Optimization of pulse train presaturation for CEST imaging in clinical scanners

Benjamin Schmitt, Moritz Zaiß, Jinyuan Zhou, Peter Bachert

Research output: Contribution to journalArticle

Abstract

Chemical exchange saturation transfer (CEST) imaging depends on the performance of radiofrequency saturation during the experiment. Scanner specifications, in particular limited pulse width and duty-cycle, and specific absorption rate guidelines restrict the full exploitation of CEST effects in clinical MR systems. The purpose of this study was to optimize techniques for effective pulse train presaturation for CEST imaging in a whole-body MR scanner. Theoretical analysis and simulations of the spectral properties of radiofrequency pulse trains demonstrated the significance of pulse width τP and interpulse delay τD for effective and selective labeling of a chemically exchanging proton pool. CEST experiments with model solutions, e.g., creatine dissolved in water, showed best performance of pulse trains with τP = τD = 100 msec, regarding minimum direct water saturation in z-spectra and distinct magnetization transfer ratio asymmetry that can be determined quantitatively. Saturation efficiency of trains of Gaussian-shaped radiofrequency pulses using this timing was evaluated in MR imagers with field strengths of 1.5, 3, and 7 T. The proposed saturation pulse train does not require hardware modifications, offers low specific absorption rate, and can be used in a standard clinical setup.

Original languageEnglish (US)
Pages (from-to)1620-1629
Number of pages10
JournalMagnetic Resonance in Medicine
Volume65
Issue number6
DOIs
StatePublished - Jun 2011

Fingerprint

Creatine
Water
Protons
Guidelines

Keywords

  • CEST imaging
  • clinical MR scanner
  • pulse train
  • selective spin labeling

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Optimization of pulse train presaturation for CEST imaging in clinical scanners. / Schmitt, Benjamin; Zaiß, Moritz; Zhou, Jinyuan; Bachert, Peter.

In: Magnetic Resonance in Medicine, Vol. 65, No. 6, 06.2011, p. 1620-1629.

Research output: Contribution to journalArticle

Schmitt, Benjamin ; Zaiß, Moritz ; Zhou, Jinyuan ; Bachert, Peter. / Optimization of pulse train presaturation for CEST imaging in clinical scanners. In: Magnetic Resonance in Medicine. 2011 ; Vol. 65, No. 6. pp. 1620-1629.
@article{d395d29ab701456490a943575b5ac17c,
title = "Optimization of pulse train presaturation for CEST imaging in clinical scanners",
abstract = "Chemical exchange saturation transfer (CEST) imaging depends on the performance of radiofrequency saturation during the experiment. Scanner specifications, in particular limited pulse width and duty-cycle, and specific absorption rate guidelines restrict the full exploitation of CEST effects in clinical MR systems. The purpose of this study was to optimize techniques for effective pulse train presaturation for CEST imaging in a whole-body MR scanner. Theoretical analysis and simulations of the spectral properties of radiofrequency pulse trains demonstrated the significance of pulse width τP and interpulse delay τD for effective and selective labeling of a chemically exchanging proton pool. CEST experiments with model solutions, e.g., creatine dissolved in water, showed best performance of pulse trains with τP = τD = 100 msec, regarding minimum direct water saturation in z-spectra and distinct magnetization transfer ratio asymmetry that can be determined quantitatively. Saturation efficiency of trains of Gaussian-shaped radiofrequency pulses using this timing was evaluated in MR imagers with field strengths of 1.5, 3, and 7 T. The proposed saturation pulse train does not require hardware modifications, offers low specific absorption rate, and can be used in a standard clinical setup.",
keywords = "CEST imaging, clinical MR scanner, pulse train, selective spin labeling",
author = "Benjamin Schmitt and Moritz Zai{\ss} and Jinyuan Zhou and Peter Bachert",
year = "2011",
month = "6",
doi = "10.1002/mrm.22750",
language = "English (US)",
volume = "65",
pages = "1620--1629",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "6",

}

TY - JOUR

T1 - Optimization of pulse train presaturation for CEST imaging in clinical scanners

AU - Schmitt, Benjamin

AU - Zaiß, Moritz

AU - Zhou, Jinyuan

AU - Bachert, Peter

PY - 2011/6

Y1 - 2011/6

N2 - Chemical exchange saturation transfer (CEST) imaging depends on the performance of radiofrequency saturation during the experiment. Scanner specifications, in particular limited pulse width and duty-cycle, and specific absorption rate guidelines restrict the full exploitation of CEST effects in clinical MR systems. The purpose of this study was to optimize techniques for effective pulse train presaturation for CEST imaging in a whole-body MR scanner. Theoretical analysis and simulations of the spectral properties of radiofrequency pulse trains demonstrated the significance of pulse width τP and interpulse delay τD for effective and selective labeling of a chemically exchanging proton pool. CEST experiments with model solutions, e.g., creatine dissolved in water, showed best performance of pulse trains with τP = τD = 100 msec, regarding minimum direct water saturation in z-spectra and distinct magnetization transfer ratio asymmetry that can be determined quantitatively. Saturation efficiency of trains of Gaussian-shaped radiofrequency pulses using this timing was evaluated in MR imagers with field strengths of 1.5, 3, and 7 T. The proposed saturation pulse train does not require hardware modifications, offers low specific absorption rate, and can be used in a standard clinical setup.

AB - Chemical exchange saturation transfer (CEST) imaging depends on the performance of radiofrequency saturation during the experiment. Scanner specifications, in particular limited pulse width and duty-cycle, and specific absorption rate guidelines restrict the full exploitation of CEST effects in clinical MR systems. The purpose of this study was to optimize techniques for effective pulse train presaturation for CEST imaging in a whole-body MR scanner. Theoretical analysis and simulations of the spectral properties of radiofrequency pulse trains demonstrated the significance of pulse width τP and interpulse delay τD for effective and selective labeling of a chemically exchanging proton pool. CEST experiments with model solutions, e.g., creatine dissolved in water, showed best performance of pulse trains with τP = τD = 100 msec, regarding minimum direct water saturation in z-spectra and distinct magnetization transfer ratio asymmetry that can be determined quantitatively. Saturation efficiency of trains of Gaussian-shaped radiofrequency pulses using this timing was evaluated in MR imagers with field strengths of 1.5, 3, and 7 T. The proposed saturation pulse train does not require hardware modifications, offers low specific absorption rate, and can be used in a standard clinical setup.

KW - CEST imaging

KW - clinical MR scanner

KW - pulse train

KW - selective spin labeling

UR - http://www.scopus.com/inward/record.url?scp=79958252245&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79958252245&partnerID=8YFLogxK

U2 - 10.1002/mrm.22750

DO - 10.1002/mrm.22750

M3 - Article

C2 - 21337418

AN - SCOPUS:79958252245

VL - 65

SP - 1620

EP - 1629

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 6

ER -