T1 - Known-component 3D image reconstruction for improved intraoperative imaging in spine surgery
T2 - A clinical pilot study
AU - Zhang, Xiaoxuan
AU - Uneri, Ali
AU - Webster Stayman, J.
AU - Zygourakis, Corinna C.
AU - Lo, Sheng fu L.
AU - Theodore, Nicholas
AU - Siewerdsen, Jeffrey H.
N1 - Publisher Copyright:
© 2019 American Association of Physicists in Medicine
PY - 2019/8
Y1 - 2019/8
N2 - Purpose: Intraoperative imaging plays an increased role in support of surgical guidance and quality assurance for interventional approaches. However, image quality sufficient to detect complications and provide quantitative assessment of the surgical product is often confounded by image noise and artifacts. In this work, we translated a three-dimensional model-based image reconstruction (referred to as “Known-Component Reconstruction,” KC-Recon) for the first time to clinical studies with the aim of resolving both limitations. Methods: KC-Recon builds upon a penalized weighted least-squares (PWLS) method by incorporating models of surgical instrumentation (“known components”) within a joint image registration–reconstruction process to improve image quality. Under IRB approval, a clinical pilot study was conducted with 17 spine surgery patients imaged under informed consent using the O-arm cone-beam CT system (Medtronic, Littleton MA) before and after spinal instrumentation. Volumetric images were generated for each patient using KC-Recon in comparison to conventional filtered backprojection (FBP). Imaging performance prior to instrumentation (“preinstrumentation”) was evaluated in terms of soft-tissue contrast-to-noise ratio (CNR) and spatial resolution. The quality of images obtained after the instrumentation (“postinstrumentation”) was assessed by quantifying the magnitude of metal artifacts (blooming and streaks) arising from pedicle screws. The potential low-dose advantages of the algorithm were tested by simulating low-dose data (down to one-tenth of the dose of standard protocols) from images acquired at normal dose. Results: Preinstrumentation images (at normal clinical dose and matched resolution) exhibited an average 24.0% increase in soft-tissue CNR with KC-Recon compared to FBP (N = 16, P = 0.02), improving visualization of paraspinal muscles, major vessels, and other soft-tissues about the spine and abdomen. For a total of 72 screws in postinstrumentation images, KC-Recon yielded a significant reduction in metal artifacts: 66.3% reduction in overestimation of screw shaft width due to blooming (P < 0.0001) and reduction in streaks at the screw tip (65.8% increase in attenuation accuracy, P < 0.0001), enabling clearer depiction of the screw within the pedicle and vertebral body for an assessment of breach. Depending on the imaging task, dose reduction up to an order of magnitude appeared feasible while maintaining soft-tissue visibility and metal artifact reduction. Conclusions: KC-Recon offers a promising means to improve visualization in the presence of surgical instrumentation and reduce patient dose in image-guided procedures. The improved soft-tissue visibility could facilitate the use of cone-beam CT to soft-tissue surgeries, and the ability to precisely quantify and visualize instrument placement could provide a valuable check against complications in the operating room (cf., postoperative CT).
AB - Purpose: Intraoperative imaging plays an increased role in support of surgical guidance and quality assurance for interventional approaches. However, image quality sufficient to detect complications and provide quantitative assessment of the surgical product is often confounded by image noise and artifacts. In this work, we translated a three-dimensional model-based image reconstruction (referred to as “Known-Component Reconstruction,” KC-Recon) for the first time to clinical studies with the aim of resolving both limitations. Methods: KC-Recon builds upon a penalized weighted least-squares (PWLS) method by incorporating models of surgical instrumentation (“known components”) within a joint image registration–reconstruction process to improve image quality. Under IRB approval, a clinical pilot study was conducted with 17 spine surgery patients imaged under informed consent using the O-arm cone-beam CT system (Medtronic, Littleton MA) before and after spinal instrumentation. Volumetric images were generated for each patient using KC-Recon in comparison to conventional filtered backprojection (FBP). Imaging performance prior to instrumentation (“preinstrumentation”) was evaluated in terms of soft-tissue contrast-to-noise ratio (CNR) and spatial resolution. The quality of images obtained after the instrumentation (“postinstrumentation”) was assessed by quantifying the magnitude of metal artifacts (blooming and streaks) arising from pedicle screws. The potential low-dose advantages of the algorithm were tested by simulating low-dose data (down to one-tenth of the dose of standard protocols) from images acquired at normal dose. Results: Preinstrumentation images (at normal clinical dose and matched resolution) exhibited an average 24.0% increase in soft-tissue CNR with KC-Recon compared to FBP (N = 16, P = 0.02), improving visualization of paraspinal muscles, major vessels, and other soft-tissues about the spine and abdomen. For a total of 72 screws in postinstrumentation images, KC-Recon yielded a significant reduction in metal artifacts: 66.3% reduction in overestimation of screw shaft width due to blooming (P < 0.0001) and reduction in streaks at the screw tip (65.8% increase in attenuation accuracy, P < 0.0001), enabling clearer depiction of the screw within the pedicle and vertebral body for an assessment of breach. Depending on the imaging task, dose reduction up to an order of magnitude appeared feasible while maintaining soft-tissue visibility and metal artifact reduction. Conclusions: KC-Recon offers a promising means to improve visualization in the presence of surgical instrumentation and reduce patient dose in image-guided procedures. The improved soft-tissue visibility could facilitate the use of cone-beam CT to soft-tissue surgeries, and the ability to precisely quantify and visualize instrument placement could provide a valuable check against complications in the operating room (cf., postoperative CT).
KW - cone-beam CT
KW - image-guided procedures
KW - intraoperative imaging
KW - model-based image reconstruction
KW - patient safety
UR - http://www.scopus.com/inward/record.url?scp=85068415403&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85068415403&partnerID=8YFLogxK
U2 - 10.1002/mp.13652
DO - 10.1002/mp.13652
M3 - Article
C2 - 31180586
AN - SCOPUS:85068415403
SN - 0094-2405
VL - 46
SP - 3483
EP - 3495
JO - Medical physics
JF - Medical physics
IS - 8
ER -