Low-dose or better resolution ultrafast cardiac SPECT: NCAT reconstruction and GATE study

Shruti Agarwal, Joyoni Dey

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

SPECT is primarily used in the clinic for cardiac myocardial perfusion imaging. However, for SPECT, sensitivity is impaired due to the need for collimation. System resolution is poor as well. New systems including those with multi-pinhole collimation has shown high sensitive acquisitions. Our previous work showed that using a curved detector instead of flat we can obtain resolution improvements as much as 29% on the average or else can sensitivity gains as much as 2.25 for similar resolutions. Here we use our voxel-driven MLEM reconstruction algorithm on an NCAT phantom for pinhole reconstruction to evaluate a lower dose or a higher resolution setting. The detector behind each pinhole can be flat or curved. We show that with the curved detector system (at the higher sensitivity setting, for equivalent resolution as the flat detector system), we can reduce the dose by about a factor of 1/2 still obtain equivalent image-quality as the flat-detector. Alternatively we have a better resolution compared to flat-detector system. In the second part of the study we investigated a practicable trapezoidal detector with the same height as our paraboloid detector using GATE SPECT simulator for pinhole geometry. Using point sources we investigated the sensitivity of the trapezoidal and an equivalent base flat detector for the same pinhole aperture size of 4mm. We then compared the sensitivity of each point to the approximate expected value for each point We also measured the FWHM of each point and obtained the average resolution gain of the trapezoidal over the flat detector. This constitutes the better resolution design. On the average we obtained a 31.3% improvement of resolution Then we used the trapezoidal detector with a larger pinhole aperture to match the flat-detector resolution on the average to obtain a higher sensitivity design, obtaining a sensitivity gain factor of 2.26, for similar average resolution for the two systems.

Original languageEnglish (US)
Title of host publication2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011
Pages4044-4050
Number of pages7
DOIs
StatePublished - Mar 26 2012
Externally publishedYes
Event2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011 - Valencia, Spain
Duration: Oct 23 2011Oct 29 2011

Publication series

NameIEEE Nuclear Science Symposium Conference Record
ISSN (Print)1095-7863

Other

Other2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011
CountrySpain
CityValencia
Period10/23/1110/29/11

Fingerprint

GARP Atlantic Tropical Experiment
Single-Photon Emission-Computed Tomography
dosage
detectors
pinholes
Myocardial Perfusion Imaging
sensitivity
collimation
apertures
point sources
simulators
acquisition

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Cite this

Agarwal, S., & Dey, J. (2012). Low-dose or better resolution ultrafast cardiac SPECT: NCAT reconstruction and GATE study. In 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011 (pp. 4044-4050). [6153769] (IEEE Nuclear Science Symposium Conference Record). https://doi.org/10.1109/NSSMIC.2011.6153769

Low-dose or better resolution ultrafast cardiac SPECT : NCAT reconstruction and GATE study. / Agarwal, Shruti; Dey, Joyoni.

2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011. 2012. p. 4044-4050 6153769 (IEEE Nuclear Science Symposium Conference Record).

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Agarwal, S & Dey, J 2012, Low-dose or better resolution ultrafast cardiac SPECT: NCAT reconstruction and GATE study. in 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011., 6153769, IEEE Nuclear Science Symposium Conference Record, pp. 4044-4050, 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011, Valencia, Spain, 10/23/11. https://doi.org/10.1109/NSSMIC.2011.6153769
Agarwal S, Dey J. Low-dose or better resolution ultrafast cardiac SPECT: NCAT reconstruction and GATE study. In 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011. 2012. p. 4044-4050. 6153769. (IEEE Nuclear Science Symposium Conference Record). https://doi.org/10.1109/NSSMIC.2011.6153769
Agarwal, Shruti ; Dey, Joyoni. / Low-dose or better resolution ultrafast cardiac SPECT : NCAT reconstruction and GATE study. 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011. 2012. pp. 4044-4050 (IEEE Nuclear Science Symposium Conference Record).
@inproceedings{fa656bdc54c24cc3ab69be27b728a471,
title = "Low-dose or better resolution ultrafast cardiac SPECT: NCAT reconstruction and GATE study",
abstract = "SPECT is primarily used in the clinic for cardiac myocardial perfusion imaging. However, for SPECT, sensitivity is impaired due to the need for collimation. System resolution is poor as well. New systems including those with multi-pinhole collimation has shown high sensitive acquisitions. Our previous work showed that using a curved detector instead of flat we can obtain resolution improvements as much as 29{\%} on the average or else can sensitivity gains as much as 2.25 for similar resolutions. Here we use our voxel-driven MLEM reconstruction algorithm on an NCAT phantom for pinhole reconstruction to evaluate a lower dose or a higher resolution setting. The detector behind each pinhole can be flat or curved. We show that with the curved detector system (at the higher sensitivity setting, for equivalent resolution as the flat detector system), we can reduce the dose by about a factor of 1/2 still obtain equivalent image-quality as the flat-detector. Alternatively we have a better resolution compared to flat-detector system. In the second part of the study we investigated a practicable trapezoidal detector with the same height as our paraboloid detector using GATE SPECT simulator for pinhole geometry. Using point sources we investigated the sensitivity of the trapezoidal and an equivalent base flat detector for the same pinhole aperture size of 4mm. We then compared the sensitivity of each point to the approximate expected value for each point We also measured the FWHM of each point and obtained the average resolution gain of the trapezoidal over the flat detector. This constitutes the better resolution design. On the average we obtained a 31.3{\%} improvement of resolution Then we used the trapezoidal detector with a larger pinhole aperture to match the flat-detector resolution on the average to obtain a higher sensitivity design, obtaining a sensitivity gain factor of 2.26, for similar average resolution for the two systems.",
author = "Shruti Agarwal and Joyoni Dey",
year = "2012",
month = "3",
day = "26",
doi = "10.1109/NSSMIC.2011.6153769",
language = "English (US)",
isbn = "9781467301183",
series = "IEEE Nuclear Science Symposium Conference Record",
pages = "4044--4050",
booktitle = "2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011",

