T1 - Monte Carlo-based evaluation of inter-crystal scatter and penetration in the PET subsystem of three GE Discovery PET/CT scanners
AU - Zeraatkar, N.
AU - Ay, M. R.
AU - Ghafarian, P.
AU - Sarkar, S.
AU - Geramifar, P.
AU - Rahmim, A.
N1 - Funding Information:
This work was supported by the Research Center for Science and Technology in Medicine and Tehran University of Medical Sciences under Grant no. 9762 .
PY - 2011/12/11
Y1 - 2011/12/11
N2 - While there is continuing demand for higher resolution in PET systems the technological improvements are still challenged by the presence of inter-crystal scatter (ICS) and inter-crystal penetration phenomena in PET detectors, which play an important role in deterioration of the spatial resolution. Both ICS and penetration have deteriorative impact on spatial resolution of PET scanners because they can lead to inaccurate incident crystal assignments. As such, an understanding of the quantitative behavior of ICS and penetration can be beneficial whether for design of a more optimized PET detection system or for more accurate modeling of ICS and penetration effects within the image reconstruction system matrix in order to enhance the quality of reconstructed images. In this work we analyzed the quantity of ICS and penetrated events in the form of coincidences, in contrast with the other studies that have assessed ICS and penetration in the form of single photons. This was performed in the PET subsystem of three GE whole-body PET/CT scanners: Discovery RX (DRX), Discovery ST (DST), and Discovery STE (DSTE). Furthermore, as a novel study, we discriminated between ICS vs. penetration events. In order to do this, we employed the GATE (Geant4 Application for Tomographic Emission) Monte Carlo (MC) toolkit for our simulations and used our previously validated GATE models of the scanners. Developing an algorithm, purely true coincidences were discriminated from ICS- and/or penetration-induced (ICSP) coincidences. ICSP coincidences were also categorized into three groups: group-1 consisted of coincidence event(s) only affected by penetration (one or both). Group-2 includes coincidences where one event is affected by ICS (possibly including penetration), while the other event is not affected by ICS (i.e. penetration or no mispositioning at all). Finally in group-3, both events are affected by ICS (possibly also including penetration). The results showed that the most magnificent quantitative variations of ICS-P occur along radial direction. In DRX, more than 55% of the true coincidences are mispositioned due to ICS and/or penetration when the source is located at the end of the transaxial field of view (FOV). This value for DST and DSTE is about 45%. Incidentally, the results revealed that the quantities of ICSP coincidences in the DST and DSTE are almost equal, while there is much smaller ICSP in the DRX.
AB - While there is continuing demand for higher resolution in PET systems the technological improvements are still challenged by the presence of inter-crystal scatter (ICS) and inter-crystal penetration phenomena in PET detectors, which play an important role in deterioration of the spatial resolution. Both ICS and penetration have deteriorative impact on spatial resolution of PET scanners because they can lead to inaccurate incident crystal assignments. As such, an understanding of the quantitative behavior of ICS and penetration can be beneficial whether for design of a more optimized PET detection system or for more accurate modeling of ICS and penetration effects within the image reconstruction system matrix in order to enhance the quality of reconstructed images. In this work we analyzed the quantity of ICS and penetrated events in the form of coincidences, in contrast with the other studies that have assessed ICS and penetration in the form of single photons. This was performed in the PET subsystem of three GE whole-body PET/CT scanners: Discovery RX (DRX), Discovery ST (DST), and Discovery STE (DSTE). Furthermore, as a novel study, we discriminated between ICS vs. penetration events. In order to do this, we employed the GATE (Geant4 Application for Tomographic Emission) Monte Carlo (MC) toolkit for our simulations and used our previously validated GATE models of the scanners. Developing an algorithm, purely true coincidences were discriminated from ICS- and/or penetration-induced (ICSP) coincidences. ICSP coincidences were also categorized into three groups: group-1 consisted of coincidence event(s) only affected by penetration (one or both). Group-2 includes coincidences where one event is affected by ICS (possibly including penetration), while the other event is not affected by ICS (i.e. penetration or no mispositioning at all). Finally in group-3, both events are affected by ICS (possibly also including penetration). The results showed that the most magnificent quantitative variations of ICS-P occur along radial direction. In DRX, more than 55% of the true coincidences are mispositioned due to ICS and/or penetration when the source is located at the end of the transaxial field of view (FOV). This value for DST and DSTE is about 45%. Incidentally, the results revealed that the quantities of ICSP coincidences in the DST and DSTE are almost equal, while there is much smaller ICSP in the DRX.
KW - Discovery RX
KW - Discovery ST
KW - Discovery STE
KW - GATE
KW - Inter-crystal scatter
KW - Monte Carlo simulations
KW - Penetration
UR - http://www.scopus.com/inward/record.url?scp=80455174681&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=80455174681&partnerID=8YFLogxK
U2 - 10.1016/j.nima.2011.07.049
DO - 10.1016/j.nima.2011.07.049
M3 - Article
AN - SCOPUS:80455174681
VL - 659
SP - 508
EP - 514
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
SN - 0168-9002
IS - 1
ER -