Pathology of balloon-expandable and self-expanding stents following MRI-guided transapical aortic valve replacement

Bogdan A. Kindzelski, Ming Li, Dumitru Mazilu, Keith A. Horvath

Research output: Contribution to journalArticle

Abstract

BACKGROUND AND AIM OF THE STUDY: Balloon-expandable (BE) and self-expanding (SE) prostheses are both used for transcatheter aortic valve replacement (TAVR), but differences in long-term outcome using these types of device are unknown. The study aim was to monitor the histopathology, echocardiographic findings and structural integrity of BE and SE stents in a preclinical model for up to six months after TAVR.

METHODS: Real-time magnetic resonance imaging (rtMRI)-guided TAVR was performed in 22 Yucatan pigs using either a BE (n = 10) or a SE (n = 12) prosthesis. Follow up echocardiography and MRI studies were performed at one-, three-, and six-month intervals. Additionally, high-contrast radiography was used to assess for strut fractures. The pigs were sacrificed after six months and tissues taken for histopathologic analysis.

RESULTS: Stent malapposition was found in seven BE prostheses (70%), and in three SE prostheses (25%) (p = 0.046). Three of the SE group (25%) had a partial left coronary artery obstruction. The incidence and severity of aortic regurgitation were similar between the BE and SE groups. Three BE prostheses (30%) and one SE prosthesis (8.3%) had a gap between the stent frame and aorta. The mean (±SD) number of strut fractures was 6.1 ± 3.45 and 1.17 ± 2.32 in the BE and SE groups, respectively (p = 0.002). In addition, two implanted BE prostheses (20%) had a consequential compressed stent frame appearance.

CONCLUSION: Long-term pathologic examination of necropsy specimens from a preclinical model of rtMRI-guided TAVR showed SE stent prostheses to be superior to their BE counterparts in terms of correct valve apposition and durability. These results may be attributed to the differing deployment methods and associated expansion forces employed by the BE and SE stents.

Original languageEnglish (US)
Pages (from-to)139-147
Number of pages9
JournalJournal of Heart Valve Disease
Volume24
Issue number2
StatePublished - Mar 1 2015
Externally publishedYes

Fingerprint

Aortic Valve
Prostheses and Implants
Stents
Pathology
Swine
Magnetic Resonance Imaging
Aortic Valve Insufficiency
Radiography
Echocardiography
Aorta
Coronary Vessels
Equipment and Supplies
Transcatheter Aortic Valve Replacement
Incidence

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Pathology of balloon-expandable and self-expanding stents following MRI-guided transapical aortic valve replacement. / Kindzelski, Bogdan A.; Li, Ming; Mazilu, Dumitru; Horvath, Keith A.

In: Journal of Heart Valve Disease, Vol. 24, No. 2, 01.03.2015, p. 139-147.

Research output: Contribution to journalArticle

@article{5c204ab852e8480687b583e7eab579ca,
title = "Pathology of balloon-expandable and self-expanding stents following MRI-guided transapical aortic valve replacement",
abstract = "BACKGROUND AND AIM OF THE STUDY: Balloon-expandable (BE) and self-expanding (SE) prostheses are both used for transcatheter aortic valve replacement (TAVR), but differences in long-term outcome using these types of device are unknown. The study aim was to monitor the histopathology, echocardiographic findings and structural integrity of BE and SE stents in a preclinical model for up to six months after TAVR.METHODS: Real-time magnetic resonance imaging (rtMRI)-guided TAVR was performed in 22 Yucatan pigs using either a BE (n = 10) or a SE (n = 12) prosthesis. Follow up echocardiography and MRI studies were performed at one-, three-, and six-month intervals. Additionally, high-contrast radiography was used to assess for strut fractures. The pigs were sacrificed after six months and tissues taken for histopathologic analysis.RESULTS: Stent malapposition was found in seven BE prostheses (70{\%}), and in three SE prostheses (25{\%}) (p = 0.046). Three of the SE group (25{\%}) had a partial left coronary artery obstruction. The incidence and severity of aortic regurgitation were similar between the BE and SE groups. Three BE prostheses (30{\%}) and one SE prosthesis (8.3{\%}) had a gap between the stent frame and aorta. The mean (±SD) number of strut fractures was 6.1 ± 3.45 and 1.17 ± 2.32 in the BE and SE groups, respectively (p = 0.002). In addition, two implanted BE prostheses (20{\%}) had a consequential compressed stent frame appearance.CONCLUSION: Long-term pathologic examination of necropsy specimens from a preclinical model of rtMRI-guided TAVR showed SE stent prostheses to be superior to their BE counterparts in terms of correct valve apposition and durability. These results may be attributed to the differing deployment methods and associated expansion forces employed by the BE and SE stents.",
author = "Kindzelski, {Bogdan A.} and Ming Li and Dumitru Mazilu and Horvath, {Keith A.}",
year = "2015",
month = "3",
day = "1",
language = "English (US)",
volume = "24",
pages = "139--147",
journal = "Journal of Heart Valve Disease",
issn = "0966-8519",
publisher = "ICR Publishers Ltd",
number = "2",

}

TY - JOUR

T1 - Pathology of balloon-expandable and self-expanding stents following MRI-guided transapical aortic valve replacement

AU - Kindzelski, Bogdan A.

