TY - JOUR
T1 - Performance of point-of-care xpert HIV-1 plasma viral load assay at a tertiary HIV care centre in Southern India
AU - Swathirajan, Chinnambedu Ravichandran
AU - Vignesh, Ramachandran
AU - Boobalan, Jayaseelan
AU - Solomon, Sunil Suhas
AU - Saravanan, Shanmugam
AU - Balakrishnan, Pachamuthu
N1 - Publisher Copyright:
© 2017 The Authors.
Copyright:
Copyright 2017 Elsevier B.V., All rights reserved.
PY - 2017/10
Y1 - 2017/10
N2 - Background. Sustainable suppression of HIV replication forms the basis of anti-retroviral therapy (ART) medication. Thus, reliable quantification of HIV viral load has become an essential factor to monitor the effectiveness of the ART. Longer turnaround-time (TAT), batch testing and technical skills are major drawbacks of standard real-time PCR assays. Methods. The performance of the point-of-care Xpert HIV-1 viral load assay was evaluated against the Abbott RealTime PCR m2000rt system. A total of 96 plasma specimens ranging from 2.5 log10 copies ml-1 to 4.99 log10 copies ml-1 and proficiency testing panel specimens were used. Precision and accuracy were checked using the Pearson correlation coefficient test and Bland-Altman analysis. Results. Compared to the Abbott RealTime PCR, the Xpert HIV-1 viral load assay showed a good correlation (Pearson r=0.81; P<0.0001) with a mean difference of 0.27 log10 copies ml-1 (95%CI, -0.41 to 0.96 log10 copies ml-1; SD, 0.35 log10 copies ml-1). Conclusion. Reliable and ease of testing individual specimens could make the Xpert HIV-1 viral load assay an efficient alternative method for ART monitoring in clinical management of HIV disease in resource-limited settings. The rapid test results (less than 2 h) could help in making an immediate clinical decision, which further strengthens patient care.
AB - Background. Sustainable suppression of HIV replication forms the basis of anti-retroviral therapy (ART) medication. Thus, reliable quantification of HIV viral load has become an essential factor to monitor the effectiveness of the ART. Longer turnaround-time (TAT), batch testing and technical skills are major drawbacks of standard real-time PCR assays. Methods. The performance of the point-of-care Xpert HIV-1 viral load assay was evaluated against the Abbott RealTime PCR m2000rt system. A total of 96 plasma specimens ranging from 2.5 log10 copies ml-1 to 4.99 log10 copies ml-1 and proficiency testing panel specimens were used. Precision and accuracy were checked using the Pearson correlation coefficient test and Bland-Altman analysis. Results. Compared to the Abbott RealTime PCR, the Xpert HIV-1 viral load assay showed a good correlation (Pearson r=0.81; P<0.0001) with a mean difference of 0.27 log10 copies ml-1 (95%CI, -0.41 to 0.96 log10 copies ml-1; SD, 0.35 log10 copies ml-1). Conclusion. Reliable and ease of testing individual specimens could make the Xpert HIV-1 viral load assay an efficient alternative method for ART monitoring in clinical management of HIV disease in resource-limited settings. The rapid test results (less than 2 h) could help in making an immediate clinical decision, which further strengthens patient care.
KW - HIV
KW - Plasma viral load
KW - Point-of-care
KW - Xpert
UR - http://www.scopus.com/inward/record.url?scp=85031416664&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85031416664&partnerID=8YFLogxK
U2 - 10.1099/jmm.0.000514
DO - 10.1099/jmm.0.000514
M3 - Article
C2 - 28901908
AN - SCOPUS:85031416664
SN - 0022-2615
VL - 66
SP - 1379
EP - 1382
JO - Journal of Medical Microbiology
JF - Journal of Medical Microbiology
IS - 10
M1 - 000514
ER -