Erratum to ‘2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing’ (Journal of Arrhythmia (2016) 32(1) (1–28) (S1880427615001696) (10.1016/j.joa.2015.12.001))

Bruce L. Wilkoff, Laurent Fauchier, Martin K. Stiles, Carlos A. Morillo, Sana M. Al-Khatib, Jesœs Almendral, Luis Aguinaga, Ronald D. Berger, Alejandro Cuesta, James P. Daubert, Sergio Dubner, Kenneth A. Ellenbogen, N. A. Mark Estes, Guilherme Fenelon, Fermin C. Garcia, Maurizio Gasparini, David E. Haines, Jeff S. Healey, Jodie L. Hurtwitz, Roberto KeeganChristof Kolb, Karl Heinz Kuck, Germanas Marinskis, Martino Martinelli, Mark McGuire, Luis G. Molina, Ken Okumura, Alessandro Proclemer, Andrea M. Russo, Jagmeet P. Singh, Charles D. Swerdlow, Wee Siong Teo, William Uribe, Sami Viskin, Chun Chieh Wang, Shu Zhang

Research output: Contribution to journalComment/debatepeer-review

1 Scopus citations

Abstract

There was an error on the above article. The sentence “Even if a dual-chamber ICD is implanted, dual-chamber discrimination should be programmed only if the atrial lead becomes chronic or if atrial sensing is unreliable” is not correct. It should read: “Even if a dual- chamber ICD is implanted, dual-chamber discrimination should be programmed only if the atrial lead becomes chronic or if atrial sensing is reliable”.

Original languageEnglish (US)
Pages (from-to)441-442
Number of pages2
Journaljournal of arrhythmia
Volume32
Issue number5
DOIs
StatePublished - Oct 1 2016

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Erratum to ‘2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing’ (Journal of Arrhythmia (2016) 32(1) (1–28) (S1880427615001696) (10.1016/j.joa.2015.12.001))'. Together they form a unique fingerprint.