Intercomparison of instruments used for safety and performance measurements of ultraviolet germicidal irradiation lamps.

Nicholas G. Reed, Stephen Wengraitis, David H. Sliney

Research output: Contribution to journalArticle

Abstract

Public health concerns such as multidrug-resistant tuberculosis, bioterrorism, pandemic influenza, and sick building syndrome have brought about increased interest in the use of ultraviolet germicidal irradiation (UVGI) to prevent the spread of airborne infection. UVGI lamps require that radiometric measurements be performed to ensure their safe and effective use. This study evaluates 10 detectors that measure the ultraviolet radiation hazard of low-pressure mercury UVGI lamps, including a polychromator spectroradiometer, narrowband detectors designed to measure the ultraviolet radiation in a short range of wavelengths, and broadband detectors with a varying spectral response designed to follow the UV hazard action spectrum. The angular responses, spectral responses, and linearity of the detectors were measured and compared. The agreement between the measured angular responses and the ideal cosine responses varied widely among the detectors, and in general, the detectors with diffusing optics agreed significantly better with the ideal cosine response. The spectral responses at 254 nm also varied widely among the detectors, and, in general, the narrowband detectors agreed more closely with the 254 nm irradiances measured under the same conditions by a double monochromator spectroradiometer. All detectors displayed good linearity. The angular and spectral response data were then used to develop correction factors for the effective irradiance measurements of two UVGI sources, each measured at 10, 20, and 30 cm. The measured effective irradiances were compared with those measured by a double monochromator spectroradiometer with an integrating sphere input optic. Prior to correction, the effective irradiances measured by the detectors varied widely, ranging from 0.29 to 2.7 times those measured by the spectroradiometer. The application of cosine and spectral response correction factors significantly improved the agreement for the effective irradiances measured by all of the detectors, typically to within 10-20%. Awareness of these detector characteristics can play a key role in ensuring the accuracy of health hazard measurements of UVGI lamps.

Original languageEnglish (US)
Pages (from-to)289-297
Number of pages9
JournalJournal of Occupational and Environmental Hygiene
Volume6
Issue number5
DOIs
StatePublished - May 2009
Externally publishedYes

Fingerprint

Sick Building Syndrome
Radiation
Bioterrorism
Safety
Multidrug-Resistant Tuberculosis
Pandemics
Mercury
Human Influenza
Public Health
Pressure
Health
Infection
Action Spectrum

ASJC Scopus subject areas

  • Public Health, Environmental and Occupational Health

Cite this

Intercomparison of instruments used for safety and performance measurements of ultraviolet germicidal irradiation lamps. / Reed, Nicholas G.; Wengraitis, Stephen; Sliney, David H.

In: Journal of Occupational and Environmental Hygiene, Vol. 6, No. 5, 05.2009, p. 289-297.

Research output: Contribution to journalArticle

@article{fe578440990f4931bb7e86d944bacf82,
title = "Intercomparison of instruments used for safety and performance measurements of ultraviolet germicidal irradiation lamps.",
abstract = "Public health concerns such as multidrug-resistant tuberculosis, bioterrorism, pandemic influenza, and sick building syndrome have brought about increased interest in the use of ultraviolet germicidal irradiation (UVGI) to prevent the spread of airborne infection. UVGI lamps require that radiometric measurements be performed to ensure their safe and effective use. This study evaluates 10 detectors that measure the ultraviolet radiation hazard of low-pressure mercury UVGI lamps, including a polychromator spectroradiometer, narrowband detectors designed to measure the ultraviolet radiation in a short range of wavelengths, and broadband detectors with a varying spectral response designed to follow the UV hazard action spectrum. The angular responses, spectral responses, and linearity of the detectors were measured and compared. The agreement between the measured angular responses and the ideal cosine responses varied widely among the detectors, and in general, the detectors with diffusing optics agreed significantly better with the ideal cosine response. The spectral responses at 254 nm also varied widely among the detectors, and, in general, the narrowband detectors agreed more closely with the 254 nm irradiances measured under the same conditions by a double monochromator spectroradiometer. All detectors displayed good linearity. The angular and spectral response data were then used to develop correction factors for the effective irradiance measurements of two UVGI sources, each measured at 10, 20, and 30 cm. The measured effective irradiances were compared with those measured by a double monochromator spectroradiometer with an integrating sphere input optic. Prior to correction, the effective irradiances measured by the detectors varied widely, ranging from 0.29 to 2.7 times those measured by the spectroradiometer. The application of cosine and spectral response correction factors significantly improved the agreement for the effective irradiances measured by all of the detectors, typically to within 10-20{\%}. Awareness of these detector characteristics can play a key role in ensuring the accuracy of health hazard measurements of UVGI lamps.",
author = "Reed, {Nicholas G.} and Stephen Wengraitis and Sliney, {David H.}",
year = "2009",
month = "5",
doi = "10.1080/15459620902801041",
language = "English (US)",
volume = "6",
pages = "289--297",
journal = "Journal of Occupational and Environmental Hygiene",
issn = "1545-9624",
publisher = "Taylor and Francis Ltd.",
number = "5",

}

TY - JOUR

T1 - Intercomparison of instruments used for safety and performance measurements of ultraviolet germicidal irradiation lamps.

