TY - JOUR
T1 - Inter- and intrafraction patient positioning uncertainties for intracranial radiotherapy
T2 - A study of four frameless, thermoplastic mask-based immobilization strategies using daily cone-beam CT
AU - Tryggestad, Erik
AU - Christian, Matthew
AU - Ford, Eric
AU - Kut, Carmen
AU - Le, Yi
AU - Sanguineti, Giuseppe
AU - Song, Danny Y.
AU - Kleinberg, Lawrence
PY - 2011/5/1
Y1 - 2011/5/1
N2 - Purpose: To determine whether frameless thermoplastic mask-based immobilization is adequate for image-guided cranial radiosurgery. Methods and Materials: Cone-beam CT localization data from patients with intracranial tumors were studied using daily pre- and posttreatment scans. The systems studied were (1) Type-S IMRT (head only) mask (Civco) with head cushion; (2) Uni-Frame mask (Civco) with head cushion, coupled with a BlueBag body immobilizer (Medical Intelligence); (3) Type-S head and shoulder mask with head and shoulder cushion (Civco); (4) same as previous, coupled with a mouthpiece. The comparative metrics were translational shift magnitude and average rotation angle; systematic inter-, random inter-, and random intrafraction positioning error was computed. For strategies 1-4, respectively, the analysis for interfraction variability included data from 20, 9, 81, and 11 patients, whereas that for intrafraction variability included a subset of 7, 9, 16, and 8 patients. The results were compared for statistical significance using an analysis of variance test. Results: Immobilization system 4 provided the best overall accuracy and stability. The mean interfraction translational shifts (± SD) were 2.3 (± 1.4), 2.2 (± 1.1), 2.7 (± 1.5), and 2.1 (± 1.0) mm whereas intrafraction motion was 1.1 (± 1.2), 1.1 (± 1.1), 0.7 (± 0.9), and 0.7 (± 0.8) mm for devices 1-4, respectively. No significant correlation between intrafraction motion and treatment time was evident, although intrafraction motion was not purely random. Conclusions: We find that all frameless thermoplastic mask systems studied are viable solutions for image-guided intracranial radiosurgery. With daily pretreatment corrections, symmetric PTV margins of 1 mm would likely be adequate if ideal radiation planning and targeting systems were available.
AB - Purpose: To determine whether frameless thermoplastic mask-based immobilization is adequate for image-guided cranial radiosurgery. Methods and Materials: Cone-beam CT localization data from patients with intracranial tumors were studied using daily pre- and posttreatment scans. The systems studied were (1) Type-S IMRT (head only) mask (Civco) with head cushion; (2) Uni-Frame mask (Civco) with head cushion, coupled with a BlueBag body immobilizer (Medical Intelligence); (3) Type-S head and shoulder mask with head and shoulder cushion (Civco); (4) same as previous, coupled with a mouthpiece. The comparative metrics were translational shift magnitude and average rotation angle; systematic inter-, random inter-, and random intrafraction positioning error was computed. For strategies 1-4, respectively, the analysis for interfraction variability included data from 20, 9, 81, and 11 patients, whereas that for intrafraction variability included a subset of 7, 9, 16, and 8 patients. The results were compared for statistical significance using an analysis of variance test. Results: Immobilization system 4 provided the best overall accuracy and stability. The mean interfraction translational shifts (± SD) were 2.3 (± 1.4), 2.2 (± 1.1), 2.7 (± 1.5), and 2.1 (± 1.0) mm whereas intrafraction motion was 1.1 (± 1.2), 1.1 (± 1.1), 0.7 (± 0.9), and 0.7 (± 0.8) mm for devices 1-4, respectively. No significant correlation between intrafraction motion and treatment time was evident, although intrafraction motion was not purely random. Conclusions: We find that all frameless thermoplastic mask systems studied are viable solutions for image-guided intracranial radiosurgery. With daily pretreatment corrections, symmetric PTV margins of 1 mm would likely be adequate if ideal radiation planning and targeting systems were available.
KW - Frameless thermoplastic mask immobilization
KW - Image-guided cranial radiosurgery
KW - Interfraction positioning uncertainties
KW - Intrafraction positioning uncertainties
KW - Patient positioning accuracy
UR - http://www.scopus.com/inward/record.url?scp=79954424802&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79954424802&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2010.06.022
DO - 10.1016/j.ijrobp.2010.06.022
M3 - Article
C2 - 20951506
AN - SCOPUS:79954424802
SN - 0360-3016
VL - 80
SP - 281
EP - 290
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 1
ER -