AU - Li, Ming

AU - Mazilu, Dumitru

AU - Horvath, Keith A.

PY - 2015/3/1

Y1 - 2015/3/1

N2 - BACKGROUND AND AIM OF THE STUDY: Balloon-expandable (BE) and self-expanding (SE) prostheses are both used for transcatheter aortic valve replacement (TAVR), but differences in long-term outcome using these types of device are unknown. The study aim was to monitor the histopathology, echocardiographic findings and structural integrity of BE and SE stents in a preclinical model for up to six months after TAVR.METHODS: Real-time magnetic resonance imaging (rtMRI)-guided TAVR was performed in 22 Yucatan pigs using either a BE (n = 10) or a SE (n = 12) prosthesis. Follow up echocardiography and MRI studies were performed at one-, three-, and six-month intervals. Additionally, high-contrast radiography was used to assess for strut fractures. The pigs were sacrificed after six months and tissues taken for histopathologic analysis.RESULTS: Stent malapposition was found in seven BE prostheses (70%), and in three SE prostheses (25%) (p = 0.046). Three of the SE group (25%) had a partial left coronary artery obstruction. The incidence and severity of aortic regurgitation were similar between the BE and SE groups. Three BE prostheses (30%) and one SE prosthesis (8.3%) had a gap between the stent frame and aorta. The mean (±SD) number of strut fractures was 6.1 ± 3.45 and 1.17 ± 2.32 in the BE and SE groups, respectively (p = 0.002). In addition, two implanted BE prostheses (20%) had a consequential compressed stent frame appearance.CONCLUSION: Long-term pathologic examination of necropsy specimens from a preclinical model of rtMRI-guided TAVR showed SE stent prostheses to be superior to their BE counterparts in terms of correct valve apposition and durability. These results may be attributed to the differing deployment methods and associated expansion forces employed by the BE and SE stents.

AB - BACKGROUND AND AIM OF THE STUDY: Balloon-expandable (BE) and self-expanding (SE) prostheses are both used for transcatheter aortic valve replacement (TAVR), but differences in long-term outcome using these types of device are unknown. The study aim was to monitor the histopathology, echocardiographic findings and structural integrity of BE and SE stents in a preclinical model for up to six months after TAVR.METHODS: Real-time magnetic resonance imaging (rtMRI)-guided TAVR was performed in 22 Yucatan pigs using either a BE (n = 10) or a SE (n = 12) prosthesis. Follow up echocardiography and MRI studies were performed at one-, three-, and six-month intervals. Additionally, high-contrast radiography was used to assess for strut fractures. The pigs were sacrificed after six months and tissues taken for histopathologic analysis.RESULTS: Stent malapposition was found in seven BE prostheses (70%), and in three SE prostheses (25%) (p = 0.046). Three of the SE group (25%) had a partial left coronary artery obstruction. The incidence and severity of aortic regurgitation were similar between the BE and SE groups. Three BE prostheses (30%) and one SE prosthesis (8.3%) had a gap between the stent frame and aorta. The mean (±SD) number of strut fractures was 6.1 ± 3.45 and 1.17 ± 2.32 in the BE and SE groups, respectively (p = 0.002). In addition, two implanted BE prostheses (20%) had a consequential compressed stent frame appearance.CONCLUSION: Long-term pathologic examination of necropsy specimens from a preclinical model of rtMRI-guided TAVR showed SE stent prostheses to be superior to their BE counterparts in terms of correct valve apposition and durability. These results may be attributed to the differing deployment methods and associated expansion forces employed by the BE and SE stents.

UR - http://www.scopus.com/inward/record.url?scp=84964698261&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84964698261&partnerID=8YFLogxK

M3 - Article

C2 - 26204676

AN - SCOPUS:84964698261

VL - 24

SP - 139

EP - 147

JO - Journal of Heart Valve Disease

JF - Journal of Heart Valve Disease

SN - 0966-8519

IS - 2

ER -