TY - JOUR
T1 - Metal Artifact Reduction Computed Tomography of Arthroplasty Implants
T2 - Effects of Combined Modeled Iterative Reconstruction and Dual-Energy Virtual Monoenergetic Extrapolation at Higher Photon Energies
AU - Khodarahmi, Iman
AU - Haroun, Reham R.
AU - Lee, Moses
AU - Fung, George S.K.
AU - Fuld, Matthew K.
AU - Schon, Lew C.
AU - Fishman, Elliot K.
AU - Fritz, Jan
N1 - Publisher Copyright:
© 2018 Wolters Kluwer Health, Inc. All rights reserved.
PY - 2018/12/1
Y1 - 2018/12/1
N2 - Objective The aim of this study was to compare the effects of combined virtual monoenergetic extrapolation (VME) of dual-energy computed tomography data and iterative metal artifact reduction (iMAR) at higher photon energies on low- and high-density metal artifacts and overall image quality of the ankle arthroplasty implants with iMAR, weighted filtered back projection (WFBP), and WFBP-based VME. Materials and Methods Total ankle arthroplasty implants in 6 human cadaver ankles served as surrogates for arthroplasty implants. All specimens underwent computed tomography with a 2 × 192-slice dual-source computed tomography scanner at tube voltages of 80 and tin-filtered 150 kVp to produce mixed 120 kVp equivalent polychromatic and virtual monoenergetic extrapolated images at 150 and 190 keV (VME 150 and VME 190, respectively). By implementing the WFBP and iMAR reconstruction algorithms on polychromatic, VME 150 and VME 190 data, 6 image datasets were created: WFBP-Polychromatic, iMAR-Polychromatic, WFBP-VME 150, WFBP-VME 190, iMAR-VME 150, and iMAR-VME 190. High-density and low-density artifacts were separately quantified with a threshold-based computer algorithm. After anonymization and randomization, 2 observers independently ranked the datasets for overall image quality. Repeated measures analysis of variance, Friedman, and Cohen weighted κ tests were applied for statistical analysis. A conservative P value of less than 0.001 was considered statistically significant. Results iMAR-VME 190 keV and iMAR-VME 150 keV created the least amount of high-density artifacts (all P < 0.001), whereas iMAR-Polychromatic was the most effective method to mitigate low-density streaks (P < 0.001). For low- and high-density artifacts, polychromatic iMAR acquisition was superior to WFBP-VME 150 keV and WFBP-VME 190 keV (all P < 0.001). On sharp kernel reconstructions, readers ranked the overall image quality of iMAR-Polychromatic images highest (all P < 0.001). Similarly, on soft tissue kernel reconstructions, readers ranked iMAR-Polychromatic images highest with a statistically significant difference over other techniques (all P < 0.001), except for iMAR-VME 150 keV (P = 0.356). Conclusions In computed tomography imaging of ankle arthroplasty implants, iMAR reconstruction results in fewer metal artifacts and better image quality than WFBP reconstruction for both polychromatic and virtual monoenergetic data. The combination of iMAR and VME at higher photon energies results in mixed effects on implant-induced metal artifacts, including decreased high-density and increased low-density artifacts, which in combination does not improve image quality over iMAR reconstruction of the polychromatic data. Our results suggest that, for ankle arthroplasty implants, the highest image quality is obtained by iMAR reconstruction of the polychromatic data without the need to implement VME at high-energy levels.
AB - Objective The aim of this study was to compare the effects of combined virtual monoenergetic extrapolation (VME) of dual-energy computed tomography data and iterative metal artifact reduction (iMAR) at higher photon energies on low- and high-density metal artifacts and overall image quality of the ankle arthroplasty implants with iMAR, weighted filtered back projection (WFBP), and WFBP-based VME. Materials and Methods Total ankle arthroplasty implants in 6 human cadaver ankles served as surrogates for arthroplasty implants. All specimens underwent computed tomography with a 2 × 192-slice dual-source computed tomography scanner at tube voltages of 80 and tin-filtered 150 kVp to produce mixed 120 kVp equivalent polychromatic and virtual monoenergetic extrapolated images at 150 and 190 keV (VME 150 and VME 190, respectively). By implementing the WFBP and iMAR reconstruction algorithms on polychromatic, VME 150 and VME 190 data, 6 image datasets were created: WFBP-Polychromatic, iMAR-Polychromatic, WFBP-VME 150, WFBP-VME 190, iMAR-VME 150, and iMAR-VME 190. High-density and low-density artifacts were separately quantified with a threshold-based computer algorithm. After anonymization and randomization, 2 observers independently ranked the datasets for overall image quality. Repeated measures analysis of variance, Friedman, and Cohen weighted κ tests were applied for statistical analysis. A conservative P value of less than 0.001 was considered statistically significant. Results iMAR-VME 190 keV and iMAR-VME 150 keV created the least amount of high-density artifacts (all P < 0.001), whereas iMAR-Polychromatic was the most effective method to mitigate low-density streaks (P < 0.001). For low- and high-density artifacts, polychromatic iMAR acquisition was superior to WFBP-VME 150 keV and WFBP-VME 190 keV (all P < 0.001). On sharp kernel reconstructions, readers ranked the overall image quality of iMAR-Polychromatic images highest (all P < 0.001). Similarly, on soft tissue kernel reconstructions, readers ranked iMAR-Polychromatic images highest with a statistically significant difference over other techniques (all P < 0.001), except for iMAR-VME 150 keV (P = 0.356). Conclusions In computed tomography imaging of ankle arthroplasty implants, iMAR reconstruction results in fewer metal artifacts and better image quality than WFBP reconstruction for both polychromatic and virtual monoenergetic data. The combination of iMAR and VME at higher photon energies results in mixed effects on implant-induced metal artifacts, including decreased high-density and increased low-density artifacts, which in combination does not improve image quality over iMAR reconstruction of the polychromatic data. Our results suggest that, for ankle arthroplasty implants, the highest image quality is obtained by iMAR reconstruction of the polychromatic data without the need to implement VME at high-energy levels.
KW - ankle arthroplasty
KW - dual-energy CT
KW - iterative metal artifact reduction
KW - metallic implants
KW - virtual monoenergetic extrapolation
UR - http://www.scopus.com/inward/record.url?scp=85056265176&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85056265176&partnerID=8YFLogxK
U2 - 10.1097/RLI.0000000000000497
DO - 10.1097/RLI.0000000000000497
M3 - Article
C2 - 30015677
AN - SCOPUS:85056265176
SN - 0020-9996
VL - 53
SP - 728
EP - 735
JO - Investigative Radiology
JF - Investigative Radiology
IS - 12
ER -