Effects of ultrasound section-thickness on brachytherapy needle tip localization error.

Mohammad Peikari, Thomas Kuiran Chen, Andras Lasso, Tamas Heffter, Gabor Fichtinger

Research output: Chapter in Book/Report/Conference proceedingChapter

2 Scopus citations

Abstract

Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.

Original languageEnglish (US)
Title of host publicationMedical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention
Pages299-306
Number of pages8
Volume14
EditionPt 1
StatePublished - 2011
Externally publishedYes

ASJC Scopus subject areas

  • General Medicine

Fingerprint

Dive into the research topics of 'Effects of ultrasound section-thickness on brachytherapy needle tip localization error.'. Together they form a unique fingerprint.

Cite this