T1 - Impact of fluoroscopy unit on the accuracy of a magnet-based electroanatomic mapping and navigation system
T2 - An in vitro and in vivo validation study
AU - Ahmad, Ghada
AU - Hussein, Ayman A.
AU - Mesubi, Olurotimi
AU - Tian, Jing
AU - Ferieg, Hanaa
AU - Elmaaty, Mervat A.
AU - Hamdy, Amal
AU - Ego-Osuala, Kelechi
AU - Jimenez, Alejandro
AU - See, Vincent
AU - Saliaris, Anastasios
AU - Shorofsky, Stephen
AU - Dickfeld, Timm
PY - 2014/2
Y1 - 2014/2
N2 - Introduction During mapping and ablation procedures, the movement of large ferromagnetic items (i.e., fluoroscopic equipment) introduce heterogeneities in the electromagnetic field, which may affect the accuracy of electromagnet-based navigation. We aimed to assess the impact of common periprocedural fluoroscopic equipment movement on the accuracy of an electromagnet-based navigation system. Methods and Results The impact of fluoroscopic equipment movement on the accuracy of the Carto® 3 System (Biosense Webster, Inc., Diamond Bar, CA, USA) was assessed both in vitro (n = 20 patients, phantom model) and in vivo (n = 18 patients). Location recordings were obtained with unchanged catheter position for fluoroscopic equipment rotational movements (RMs) and maximal to closest distance (MD to CD) to phantom/patient. The effects of both single- and biplane fluoroscopy were assessed. In vitro, the movement of fluoroscopic equipment resulted in an average catheter location estimation error of 0.8 mm (interquartile range 0.3-1.3). The maximal location estimation errors with MD to CD movement and RM were 2.3 mm and 1.3 mm, respectively. Changing from single-plane to biplane setup resulted in an average location estimation change of 1.5 mm (maximum 2.1). Larger location changes were observed in vivo (2.9 mm vs 0.8 mm, P < 0.0001) with 28.7% of these exceeded 4 mm versus none of the in vitro measurements (P < 0.0001). Conclusion Although fluoroscopy manipulation affected the accuracy of the Carto® 3 System, the in vitro data suggest that these inaccuracies are likely of limited clinical consequences. The larger in vivo inaccuracies are most likely due to nonferromagnetic interferences, such as respiratory or cardiac movements.
AB - Introduction During mapping and ablation procedures, the movement of large ferromagnetic items (i.e., fluoroscopic equipment) introduce heterogeneities in the electromagnetic field, which may affect the accuracy of electromagnet-based navigation. We aimed to assess the impact of common periprocedural fluoroscopic equipment movement on the accuracy of an electromagnet-based navigation system. Methods and Results The impact of fluoroscopic equipment movement on the accuracy of the Carto® 3 System (Biosense Webster, Inc., Diamond Bar, CA, USA) was assessed both in vitro (n = 20 patients, phantom model) and in vivo (n = 18 patients). Location recordings were obtained with unchanged catheter position for fluoroscopic equipment rotational movements (RMs) and maximal to closest distance (MD to CD) to phantom/patient. The effects of both single- and biplane fluoroscopy were assessed. In vitro, the movement of fluoroscopic equipment resulted in an average catheter location estimation error of 0.8 mm (interquartile range 0.3-1.3). The maximal location estimation errors with MD to CD movement and RM were 2.3 mm and 1.3 mm, respectively. Changing from single-plane to biplane setup resulted in an average location estimation change of 1.5 mm (maximum 2.1). Larger location changes were observed in vivo (2.9 mm vs 0.8 mm, P < 0.0001) with 28.7% of these exceeded 4 mm versus none of the in vitro measurements (P < 0.0001). Conclusion Although fluoroscopy manipulation affected the accuracy of the Carto® 3 System, the in vitro data suggest that these inaccuracies are likely of limited clinical consequences. The larger in vivo inaccuracies are most likely due to nonferromagnetic interferences, such as respiratory or cardiac movements.
KW - arrhythmias
KW - mapping
KW - navigation systems
UR - http://www.scopus.com/inward/record.url?scp=84893795052&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84893795052&partnerID=8YFLogxK
U2 - 10.1111/pace.12231
DO - 10.1111/pace.12231
M3 - Article
C2 - 24102109
AN - SCOPUS:84893795052
SN - 0147-8389
VL - 37
SP - 157
EP - 163
JO - PACE - Pacing and Clinical Electrophysiology
JF - PACE - Pacing and Clinical Electrophysiology
IS - 2
ER -