N2 - Previously, a static and adjustable image overlay systems were proposed for aiding needle interventions. The system was either fixed to a scanner or mounted over a large articulated counterbalanced arm. Certain drawbacks associated with these systems limited the clinical translation. In order to minimize these limitations, we present the mobile image overlay system with the objective of reduced system weight, smaller dimension, and increased tracking accuracy. The design study includes optimal workspace definition, selection of display device, mirror, and laser source. The laser plane alignment, phantom design, image overlay plane calibration, and system accuracy validation methods are discussed. The virtual image is generated by a tablet device and projected into the patient by using a beamsplitter mirror. The viewbox weight (1.0kg) was reduced by 8.2 times and image overlay plane tracking precision (0.21mm, STD=0.05) was improved by 5 times compared to previous system. The automatic self-calibration of the image overlay plane was achieved in two simple steps and can be done away from patient table. The fiducial registration error of the physical phantom to scanned image volume registration was 1.35mm (STD=0.11). The reduced system weight and increased accuracy of optical tracking should enable the system to be hand held by the physician and explore the image volume over the patient for needle interventions.
AB - Previously, a static and adjustable image overlay systems were proposed for aiding needle interventions. The system was either fixed to a scanner or mounted over a large articulated counterbalanced arm. Certain drawbacks associated with these systems limited the clinical translation. In order to minimize these limitations, we present the mobile image overlay system with the objective of reduced system weight, smaller dimension, and increased tracking accuracy. The design study includes optimal workspace definition, selection of display device, mirror, and laser source. The laser plane alignment, phantom design, image overlay plane calibration, and system accuracy validation methods are discussed. The virtual image is generated by a tablet device and projected into the patient by using a beamsplitter mirror. The viewbox weight (1.0kg) was reduced by 8.2 times and image overlay plane tracking precision (0.21mm, STD=0.05) was improved by 5 times compared to previous system. The automatic self-calibration of the image overlay plane was achieved in two simple steps and can be done away from patient table. The fiducial registration error of the physical phantom to scanned image volume registration was 1.35mm (STD=0.11). The reduced system weight and increased accuracy of optical tracking should enable the system to be hand held by the physician and explore the image volume over the patient for needle interventions.
UR - http://www.scopus.com/inward/record.url?scp=84929493367&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84929493367&partnerID=8YFLogxK
U2 - 10.1109/EMBC.2014.6945035
DO - 10.1109/EMBC.2014.6945035
M3 - Conference contribution
C2 - 25571403
AN - SCOPUS:84929493367
T3 - 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014
SP - 6159
EP - 6162
BT - 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014
Y2 - 26 August 2014 through 30 August 2014
ER -