Evaluation of cardiac cone—beam single photon emission computed tomography using observer performance experiments and receiver operating characteristic analysis

Benjamin M.W. Tsui, James A. Terry, Grant T. Gullberg

Research output: Contribution to journalArticlepeer-review

Abstract

RATIONALE AND OBJECTIVES. Single photon emission computed tomography (SPECT) with a cone-beam collimator improves the trade-off between detection efficiency and spatial resolution for cardiac imaging. However, acquisitions using orbits where the focus remains in a plane do not provide sufficient data for exact reconstruction. In the current study the authors evaluate the clinical utility of planar-orbit cone-beam SPECT in detecting a simple myocardial defect. METHODS. Observer performance experiments compared high-resolution cone-beam with same-resolution parallel-hole and fan-beam collimator designs in myocardial defect detection using a computer-simulated cardiac model. The uptake of Thallium-201 in the myocardium and other tissue organs was modeled by a mathematical three-dimensional upper torso phantom from which physically realistic projections were simulated. Eight observers viewed reconstructed transaxial images from the three collimator designs and indicated the certainty with which they detected a Gaussian-shaped defect at a specified location. RESULTS. The area under the receiver operating characteristic curve indicated that the cone-beam design, regardless of slice position, was superior to the fan-beam, which in turn was superior to the parallel-hole design for the specified detection task. conclusions. The observer study demonstrated that reconstruction artifacts resulting from insufficient data sampling do not hinder obtaining improved diagnostic information from planar-orbit cone-beam cardiac SPECT images compared to conventional cardiac SPECT using parallel-hole and fan-beam collimators.

Original languageEnglish (US)
Pages (from-to)1101-1112
Number of pages12
JournalInvestigative radiology
Volume28
Issue number12
DOIs
StatePublished - Dec 1993
Externally publishedYes

Keywords

  • Cardiac
  • Cone-beam
  • Receiver operating characteristic
  • Single photon emission computed tomography

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint Dive into the research topics of 'Evaluation of cardiac cone—beam single photon emission computed tomography using observer performance experiments and receiver operating characteristic analysis'. Together they form a unique fingerprint.