fibroids. This study examines the benefits and detriments of using a frequency that is significantly lower than that used in clinical systems (1-1.5 MHz). Methods: Ultrasound simulations were performed using the anatomies of five fibroid patients derived from 3D MRI. Using electronic steering solely, the ultrasound focus from a flat, 6500-element phased array was translated around the volume of the fibroids in various patterns to assess the feasibility of completing full treatments from fixed physical locations. Successive temperature maps were generated by numerically solving the bioheat equation. Using a thermal dose model, the bioeffects of these simulations were quantified and analyzed. Results: The simulations indicate that such an array could be used to perform fibroid treatments to 18-EM43 at an average rate of 90 ± 20 cm3/h without physically moving the transducer array. On average, the maximum near-field thermal dose for each patient was below 4-EM43. Fibroid tissue could be treated as close as 40-mm to the spine without reaching temperatures expected to cause pain or damage. Conclusions: Fibroids were successfully targeted and treated from a single transducer position to acceptable extents and without causing damage in the near- or far-field. Compared to clinical systems, treatment rates were good. The proposed treatment paradigm is a promising alternative to existing systems and warrants further investigation.

AB - Purpose: Assess the feasibility of using large-aperture, flat ultrasonic transducer arrays with 6500 small elements operating at 500 kHz without the use of any mechanical components for the thermal coagulation of uterine fibroids. This study examines the benefits and detriments of using a frequency that is significantly lower than that used in clinical systems (1-1.5 MHz). Methods: Ultrasound simulations were performed using the anatomies of five fibroid patients derived from 3D MRI. Using electronic steering solely, the ultrasound focus from a flat, 6500-element phased array was translated around the volume of the fibroids in various patterns to assess the feasibility of completing full treatments from fixed physical locations. Successive temperature maps were generated by numerically solving the bioheat equation. Using a thermal dose model, the bioeffects of these simulations were quantified and analyzed. Results: The simulations indicate that such an array could be used to perform fibroid treatments to 18-EM43 at an average rate of 90 ± 20 cm3/h without physically moving the transducer array. On average, the maximum near-field thermal dose for each patient was below 4-EM43. Fibroid tissue could be treated as close as 40-mm to the spine without reaching temperatures expected to cause pain or damage. Conclusions: Fibroids were successfully targeted and treated from a single transducer position to acceptable extents and without causing damage in the near- or far-field. Compared to clinical systems, treatment rates were good. The proposed treatment paradigm is a promising alternative to existing systems and warrants further investigation.

KW - focused ultrasound

KW - HIFU

KW - near-field heating

KW - phased arrays

KW - thermal coagulation

KW - uterine fibroids

UR - http://www.scopus.com/inward/record.url?scp=84903122626&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84903122626&partnerID=8YFLogxK

U2 - 10.1118/1.4883777

DO - 10.1118/1.4883777

M3 - Article

C2 - 24989412

AN - SCOPUS:84903122626

VL - 41

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 7

M1 - 072902

ER -