Evaluation of the ΔV 4D CT ventilation calculation method using in vivo xenon CT ventilation data and comparison to other methods

Geoffrey G. Zhang, Kujtim Latifi, Kaifang Du, Joseph M. Reinhardt, Gary E. Christensen, Kai Ding, Vladimir Feygelman, Eduardo G. Moros

Research output: Contribution to journalArticlepeer-review

6 Scopus citations

Abstract

Ventilation distribution calculation using 4D CT has shown promising potential in several clinical applications. This study evaluated the direct geometric ventilation calculation method, namely the ΔV method, with xenon-enhanced CT (XeCT) ventilation data from four sheep, and compared it with two other published meth-ods, the Jacobian and the Hounsfield unit (HU) methods. Spearman correlation coefficient (SCC) and Dice similarity coefficient (DSC) were used for the evaluation and comparison. The average SCC with one standard deviation was 0.44 ± 0.13 with a range between 0.29 and 0.61 between the XeCT and ΔV ventilation distributions. The average DSC value for lower 30% ventilation volumes between the XeCT and ΔV ventilation distributions was 0.55 ± 0.07 with a range between 0.48 and 0.63. Ventilation difference introduced by deformable image registration errors improved with smoothing. In conclusion, ventilation distributions generated using ΔV-4D CT and deformable image registration are in reasonably agreement with the in vivo XeCT measured ventilation distribution.

Original languageEnglish (US)
Pages (from-to)5985
Number of pages1
JournalJournal of Applied Clinical Medical Physics
Volume17
Issue number2
StatePublished - Mar 8 2016

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluation of the ΔV 4D CT ventilation calculation method using in vivo xenon CT ventilation data and comparison to other methods'. Together they form a unique fingerprint.

Cite this