Research validity of measurements when reusing icare probes

Michelle S. Lee, Brad Barnett, Jing Tian, Scott McCabe, Eric L Singman

Research output: Contribution to journalArticle

Abstract

Background: It has been reported that reusing Icare tonometer probes may not pose a considerable risk of transmission of infection, thereby supporting this practice under extraordinary circumstances, such as mass glaucoma screenings in developing countries. The present study sought to determine whether reusing the probes reduced the validity of pressure readings. Methods: Thousands of measurements were made with the same Icare probe on a purpose-built calibration device. A second set of measurements was made during which the Icare probe head was wiped with an alcohol pad between each reading. Results: Statistically different readings were observed between the no-wipe and wipe groups for most of the settings evaluated. In addition, statistically significant readings were noted for a few of the settings as the number of readings increased for either group. However, these differences would not be considered clinically significant. Conclusion: This study suggests that reusing Icare tonometer probes, even for hundreds of measurements, would not reduce the clinical utility of the pressure readings. For mass screenings in developing countries where cost savings is crucial, considering the reuse of Icare tonometer probes seems reasonable.

Original languageEnglish (US)
Pages (from-to)468-473
Number of pages6
JournalBiomedical Instrumentation and Technology
Volume51
Issue number6
StatePublished - Nov 1 2017

Fingerprint

Developing countries
Screening
Alcohols
Calibration
Costs

ASJC Scopus subject areas

  • Biomedical Engineering
  • Computer Networks and Communications

Cite this

Research validity of measurements when reusing icare probes. / Lee, Michelle S.; Barnett, Brad; Tian, Jing; McCabe, Scott; Singman, Eric L.

In: Biomedical Instrumentation and Technology, Vol. 51, No. 6, 01.11.2017, p. 468-473.

Research output: Contribution to journalArticle

Lee, Michelle S. ; Barnett, Brad ; Tian, Jing ; McCabe, Scott ; Singman, Eric L. / Research validity of measurements when reusing icare probes. In: Biomedical Instrumentation and Technology. 2017 ; Vol. 51, No. 6. pp. 468-473.
@article{9c6a340842e747d39cbbf4c3ce5bf744,
title = "Research validity of measurements when reusing icare probes",
abstract = "Background: It has been reported that reusing Icare tonometer probes may not pose a considerable risk of transmission of infection, thereby supporting this practice under extraordinary circumstances, such as mass glaucoma screenings in developing countries. The present study sought to determine whether reusing the probes reduced the validity of pressure readings. Methods: Thousands of measurements were made with the same Icare probe on a purpose-built calibration device. A second set of measurements was made during which the Icare probe head was wiped with an alcohol pad between each reading. Results: Statistically different readings were observed between the no-wipe and wipe groups for most of the settings evaluated. In addition, statistically significant readings were noted for a few of the settings as the number of readings increased for either group. However, these differences would not be considered clinically significant. Conclusion: This study suggests that reusing Icare tonometer probes, even for hundreds of measurements, would not reduce the clinical utility of the pressure readings. For mass screenings in developing countries where cost savings is crucial, considering the reuse of Icare tonometer probes seems reasonable.",
author = "Lee, {Michelle S.} and Brad Barnett and Jing Tian and Scott McCabe and Singman, {Eric L}",
year = "2017",
month = "11",
day = "1",
language = "English (US)",
volume = "51",
pages = "468--473",
journal = "Biomedical Instrumentation and Technology",
issn = "0899-8205",
publisher = "Association for the Advancement of Medical Instrumentation",
number = "6",

}

TY - JOUR

T1 - Research validity of measurements when reusing icare probes

AU - Lee, Michelle S.

AU - Barnett, Brad

AU - Tian, Jing

AU - McCabe, Scott

AU - Singman, Eric L

PY - 2017/11/1

Y1 - 2017/11/1

N2 - Background: It has been reported that reusing Icare tonometer probes may not pose a considerable risk of transmission of infection, thereby supporting this practice under extraordinary circumstances, such as mass glaucoma screenings in developing countries. The present study sought to determine whether reusing the probes reduced the validity of pressure readings. Methods: Thousands of measurements were made with the same Icare probe on a purpose-built calibration device. A second set of measurements was made during which the Icare probe head was wiped with an alcohol pad between each reading. Results: Statistically different readings were observed between the no-wipe and wipe groups for most of the settings evaluated. In addition, statistically significant readings were noted for a few of the settings as the number of readings increased for either group. However, these differences would not be considered clinically significant. Conclusion: This study suggests that reusing Icare tonometer probes, even for hundreds of measurements, would not reduce the clinical utility of the pressure readings. For mass screenings in developing countries where cost savings is crucial, considering the reuse of Icare tonometer probes seems reasonable.

AB - Background: It has been reported that reusing Icare tonometer probes may not pose a considerable risk of transmission of infection, thereby supporting this practice under extraordinary circumstances, such as mass glaucoma screenings in developing countries. The present study sought to determine whether reusing the probes reduced the validity of pressure readings. Methods: Thousands of measurements were made with the same Icare probe on a purpose-built calibration device. A second set of measurements was made during which the Icare probe head was wiped with an alcohol pad between each reading. Results: Statistically different readings were observed between the no-wipe and wipe groups for most of the settings evaluated. In addition, statistically significant readings were noted for a few of the settings as the number of readings increased for either group. However, these differences would not be considered clinically significant. Conclusion: This study suggests that reusing Icare tonometer probes, even for hundreds of measurements, would not reduce the clinical utility of the pressure readings. For mass screenings in developing countries where cost savings is crucial, considering the reuse of Icare tonometer probes seems reasonable.

UR - http://www.scopus.com/inward/record.url?scp=85035036890&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85035036890&partnerID=8YFLogxK

M3 - Article

C2 - 29161129

AN - SCOPUS:85035036890

VL - 51

SP - 468

EP - 473

JO - Biomedical Instrumentation and Technology

JF - Biomedical Instrumentation and Technology

SN - 0899-8205

IS - 6

ER -