High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses

Angela Dagirmanjian, Jeffrey S. Ross, Nancy Obuchowski, Jonathan S. Lewin, Jean A. Tkach, Paul M. Ruggieri, Thomas J. Masaryk

Research output: Contribution to journalArticle

Abstract

Objective: Factors that restrict 3D TOF MRA are limited resolution, saturation of flow, and degree of background suppression. We evaluated MRA for intracranial stenoses by using a 3D TOF technique that minimizes these factors. Materials and Methods: Twenty-nine patients underwent MRA and intraarterial digital subtraction angiography (DSA). The MRA studies were performed on a 1.5 T Siemens SP 4000 system. Integrated techniques applied to the conventional 3D TOF acquisition included the following: (a) 256 x 256 matrix with a 140 mm FOV and 0.9 mm slice thickness, yielding a 0.54 x 0.54 x 0.9 mm3 voxel; (b) tilted optimized nonsaturating excitation (TONE); and (c) magnetization transfer saturation (MTS). The intraarterial DSA was performed on a Siemens Angiostar system with a 1,024 x 1,024 noninterpolated matrix. The MRAs were reviewed by two neuroradiologists. Two hundred seventy-seven vessels were evaluated for a total of 806 segments. Vessel segments were evaluated with a 5 point scale. Results: The estimated accuracy of MRA for detecting stenosis over all intracranial vessel segments was 0.88 ± 0.03 and 0.89 ± 0.02 for the two readers, respectively. The estimated accuracy ranged from 0.94 ± 0.02 and 0.93 ± 0.02 for detecting internal carotid artery stenosis by the two readers, respectively, to 0.65 ± 0.17 and 0.71 ± 0.15 for detecting distal vertebral artery stenosis. In vessels determined by catheter angiography to be stenosis-free, reader confidence at the proximal versus distal segments was similar for the internal carotid, basilar, and posterior cerebral arteries. However, for the anterior and middle cerebral arteries, one or both readers were more confident in diagnosing the proximal segment. Conclusion: High resolution MTS TONE 3D TOF MRA is an accurate technique for the screening of medium and large vessel intracranial stenoses.

Original languageEnglish (US)
Pages (from-to)700-706
Number of pages7
JournalJournal of Computer Assisted Tomography
Volume19
Issue number5
StatePublished - 1995
Externally publishedYes

Fingerprint

Blood Vessels
Pathologic Constriction
Digital Subtraction Angiography
Vertebrobasilar Insufficiency
Posterior Cerebral Artery
Anterior Cerebral Artery
Carotid Stenosis
Middle Cerebral Artery
Angiography
Catheters

Keywords

  • Brain, blood flow
  • Magnetic resonance imaging angiography (MRA)
  • Magnetic resonance imaging, techniques
  • Magnetization transfer

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Dagirmanjian, A., Ross, J. S., Obuchowski, N., Lewin, J. S., Tkach, J. A., Ruggieri, P. M., & Masaryk, T. J. (1995). High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses. Journal of Computer Assisted Tomography, 19(5), 700-706.

High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses. / Dagirmanjian, Angela; Ross, Jeffrey S.; Obuchowski, Nancy; Lewin, Jonathan S.; Tkach, Jean A.; Ruggieri, Paul M.; Masaryk, Thomas J.

In: Journal of Computer Assisted Tomography, Vol. 19, No. 5, 1995, p. 700-706.

Research output: Contribution to journalArticle

Dagirmanjian, A, Ross, JS, Obuchowski, N, Lewin, JS, Tkach, JA, Ruggieri, PM & Masaryk, TJ 1995, 'High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses', Journal of Computer Assisted Tomography, vol. 19, no. 5, pp. 700-706.
Dagirmanjian, Angela ; Ross, Jeffrey S. ; Obuchowski, Nancy ; Lewin, Jonathan S. ; Tkach, Jean A. ; Ruggieri, Paul M. ; Masaryk, Thomas J. / High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses. In: Journal of Computer Assisted Tomography. 1995 ; Vol. 19, No. 5. pp. 700-706.
@article{1c5d8c9b3563495f9d25fdbf8a9bbbae,
title = "High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses",
abstract = "Objective: Factors that restrict 3D TOF MRA are limited resolution, saturation of flow, and degree of background suppression. We evaluated MRA for intracranial stenoses by using a 3D TOF technique that minimizes these factors. Materials and Methods: Twenty-nine patients underwent MRA and intraarterial digital subtraction angiography (DSA). The MRA studies were performed on a 1.5 T Siemens SP 4000 system. Integrated techniques applied to the conventional 3D TOF acquisition included the following: (a) 256 x 256 matrix with a 140 mm FOV and 0.9 mm slice thickness, yielding a 0.54 x 0.54 x 0.9 mm3 voxel; (b) tilted optimized nonsaturating excitation (TONE); and (c) magnetization transfer saturation (MTS). The intraarterial DSA was performed on a Siemens Angiostar system with a 1,024 x 1,024 noninterpolated matrix. The MRAs were reviewed by two neuroradiologists. Two hundred seventy-seven vessels were evaluated for a total of 806 segments. Vessel segments were evaluated with a 5 point scale. Results: The estimated accuracy of MRA for detecting stenosis over all intracranial vessel segments was 0.88 ± 0.03 and 0.89 ± 0.02 for the two readers, respectively. The estimated accuracy ranged from 0.94 ± 0.02 and 0.93 ± 0.02 for detecting internal carotid artery stenosis by the two readers, respectively, to 0.65 ± 0.17 and 0.71 ± 0.15 for detecting distal vertebral artery stenosis. In vessels determined by catheter angiography to be stenosis-free, reader confidence at the proximal versus distal segments was similar for the internal carotid, basilar, and posterior cerebral arteries. However, for the anterior and middle cerebral arteries, one or both readers were more confident in diagnosing the proximal segment. Conclusion: High resolution MTS TONE 3D TOF MRA is an accurate technique for the screening of medium and large vessel intracranial stenoses.",
keywords = "Brain, blood flow, Magnetic resonance imaging angiography (MRA), Magnetic resonance imaging, techniques, Magnetization transfer",
author = "Angela Dagirmanjian and Ross, {Jeffrey S.} and Nancy Obuchowski and Lewin, {Jonathan S.} and Tkach, {Jean A.} and Ruggieri, {Paul M.} and Masaryk, {Thomas J.}",
year = "1995",
language = "English (US)",
volume = "19",
pages = "700--706",
journal = "Journal of Computer Assisted Tomography",
issn = "0363-8715",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - High resolution, magnetization transfer saturation, variable flip angle, time-of-flight mr a in the detection of intracranial vascular stenoses

