Initial characterization of the Siemens E.CAM+: a dual-detector camera with coincidence imaging capability

S. D. Wollenweber, B. M.W. Tsui, D. S. Lalush, E. C. Frey, W. H. McCartney, C. Tocharoenchai, T. D. Rempel, V. Simcic

Research output: Chapter in Book/Report/Conference proceedingConference contribution

5 Scopus citations

Abstract

A study was carried out to characterize the performance of the new E.CAM+ dual-gamma camera imaging system with coincidence imaging capability. The intrinsic spatial resolution, scatter fraction, sensitivity, count-rate capability, energy resolution, and clinical image quality were evaluated. The results show that the dual-headed NaI[Tl]-based gamma camera will be suitable for FDG imaging in oncologic applications with superior energy resolution but inferior sensitivity as compared to traditional cylindrical PET tomographs.

Original languageEnglish (US)
Title of host publicationIEEE Nuclear Science Symposium and Medical Imaging Conference
PublisherIEEE
Pages1470-1476
Number of pages7
ISBN (Print)0780350227
StatePublished - 1999
Externally publishedYes
EventProceedings of the 1998 IEEE Nuclear Science Symposium Conference Record - Toronto, Que, Can
Duration: Nov 8 1998Nov 14 1998

Publication series

NameIEEE Nuclear Science Symposium and Medical Imaging Conference
Volume3

Other

OtherProceedings of the 1998 IEEE Nuclear Science Symposium Conference Record
CityToronto, Que, Can
Period11/8/9811/14/98

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging