Precision of 2 Low-dose Abdomen/Pelvis Cone Beam Computed Tomography Protocols for Alignment to Bone and Soft Tissue in Pediatric Patients Receiving Image Guided Radiation Therapy

Avani D. Rao, Junghoon Lee, Wei Fu, Sarah Nicholas, Sara R. Alcorn, Joseph Moore, Matthew Ladra, Mahadevappa Mahesh, Steven Bartolac, Stephanie A. Terezakis

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Purpose: To evaluate the precision of 2 low-dose cone (LD) beam computed tomography (CBCT) protocols to align to bone and soft tissue for pediatric patients receiving image guided radiation therapy (IGRT) to the abdomen and pelvis. Methods and Materials: Image-quality evaluation was done for 858 CBCT scans from 46 pediatric patients treated with IGRT from January 2015 to December 2017. The evaluations guided the development of 2 significantly dose-reduced protocols, LD-CBCT1 and a further dose-reduced LD-CBCT2. Representative scans from LD-CBCT1 and LD-CBCT2 from 8 patients with at least 1 CBCT scan from both protocols were registered separately to a bone and soft-tissue landmark on the simulation computed tomography scan. Eighteen identical blinded random offsets were applied to each patient's LD-CBCT1 and LD-CBCT2 from a starting registration that was then realigned using rigid registration. The residual offset between the baseline registration and the final registration attempt was calculated and analyzed using a 1-sided, 1 sample t test to evaluate whether LD-CBCT1, delivering a higher dose, was superior to the lower-dose LD-CBCT2 for bone and soft-tissue alignment. Results: In comparing 288 registrations with a bone landmark across 8 patients, no difference was found in the vector magnitude offsets using LD-CBCT 1 (mean [x¯], 0.73 mm; standard deviation [σ], 0.39 mm) and LD-CBCT2 (x¯, 0.74 mm; σ, 0.40 mm; P =.425). Comparing 216 registrations with a soft-tissue landmark across 6 patients, alignment using LD-CBCT2 (x¯, 1.55 mm; σ, 1.08 mm) resulted in larger differences in the vector magnitude of the offsets compared with LD-CBCT1 (x¯, 1.37 mm; σ, 0.74 mm; P =.049). Conclusions: Clinics treating pediatric patients should consider implementing a protocol mirroring LD-CBCT2 for abdomen and pelvis IGRT bone alignment. Further evaluation of the precision of LD-CBCTs for soft-tissue alignment is necessary.

Original languageEnglish (US)
Pages (from-to)e307-e313
JournalPractical Radiation Oncology
Volume9
Issue number3
DOIs
StatePublished - May 2019

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Precision of 2 Low-dose Abdomen/Pelvis Cone Beam Computed Tomography Protocols for Alignment to Bone and Soft Tissue in Pediatric Patients Receiving Image Guided Radiation Therapy'. Together they form a unique fingerprint.

Cite this