AU - Dagirmanjian, Angela

AU - Ross, Jeffrey S.

AU - Obuchowski, Nancy

AU - Lewin, Jonathan S.

AU - Tkach, Jean A.

AU - Ruggieri, Paul M.

AU - Masaryk, Thomas J.

PY - 1995

Y1 - 1995

N2 - Objective: Factors that restrict 3D TOF MRA are limited resolution, saturation of flow, and degree of background suppression. We evaluated MRA for intracranial stenoses by using a 3D TOF technique that minimizes these factors. Materials and Methods: Twenty-nine patients underwent MRA and intraarterial digital subtraction angiography (DSA). The MRA studies were performed on a 1.5 T Siemens SP 4000 system. Integrated techniques applied to the conventional 3D TOF acquisition included the following: (a) 256 x 256 matrix with a 140 mm FOV and 0.9 mm slice thickness, yielding a 0.54 x 0.54 x 0.9 mm3 voxel; (b) tilted optimized nonsaturating excitation (TONE); and (c) magnetization transfer saturation (MTS). The intraarterial DSA was performed on a Siemens Angiostar system with a 1,024 x 1,024 noninterpolated matrix. The MRAs were reviewed by two neuroradiologists. Two hundred seventy-seven vessels were evaluated for a total of 806 segments. Vessel segments were evaluated with a 5 point scale. Results: The estimated accuracy of MRA for detecting stenosis over all intracranial vessel segments was 0.88 ± 0.03 and 0.89 ± 0.02 for the two readers, respectively. The estimated accuracy ranged from 0.94 ± 0.02 and 0.93 ± 0.02 for detecting internal carotid artery stenosis by the two readers, respectively, to 0.65 ± 0.17 and 0.71 ± 0.15 for detecting distal vertebral artery stenosis. In vessels determined by catheter angiography to be stenosis-free, reader confidence at the proximal versus distal segments was similar for the internal carotid, basilar, and posterior cerebral arteries. However, for the anterior and middle cerebral arteries, one or both readers were more confident in diagnosing the proximal segment. Conclusion: High resolution MTS TONE 3D TOF MRA is an accurate technique for the screening of medium and large vessel intracranial stenoses.

AB - Objective: Factors that restrict 3D TOF MRA are limited resolution, saturation of flow, and degree of background suppression. We evaluated MRA for intracranial stenoses by using a 3D TOF technique that minimizes these factors. Materials and Methods: Twenty-nine patients underwent MRA and intraarterial digital subtraction angiography (DSA). The MRA studies were performed on a 1.5 T Siemens SP 4000 system. Integrated techniques applied to the conventional 3D TOF acquisition included the following: (a) 256 x 256 matrix with a 140 mm FOV and 0.9 mm slice thickness, yielding a 0.54 x 0.54 x 0.9 mm3 voxel; (b) tilted optimized nonsaturating excitation (TONE); and (c) magnetization transfer saturation (MTS). The intraarterial DSA was performed on a Siemens Angiostar system with a 1,024 x 1,024 noninterpolated matrix. The MRAs were reviewed by two neuroradiologists. Two hundred seventy-seven vessels were evaluated for a total of 806 segments. Vessel segments were evaluated with a 5 point scale. Results: The estimated accuracy of MRA for detecting stenosis over all intracranial vessel segments was 0.88 ± 0.03 and 0.89 ± 0.02 for the two readers, respectively. The estimated accuracy ranged from 0.94 ± 0.02 and 0.93 ± 0.02 for detecting internal carotid artery stenosis by the two readers, respectively, to 0.65 ± 0.17 and 0.71 ± 0.15 for detecting distal vertebral artery stenosis. In vessels determined by catheter angiography to be stenosis-free, reader confidence at the proximal versus distal segments was similar for the internal carotid, basilar, and posterior cerebral arteries. However, for the anterior and middle cerebral arteries, one or both readers were more confident in diagnosing the proximal segment. Conclusion: High resolution MTS TONE 3D TOF MRA is an accurate technique for the screening of medium and large vessel intracranial stenoses.

KW - Brain, blood flow

KW - Magnetic resonance imaging angiography (MRA)

KW - Magnetic resonance imaging, techniques

KW - Magnetization transfer

UR - http://www.scopus.com/inward/record.url?scp=0029118567&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029118567&partnerID=8YFLogxK

M3 - Article

C2 - 7560313

AN - SCOPUS:0029118567

VL - 19

SP - 700

EP - 706

JO - Journal of Computer Assisted Tomography

JF - Journal of Computer Assisted Tomography

SN - 0363-8715

IS - 5

ER -