Development of an image-guided robot for small animal research

Peter Kazanzides, Jenghwa Chang, Iulian Iordachita, Jack Li, C. Clifton Ling, Gabor Fichtinger

Research output: Contribution to journalArticle

Abstract

We developed a robot system that can be used for image-guided experimental procedures on small animals, where the goal is to perform physical actions at specific positions identified on a preoperative image. The animal is first placed in a fixture that is compatible with all imaging systems of interest, including PET, SPECT, CT and MRI. After imaging, the fixture is attached and registered to the robot system, where the image-guided intervention is performed. This system has been applied to perform pO2 measurements with physical probes based on tumor hypoxia images obtained in an animal PET scanner. This paper focuses on the design and validation of the robot system. The validation is performed using a phantom and includes a new method for estimating the Fiducial Localization Error (FLE) that is based on the measured Fiducial Distance Error (FDE). The results indicate that the robot system can position the measurement probe at a defined target with a mean error that is less than 0.4 mm.

Original languageEnglish (US)
Pages (from-to)357-365
Number of pages9
JournalComputer Aided Surgery
Volume12
Issue number6
DOIs
StatePublished - Nov 2007

Fingerprint

Animals
Robots
Imaging systems
Magnetic resonance imaging
Tumors
Imaging techniques
Tumor Hypoxia
Single Photon Emission Computed Tomography Computed Tomography

Keywords

  • Animal research
  • Fiducial distance error
  • Fiducial localization error
  • Hypoxia
  • microPET
  • Surgical robot

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Surgery
  • Medicine (miscellaneous)

Cite this

Kazanzides, P., Chang, J., Iordachita, I., Li, J., Ling, C. C., & Fichtinger, G. (2007). Development of an image-guided robot for small animal research. Computer Aided Surgery, 12(6), 357-365. https://doi.org/10.1080/10929080701732538

Development of an image-guided robot for small animal research. / Kazanzides, Peter; Chang, Jenghwa; Iordachita, Iulian; Li, Jack; Ling, C. Clifton; Fichtinger, Gabor.

In: Computer Aided Surgery, Vol. 12, No. 6, 11.2007, p. 357-365.

Research output: Contribution to journalArticle

Kazanzides, P, Chang, J, Iordachita, I, Li, J, Ling, CC & Fichtinger, G 2007, 'Development of an image-guided robot for small animal research', Computer Aided Surgery, vol. 12, no. 6, pp. 357-365. https://doi.org/10.1080/10929080701732538
Kazanzides P, Chang J, Iordachita I, Li J, Ling CC, Fichtinger G. Development of an image-guided robot for small animal research. Computer Aided Surgery. 2007 Nov;12(6):357-365. https://doi.org/10.1080/10929080701732538
Kazanzides, Peter ; Chang, Jenghwa ; Iordachita, Iulian ; Li, Jack ; Ling, C. Clifton ; Fichtinger, Gabor. / Development of an image-guided robot for small animal research. In: Computer Aided Surgery. 2007 ; Vol. 12, No. 6. pp. 357-365.
@article{adf196e1f432449189077f9bb68cbcd6,
title = "Development of an image-guided robot for small animal research",
abstract = "We developed a robot system that can be used for image-guided experimental procedures on small animals, where the goal is to perform physical actions at specific positions identified on a preoperative image. The animal is first placed in a fixture that is compatible with all imaging systems of interest, including PET, SPECT, CT and MRI. After imaging, the fixture is attached and registered to the robot system, where the image-guided intervention is performed. This system has been applied to perform pO2 measurements with physical probes based on tumor hypoxia images obtained in an animal PET scanner. This paper focuses on the design and validation of the robot system. The validation is performed using a phantom and includes a new method for estimating the Fiducial Localization Error (FLE) that is based on the measured Fiducial Distance Error (FDE). The results indicate that the robot system can position the measurement probe at a defined target with a mean error that is less than 0.4 mm.",
keywords = "Animal research, Fiducial distance error, Fiducial localization error, Hypoxia, microPET, Surgical robot",
author = "Peter Kazanzides and Jenghwa Chang and Iulian Iordachita and Jack Li and Ling, {C. Clifton} and Gabor Fichtinger",
year = "2007",
month = "11",
doi = "10.1080/10929080701732538",
language = "English (US)",
volume = "12",
pages = "357--365",
journal = "Computer Assisted Surgery",
issn = "1092-9088",
publisher = "Taylor and Francis Ltd.",
number = "6",

}

TY - JOUR

T1 - Development of an image-guided robot for small animal research

AU - Kazanzides, Peter

AU - Chang, Jenghwa

AU - Iordachita, Iulian

AU - Li, Jack

AU - Ling, C. Clifton

AU - Fichtinger, Gabor

PY - 2007/11

Y1 - 2007/11

N2 - We developed a robot system that can be used for image-guided experimental procedures on small animals, where the goal is to perform physical actions at specific positions identified on a preoperative image. The animal is first placed in a fixture that is compatible with all imaging systems of interest, including PET, SPECT, CT and MRI. After imaging, the fixture is attached and registered to the robot system, where the image-guided intervention is performed. This system has been applied to perform pO2 measurements with physical probes based on tumor hypoxia images obtained in an animal PET scanner. This paper focuses on the design and validation of the robot system. The validation is performed using a phantom and includes a new method for estimating the Fiducial Localization Error (FLE) that is based on the measured Fiducial Distance Error (FDE). The results indicate that the robot system can position the measurement probe at a defined target with a mean error that is less than 0.4 mm.

AB - We developed a robot system that can be used for image-guided experimental procedures on small animals, where the goal is to perform physical actions at specific positions identified on a preoperative image. The animal is first placed in a fixture that is compatible with all imaging systems of interest, including PET, SPECT, CT and MRI. After imaging, the fixture is attached and registered to the robot system, where the image-guided intervention is performed. This system has been applied to perform pO2 measurements with physical probes based on tumor hypoxia images obtained in an animal PET scanner. This paper focuses on the design and validation of the robot system. The validation is performed using a phantom and includes a new method for estimating the Fiducial Localization Error (FLE) that is based on the measured Fiducial Distance Error (FDE). The results indicate that the robot system can position the measurement probe at a defined target with a mean error that is less than 0.4 mm.

KW - Animal research

KW - Fiducial distance error

KW - Fiducial localization error

KW - Hypoxia

KW - microPET

KW - Surgical robot

UR - http://www.scopus.com/inward/record.url?scp=36849005580&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=36849005580&partnerID=8YFLogxK

U2 - 10.1080/10929080701732538

DO - 10.1080/10929080701732538

M3 - Article

VL - 12

SP - 357

EP - 365

JO - Computer Assisted Surgery

JF - Computer Assisted Surgery

SN - 1092-9088

IS - 6

ER -