AU - Reed, Nicholas G.

AU - Wengraitis, Stephen

AU - Sliney, David H.

PY - 2009/5

Y1 - 2009/5

N2 - Public health concerns such as multidrug-resistant tuberculosis, bioterrorism, pandemic influenza, and sick building syndrome have brought about increased interest in the use of ultraviolet germicidal irradiation (UVGI) to prevent the spread of airborne infection. UVGI lamps require that radiometric measurements be performed to ensure their safe and effective use. This study evaluates 10 detectors that measure the ultraviolet radiation hazard of low-pressure mercury UVGI lamps, including a polychromator spectroradiometer, narrowband detectors designed to measure the ultraviolet radiation in a short range of wavelengths, and broadband detectors with a varying spectral response designed to follow the UV hazard action spectrum. The angular responses, spectral responses, and linearity of the detectors were measured and compared. The agreement between the measured angular responses and the ideal cosine responses varied widely among the detectors, and in general, the detectors with diffusing optics agreed significantly better with the ideal cosine response. The spectral responses at 254 nm also varied widely among the detectors, and, in general, the narrowband detectors agreed more closely with the 254 nm irradiances measured under the same conditions by a double monochromator spectroradiometer. All detectors displayed good linearity. The angular and spectral response data were then used to develop correction factors for the effective irradiance measurements of two UVGI sources, each measured at 10, 20, and 30 cm. The measured effective irradiances were compared with those measured by a double monochromator spectroradiometer with an integrating sphere input optic. Prior to correction, the effective irradiances measured by the detectors varied widely, ranging from 0.29 to 2.7 times those measured by the spectroradiometer. The application of cosine and spectral response correction factors significantly improved the agreement for the effective irradiances measured by all of the detectors, typically to within 10-20%. Awareness of these detector characteristics can play a key role in ensuring the accuracy of health hazard measurements of UVGI lamps.

AB - Public health concerns such as multidrug-resistant tuberculosis, bioterrorism, pandemic influenza, and sick building syndrome have brought about increased interest in the use of ultraviolet germicidal irradiation (UVGI) to prevent the spread of airborne infection. UVGI lamps require that radiometric measurements be performed to ensure their safe and effective use. This study evaluates 10 detectors that measure the ultraviolet radiation hazard of low-pressure mercury UVGI lamps, including a polychromator spectroradiometer, narrowband detectors designed to measure the ultraviolet radiation in a short range of wavelengths, and broadband detectors with a varying spectral response designed to follow the UV hazard action spectrum. The angular responses, spectral responses, and linearity of the detectors were measured and compared. The agreement between the measured angular responses and the ideal cosine responses varied widely among the detectors, and in general, the detectors with diffusing optics agreed significantly better with the ideal cosine response. The spectral responses at 254 nm also varied widely among the detectors, and, in general, the narrowband detectors agreed more closely with the 254 nm irradiances measured under the same conditions by a double monochromator spectroradiometer. All detectors displayed good linearity. The angular and spectral response data were then used to develop correction factors for the effective irradiance measurements of two UVGI sources, each measured at 10, 20, and 30 cm. The measured effective irradiances were compared with those measured by a double monochromator spectroradiometer with an integrating sphere input optic. Prior to correction, the effective irradiances measured by the detectors varied widely, ranging from 0.29 to 2.7 times those measured by the spectroradiometer. The application of cosine and spectral response correction factors significantly improved the agreement for the effective irradiances measured by all of the detectors, typically to within 10-20%. Awareness of these detector characteristics can play a key role in ensuring the accuracy of health hazard measurements of UVGI lamps.

UR - http://www.scopus.com/inward/record.url?scp=66949119696&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=66949119696&partnerID=8YFLogxK

U2 - 10.1080/15459620902801041

DO - 10.1080/15459620902801041

M3 - Article

C2 - 19259871

AN - SCOPUS:66949119696

VL - 6

SP - 289

EP - 297

JO - Journal of Occupational and Environmental Hygiene

JF - Journal of Occupational and Environmental Hygiene

SN - 1545-9624

IS - 5

ER -