}

TY - GEN

T1 - Low-dose or better resolution ultrafast cardiac SPECT

T2 - NCAT reconstruction and GATE study

AU - Agarwal, Shruti

AU - Dey, Joyoni

PY - 2012/3/26

Y1 - 2012/3/26

N2 - SPECT is primarily used in the clinic for cardiac myocardial perfusion imaging. However, for SPECT, sensitivity is impaired due to the need for collimation. System resolution is poor as well. New systems including those with multi-pinhole collimation has shown high sensitive acquisitions. Our previous work showed that using a curved detector instead of flat we can obtain resolution improvements as much as 29% on the average or else can sensitivity gains as much as 2.25 for similar resolutions. Here we use our voxel-driven MLEM reconstruction algorithm on an NCAT phantom for pinhole reconstruction to evaluate a lower dose or a higher resolution setting. The detector behind each pinhole can be flat or curved. We show that with the curved detector system (at the higher sensitivity setting, for equivalent resolution as the flat detector system), we can reduce the dose by about a factor of 1/2 still obtain equivalent image-quality as the flat-detector. Alternatively we have a better resolution compared to flat-detector system. In the second part of the study we investigated a practicable trapezoidal detector with the same height as our paraboloid detector using GATE SPECT simulator for pinhole geometry. Using point sources we investigated the sensitivity of the trapezoidal and an equivalent base flat detector for the same pinhole aperture size of 4mm. We then compared the sensitivity of each point to the approximate expected value for each point We also measured the FWHM of each point and obtained the average resolution gain of the trapezoidal over the flat detector. This constitutes the better resolution design. On the average we obtained a 31.3% improvement of resolution Then we used the trapezoidal detector with a larger pinhole aperture to match the flat-detector resolution on the average to obtain a higher sensitivity design, obtaining a sensitivity gain factor of 2.26, for similar average resolution for the two systems.

AB - SPECT is primarily used in the clinic for cardiac myocardial perfusion imaging. However, for SPECT, sensitivity is impaired due to the need for collimation. System resolution is poor as well. New systems including those with multi-pinhole collimation has shown high sensitive acquisitions. Our previous work showed that using a curved detector instead of flat we can obtain resolution improvements as much as 29% on the average or else can sensitivity gains as much as 2.25 for similar resolutions. Here we use our voxel-driven MLEM reconstruction algorithm on an NCAT phantom for pinhole reconstruction to evaluate a lower dose or a higher resolution setting. The detector behind each pinhole can be flat or curved. We show that with the curved detector system (at the higher sensitivity setting, for equivalent resolution as the flat detector system), we can reduce the dose by about a factor of 1/2 still obtain equivalent image-quality as the flat-detector. Alternatively we have a better resolution compared to flat-detector system. In the second part of the study we investigated a practicable trapezoidal detector with the same height as our paraboloid detector using GATE SPECT simulator for pinhole geometry. Using point sources we investigated the sensitivity of the trapezoidal and an equivalent base flat detector for the same pinhole aperture size of 4mm. We then compared the sensitivity of each point to the approximate expected value for each point We also measured the FWHM of each point and obtained the average resolution gain of the trapezoidal over the flat detector. This constitutes the better resolution design. On the average we obtained a 31.3% improvement of resolution Then we used the trapezoidal detector with a larger pinhole aperture to match the flat-detector resolution on the average to obtain a higher sensitivity design, obtaining a sensitivity gain factor of 2.26, for similar average resolution for the two systems.

UR - http://www.scopus.com/inward/record.url?scp=84858690229&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84858690229&partnerID=8YFLogxK

U2 - 10.1109/NSSMIC.2011.6153769

DO - 10.1109/NSSMIC.2011.6153769

M3 - Conference contribution

AN - SCOPUS:84858690229

SN - 9781467301183

T3 - IEEE Nuclear Science Symposium Conference Record

SP - 4044

EP - 4050

BT - 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011

ER -