Plug surface defects as a late complication of silicone punctal plugs

Stella C. Paparizos, Deepak P. Edward, Smajo Osmanovic

Research output: Contribution to journalArticle

Abstract

Purpose: To report 3 cases of punctal plug surface defects as a cause of irritation complicating punctal plug therapy. Methods: Interventional case series and scanning electron microscopy (SEM) of punctal plugs. Results: Three patients were found to have punctal plug surface defects and local irritation of conjunctiva and inferonasal cornea 9, 40, and 69 months after placement of AquaFlo silicone punctal plugs for the treatment of dry eyes. SEM of the explanted plugs confirmed a defective collarette surface. In addition, unused AquaFlo plugs from different lots examined by SEM showed a rough irregular collarette surface that was not seen in unused plugs of Parasol silicone punctal plugs. DISCUSSION:: AquaFlo punctal plugs inherently have surface irregularities when compared with similar plug brands, which we suggest may predispose to plug breakdown over time, leading to mechanical irritation of the adjacent conjunctiva and cornea. This report highlights the local complications by AquaFlo punctal plugs associated with plug surface defects as a complication of punctal plug therapy.

Original languageEnglish (US)
Pages (from-to)1224-1226
Number of pages3
JournalCornea
Volume32
Issue number9
DOIs
StatePublished - Sep 2013

Fingerprint

Electron Scanning Microscopy
Conjunctiva
Cornea
Punctal Plugs
Therapeutics

Keywords

  • complications
  • punctal plug
  • scanning electron microscopy
  • surface defects

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Plug surface defects as a late complication of silicone punctal plugs. / Paparizos, Stella C.; Edward, Deepak P.; Osmanovic, Smajo.

In: Cornea, Vol. 32, No. 9, 09.2013, p. 1224-1226.

Research output: Contribution to journalArticle

Paparizos, SC, Edward, DP & Osmanovic, S 2013, 'Plug surface defects as a late complication of silicone punctal plugs', Cornea, vol. 32, no. 9, pp. 1224-1226. https://doi.org/10.1097/ICO.0b013e31829a6881
Paparizos, Stella C. ; Edward, Deepak P. ; Osmanovic, Smajo. / Plug surface defects as a late complication of silicone punctal plugs. In: Cornea. 2013 ; Vol. 32, No. 9. pp. 1224-1226.
@article{405c80b7b9ea458eb486d3e6b7713b0a,
title = "Plug surface defects as a late complication of silicone punctal plugs",
abstract = "Purpose: To report 3 cases of punctal plug surface defects as a cause of irritation complicating punctal plug therapy. Methods: Interventional case series and scanning electron microscopy (SEM) of punctal plugs. Results: Three patients were found to have punctal plug surface defects and local irritation of conjunctiva and inferonasal cornea 9, 40, and 69 months after placement of AquaFlo silicone punctal plugs for the treatment of dry eyes. SEM of the explanted plugs confirmed a defective collarette surface. In addition, unused AquaFlo plugs from different lots examined by SEM showed a rough irregular collarette surface that was not seen in unused plugs of Parasol silicone punctal plugs. DISCUSSION:: AquaFlo punctal plugs inherently have surface irregularities when compared with similar plug brands, which we suggest may predispose to plug breakdown over time, leading to mechanical irritation of the adjacent conjunctiva and cornea. This report highlights the local complications by AquaFlo punctal plugs associated with plug surface defects as a complication of punctal plug therapy.",
keywords = "complications, punctal plug, scanning electron microscopy, surface defects",
author = "Paparizos, {Stella C.} and Edward, {Deepak P.} and Smajo Osmanovic",
year = "2013",
month = "9",
doi = "10.1097/ICO.0b013e31829a6881",
language = "English (US)",
volume = "32",
pages = "1224--1226",
journal = "Cornea",
issn = "0277-3740",
publisher = "Lippincott Williams and Wilkins",
number = "9",

}

TY - JOUR

T1 - Plug surface defects as a late complication of silicone punctal plugs

AU - Paparizos, Stella C.

AU - Edward, Deepak P.

AU - Osmanovic, Smajo

PY - 2013/9

Y1 - 2013/9

N2 - Purpose: To report 3 cases of punctal plug surface defects as a cause of irritation complicating punctal plug therapy. Methods: Interventional case series and scanning electron microscopy (SEM) of punctal plugs. Results: Three patients were found to have punctal plug surface defects and local irritation of conjunctiva and inferonasal cornea 9, 40, and 69 months after placement of AquaFlo silicone punctal plugs for the treatment of dry eyes. SEM of the explanted plugs confirmed a defective collarette surface. In addition, unused AquaFlo plugs from different lots examined by SEM showed a rough irregular collarette surface that was not seen in unused plugs of Parasol silicone punctal plugs. DISCUSSION:: AquaFlo punctal plugs inherently have surface irregularities when compared with similar plug brands, which we suggest may predispose to plug breakdown over time, leading to mechanical irritation of the adjacent conjunctiva and cornea. This report highlights the local complications by AquaFlo punctal plugs associated with plug surface defects as a complication of punctal plug therapy.

AB - Purpose: To report 3 cases of punctal plug surface defects as a cause of irritation complicating punctal plug therapy. Methods: Interventional case series and scanning electron microscopy (SEM) of punctal plugs. Results: Three patients were found to have punctal plug surface defects and local irritation of conjunctiva and inferonasal cornea 9, 40, and 69 months after placement of AquaFlo silicone punctal plugs for the treatment of dry eyes. SEM of the explanted plugs confirmed a defective collarette surface. In addition, unused AquaFlo plugs from different lots examined by SEM showed a rough irregular collarette surface that was not seen in unused plugs of Parasol silicone punctal plugs. DISCUSSION:: AquaFlo punctal plugs inherently have surface irregularities when compared with similar plug brands, which we suggest may predispose to plug breakdown over time, leading to mechanical irritation of the adjacent conjunctiva and cornea. This report highlights the local complications by AquaFlo punctal plugs associated with plug surface defects as a complication of punctal plug therapy.

KW - complications

KW - punctal plug

KW - scanning electron microscopy

KW - surface defects

UR - http://www.scopus.com/inward/record.url?scp=84883285751&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84883285751&partnerID=8YFLogxK

U2 - 10.1097/ICO.0b013e31829a6881

DO - 10.1097/ICO.0b013e31829a6881

M3 - Article

C2 - 23860429

AN - SCOPUS:84883285751

VL - 32

SP - 1224

EP - 1226

JO - Cornea

JF - Cornea

SN - 0277-3740

IS - 9

ER -