Nummular dystrophic calcification of an Ahmed glaucoma valve in a child

Arif O. Khan, Hind Al-Katan, Deepak P. Edward

Research output: Contribution to journalArticle

Abstract

We report the case of an otherwise-healthy 4.5-year-old girl with primary congenital glaucoma who had multiple nummular calcium deposits on the dorsal plate of a previously-implanted Ahmed glaucoma valve. The child had received the implant in the right eye at 6 months of age and developed increased intraocular pressure requiring surgical excision of surrounding fibrous encapsulation. Intraoperative inspection revealed that the dorsal plate surface was covered with multiple white nummular lesions; staining and X-ray microanalysis confirmed the lesions to be calcium deposits. Calcium deposits have been documented for various ophthalmic implants, including silicone scleral buckles and intraocular lenses, but to the best of our knowledge they have not previously been described in association with a glaucoma drainage device.

Fingerprint

Glaucoma
Calcium
Electron Probe Microanalysis
Intraocular Lenses
Silicones
Intraocular Pressure
Drainage
Staining and Labeling
Equipment and Supplies

ASJC Scopus subject areas

  • Ophthalmology
  • Pediatrics, Perinatology, and Child Health

Cite this

@article{49cb3b4d34d145e7b8946fe507452df4,
title = "Nummular dystrophic calcification of an Ahmed glaucoma valve in a child",
abstract = "We report the case of an otherwise-healthy 4.5-year-old girl with primary congenital glaucoma who had multiple nummular calcium deposits on the dorsal plate of a previously-implanted Ahmed glaucoma valve. The child had received the implant in the right eye at 6 months of age and developed increased intraocular pressure requiring surgical excision of surrounding fibrous encapsulation. Intraoperative inspection revealed that the dorsal plate surface was covered with multiple white nummular lesions; staining and X-ray microanalysis confirmed the lesions to be calcium deposits. Calcium deposits have been documented for various ophthalmic implants, including silicone scleral buckles and intraocular lenses, but to the best of our knowledge they have not previously been described in association with a glaucoma drainage device.",
author = "Khan, {Arif O.} and Hind Al-Katan and Edward, {Deepak P.}",
year = "2012",
month = "8",
doi = "10.1016/j.jaapos.2012.04.004",
language = "English (US)",
volume = "16",
pages = "401--402",
journal = "Journal of AAPOS",
issn = "1091-8531",
publisher = "Mosby Inc.",
number = "4",

}

TY - JOUR

T1 - Nummular dystrophic calcification of an Ahmed glaucoma valve in a child

AU - Khan, Arif O.

AU - Al-Katan, Hind

AU - Edward, Deepak P.

PY - 2012/8

Y1 - 2012/8

N2 - We report the case of an otherwise-healthy 4.5-year-old girl with primary congenital glaucoma who had multiple nummular calcium deposits on the dorsal plate of a previously-implanted Ahmed glaucoma valve. The child had received the implant in the right eye at 6 months of age and developed increased intraocular pressure requiring surgical excision of surrounding fibrous encapsulation. Intraoperative inspection revealed that the dorsal plate surface was covered with multiple white nummular lesions; staining and X-ray microanalysis confirmed the lesions to be calcium deposits. Calcium deposits have been documented for various ophthalmic implants, including silicone scleral buckles and intraocular lenses, but to the best of our knowledge they have not previously been described in association with a glaucoma drainage device.

AB - We report the case of an otherwise-healthy 4.5-year-old girl with primary congenital glaucoma who had multiple nummular calcium deposits on the dorsal plate of a previously-implanted Ahmed glaucoma valve. The child had received the implant in the right eye at 6 months of age and developed increased intraocular pressure requiring surgical excision of surrounding fibrous encapsulation. Intraoperative inspection revealed that the dorsal plate surface was covered with multiple white nummular lesions; staining and X-ray microanalysis confirmed the lesions to be calcium deposits. Calcium deposits have been documented for various ophthalmic implants, including silicone scleral buckles and intraocular lenses, but to the best of our knowledge they have not previously been described in association with a glaucoma drainage device.

UR - http://www.scopus.com/inward/record.url?scp=84865490275&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84865490275&partnerID=8YFLogxK

U2 - 10.1016/j.jaapos.2012.04.004

DO - 10.1016/j.jaapos.2012.04.004

M3 - Article

VL - 16

SP - 401

EP - 402

JO - Journal of AAPOS

JF - Journal of AAPOS

SN - 1091-8531

IS - 4

ER -