Intra- and inter-radiation therapist reproducibility of daily isocenter verification using prostatic fiducial markers

Karen L. Ullman, Holly Ning, Robert C. Susil, Asna Ayele, Lucresse Jocelyn, Jan Havelos, Peter Guion, Huchen Xie, Guang Li, Barbara C. Arora, Angela Cannon, Robert W. Miller, C. Norman Coleman, Kevin Camphausen, Cynthia Ménard

Research output: Contribution to journalArticlepeer-review

20 Scopus citations

Abstract

Background: We sought to determine the intra- and inter-radiation therapist reproducibility of a previously established matching technique for daily verification and correction of isocenter position relative to intraprostatic fiducial markers (FM). Materials and methods: With the patient in the treatment position, anterior-posterior and left lateral electronic images are acquired on an amorphous silicon flat panel electronic portal imaging device. After each portal image is acquired, the therapist manually translates and aligns the fiducial markers in the image to the marker contours on the digitally reconstructed radiograph. The distances between the planned and actual isocenter location is displayed. In order to determine the reproducibility of this technique, four therapists repeated and recorded this operation two separate times on 20 previously acquired portal image datasets from two patients. The data were analyzed to obtain the mean variability in the distances measured between and within observers. Results: The mean and median intra-observer variability ranged from 0.4 to 0.7 mm and 0.3 to 0.6 mm respectively with a standard deviation of 0.4 to 1.0 mm. Inter-observer results were similar with a mean variability of 0.9 mm, a median of 0.6 mm, and a standard deviation of 0.7 mm. When using a 5 mm threshold, only 0.5% of treatments will undergo a table shift due to intra or inter-observer error, increasing to an error rate of 2.4% if this threshold were reduced to 3 mm. Conclusion: We have found high reproducibility with a previously established method for daily verification and correction of isocenter position relative to prostatic fiducial markers using electronic portal imaging.

Original languageEnglish (US)
Article number2
JournalRadiation Oncology
Volume1
Issue number1
DOIs
StatePublished - Feb 28 2006
Externally publishedYes

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Intra- and inter-radiation therapist reproducibility of daily isocenter verification using prostatic fiducial markers'. Together they form a unique fingerprint.

Cite this