Abstract
Postpartum hemorrhage (PPH) is the leading cause of maternal mortality worldwide, and effective interventions for addressing PPH are urgently needed. Uterine balloon tamponade (UBT) is a technique to control PPH. Commercially available UBT devices are expensive and frequently require imaging technology to ensure placement. Condom-catheter uterine tamponade (C-UBT) is a technique appropriate for low-resource settings. Testing of the C-UBT is needed to better understand and optimize this technique for use in a variety of clinical settings including low-resource contexts. We describe here the design, development, and bench-top validation of a reusable C-UBT device optimized for low-resource settings. The device was tested in three differently sized uterine models using a variety of condom balloon configurations. Intrauterine wall pressure application was measured to evaluate the device capacity to apply pressure of at least 90 mmHg, estimating the mean arterial pressure within the uterine vasculature. Bench-top experimental validation of pressure exerted in uterine models demonstrated the device's capability of reaching hemostatic pressure in uterine volumes ranging from 170 to 1740 mL. Device adaptability and versatility were shown through its ability to reach the target pressure of 90 mmHg in different uterine sizes by varying balloon parameters, including condom thickness and condom configuration. The results of this study show the potential of a low-cost, reusable C-UBT device optimized to treat PPH in a variety of clinical settings, including low-resource contexts.
Original language | English (US) |
---|---|
Article number | 014503 EN |
Journal | Journal of Medical Devices, Transactions of the ASME |
Volume | 14 |
Issue number | 1 |
DOIs | |
State | Published - Mar 1 2020 |
Keywords
- condom catheter
- global health
- obstetrical hemorrhage
- uterine balloon tamponade
ASJC Scopus subject areas
- Medicine (miscellaneous)
- Biomedical Engineering