Abstract
We report new optics designs for an optical coherence tomography (OCT) balloon imaging catheter to achieve diffraction-limited high resolution at a large working distance and enable the correction of severe astigmatism in the catheter. The designs employed a 1 mm diameter gradient-index lens of a properly chosen pitch number and a glass rod spacer to fully utilize the available NA of the miniature optics. Astigmatism caused by the balloon tubing was analyzed, and a method based on a cylindrical reflector was proposed and demonstrated to compensate the astigmatism. A catheter based on the new designs was successfully developed with a measured diffraction-limited lateral resolution of ∼21 μm, a working distance of ∼ 11-12 mm, and a round-shape beam profile. The performance of the OCT balloon catheter was demonstrated by 3D fullcircumferential imaging of a swine esophagus in vivo along with a high-speed, Fourier-domain, mode-locked swept-source OCT system.
Original language | English (US) |
---|---|
Pages (from-to) | 1943-1945 |
Number of pages | 3 |
Journal | Optics Letters |
Volume | 34 |
Issue number | 13 |
DOIs | |
State | Published - Jul 1 2009 |
ASJC Scopus subject areas
- Atomic and Molecular Physics, and Optics