Effects of ultrasound section-thickness on brachytherapy needle tip localization error.

Mohammad Peikari, Thomas Kuiran Chen, Andras Lasso, Tamas Heffter, Gabor Fichtinger

Research output: Chapter in Book/Report/Conference proceedingChapter

Abstract

Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.

Original languageEnglish (US)
Title of host publicationMedical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention
Pages299-306
Number of pages8
Volume14
EditionPt 1
StatePublished - 2011
Externally publishedYes

Fingerprint

Brachytherapy
Needles
Artifacts
Transducers
Calibration

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Peikari, M., Chen, T. K., Lasso, A., Heffter, T., & Fichtinger, G. (2011). Effects of ultrasound section-thickness on brachytherapy needle tip localization error. In Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention (Pt 1 ed., Vol. 14, pp. 299-306)

Effects of ultrasound section-thickness on brachytherapy needle tip localization error. / Peikari, Mohammad; Chen, Thomas Kuiran; Lasso, Andras; Heffter, Tamas; Fichtinger, Gabor.

Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention. Vol. 14 Pt 1. ed. 2011. p. 299-306.

Research output: Chapter in Book/Report/Conference proceedingChapter

Peikari, M, Chen, TK, Lasso, A, Heffter, T & Fichtinger, G 2011, Effects of ultrasound section-thickness on brachytherapy needle tip localization error. in Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention. Pt 1 edn, vol. 14, pp. 299-306.
Peikari M, Chen TK, Lasso A, Heffter T, Fichtinger G. Effects of ultrasound section-thickness on brachytherapy needle tip localization error. In Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention. Pt 1 ed. Vol. 14. 2011. p. 299-306
Peikari, Mohammad ; Chen, Thomas Kuiran ; Lasso, Andras ; Heffter, Tamas ; Fichtinger, Gabor. / Effects of ultrasound section-thickness on brachytherapy needle tip localization error. Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention. Vol. 14 Pt 1. ed. 2011. pp. 299-306
@inbook{2fedac394b7f47279da8782852846443,
title = "Effects of ultrasound section-thickness on brachytherapy needle tip localization error.",
abstract = "Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.",
author = "Mohammad Peikari and Chen, {Thomas Kuiran} and Andras Lasso and Tamas Heffter and Gabor Fichtinger",
year = "2011",
language = "English (US)",
volume = "14",
pages = "299--306",
booktitle = "Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention",
edition = "Pt 1",

}

TY - CHAP

T1 - Effects of ultrasound section-thickness on brachytherapy needle tip localization error.

AU - Peikari, Mohammad

AU - Chen, Thomas Kuiran

AU - Lasso, Andras

AU - Heffter, Tamas

AU - Fichtinger, Gabor

PY - 2011

Y1 - 2011

N2 - Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.

AB - Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.

UR - http://www.scopus.com/inward/record.url?scp=82255185859&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=82255185859&partnerID=8YFLogxK

M3 - Chapter

C2 - 22003630

AN - SCOPUS:82255185859

VL - 14

SP - 299

EP - 306

BT - Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention

ER -