Feasibility of pulmonary artery pressure measurements in infants through aorto-pulmonary shunts using a micromanometer pressure wire

A. D. Everett, G. P. Matherne

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Assessment of pulmonary artery pressure is an essential element in the evaluation of children palliated with surgical aorto-pulmonary shunts prior to definitive surgical repair. We report the ease of use and accuracy of a 0.014 inch micromanometer pressure wire for the measurement of pulmonary artery pressures in children with aorto-pulmonary artery shunts. The study population consisted of 11 infants and children with either a 3.5 mm modified Blalock-Taussig shunt from the subclavian artery to the branch pulmonary artery after stage 1 Norwood repair for hypoplastic left heart syndrome or palliative staged repair for tetralogy of Fallot, or a central shunt for pulmonary atresia or double outlet right ventricle. The unique features of the micromanometer pressure wire allowed rapid access and accurate measurement of pulmonary pressures in all patients studied. We conclude that the micromanometer pressure wire is a unique and accurate alternative device for rapid and safe determinations of pulmonary artery pressures in children with aorto-pulmonary artery shunts.

Original languageEnglish (US)
Pages (from-to)336-337
Number of pages2
JournalPediatric Cardiology
Volume24
Issue number4
DOIs
StatePublished - Jul 2003
Externally publishedYes

Keywords

  • Fontan
  • PTFE shunts
  • Single ventricle

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Feasibility of pulmonary artery pressure measurements in infants through aorto-pulmonary shunts using a micromanometer pressure wire'. Together they form a unique fingerprint.

Cite this