Characterization of a fluoroscopic imaging system for kV and MV radiography

D. G. Drake, D. A. Jaffray, John Wong

Research output: Contribution to journalArticle

Abstract

An on-line kilovoltage (kV) imaging system has been implemented on a medical linear accelerator to verify radiotherapy field placement. A kV x-ray tube is mounted on the accelerator at 90°to the megavoltage (MV) source and shares the same isocenter. Nearly identical CCD-based fluoroscopic imagers are mounted opposite the two x-ray sources. These systems are being used in a clinical study of patient setup error that examines the advantage of kV imaging for on-line localization. In the investigation reported here, the imaging performance of the kV and MV systems are characterized to provide support to the conclusions of the studies of setup error. A spatial- frequency-dependent linear systems model is used to predict the detective quantum efficiencies (DQEs) of the two systems. Each is divided into a series of gain and spreading stages. The parameters of each stage are either measured or obtained from the literature. The model predicts the system gain to within 7% of the measured gain for the MV system and to within 10% for the kV system. The systems' noise power spectra (NPSs) and modulation transfer functions (MTFs) are measured to construct the measured DQEs. X-ray fluences are calculated using modeled polyenergetic spectra. Measured DQEs agree well with those predicted by the model. The model reveals that the MV system is well optimized, and is x-ray quantum noise limited at low spatial frequencies. The kV system is suboptimal, but for purposes of patient positioning yields images superior to those produced by the MV system. This is attributed to the kV system's higher DQE and to the inherently higher contrasts present at kV energies. (C) 2000 American Association of Physicists in Medicine.

Original languageEnglish (US)
Pages (from-to)898-905
Number of pages8
JournalMedical Physics
Volume27
Issue number5
DOIs
StatePublished - May 2000
Externally publishedYes

Fingerprint

Radiography
X-Rays
Patient Positioning
Particle Accelerators
Noise
Linear Models
Radiotherapy

Keywords

  • Detective quantum efficiency
  • Kilovoltage
  • Megavoltage
  • Portal imaging

ASJC Scopus subject areas

  • Biophysics

Cite this

Characterization of a fluoroscopic imaging system for kV and MV radiography. / Drake, D. G.; Jaffray, D. A.; Wong, John.

In: Medical Physics, Vol. 27, No. 5, 05.2000, p. 898-905.

Research output: Contribution to journalArticle

Drake, D. G. ; Jaffray, D. A. ; Wong, John. / Characterization of a fluoroscopic imaging system for kV and MV radiography. In: Medical Physics. 2000 ; Vol. 27, No. 5. pp. 898-905.
@article{86668359d4014c4ba941b4de54293c20,
title = "Characterization of a fluoroscopic imaging system for kV and MV radiography",
abstract = "An on-line kilovoltage (kV) imaging system has been implemented on a medical linear accelerator to verify radiotherapy field placement. A kV x-ray tube is mounted on the accelerator at 90°to the megavoltage (MV) source and shares the same isocenter. Nearly identical CCD-based fluoroscopic imagers are mounted opposite the two x-ray sources. These systems are being used in a clinical study of patient setup error that examines the advantage of kV imaging for on-line localization. In the investigation reported here, the imaging performance of the kV and MV systems are characterized to provide support to the conclusions of the studies of setup error. A spatial- frequency-dependent linear systems model is used to predict the detective quantum efficiencies (DQEs) of the two systems. Each is divided into a series of gain and spreading stages. The parameters of each stage are either measured or obtained from the literature. The model predicts the system gain to within 7{\%} of the measured gain for the MV system and to within 10{\%} for the kV system. The systems' noise power spectra (NPSs) and modulation transfer functions (MTFs) are measured to construct the measured DQEs. X-ray fluences are calculated using modeled polyenergetic spectra. Measured DQEs agree well with those predicted by the model. The model reveals that the MV system is well optimized, and is x-ray quantum noise limited at low spatial frequencies. The kV system is suboptimal, but for purposes of patient positioning yields images superior to those produced by the MV system. This is attributed to the kV system's higher DQE and to the inherently higher contrasts present at kV energies. (C) 2000 American Association of Physicists in Medicine.",
keywords = "Detective quantum efficiency, Kilovoltage, Megavoltage, Portal imaging",
author = "Drake, {D. G.} and Jaffray, {D. A.} and John Wong",
year = "2000",
month = "5",
doi = "10.1118/1.598955",
language = "English (US)",
volume = "27",
pages = "898--905",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - Characterization of a fluoroscopic imaging system for kV and MV radiography

AU - Drake, D. G.

AU - Jaffray, D. A.

AU - Wong, John

PY - 2000/5

Y1 - 2000/5

N2 - An on-line kilovoltage (kV) imaging system has been implemented on a medical linear accelerator to verify radiotherapy field placement. A kV x-ray tube is mounted on the accelerator at 90°to the megavoltage (MV) source and shares the same isocenter. Nearly identical CCD-based fluoroscopic imagers are mounted opposite the two x-ray sources. These systems are being used in a clinical study of patient setup error that examines the advantage of kV imaging for on-line localization. In the investigation reported here, the imaging performance of the kV and MV systems are characterized to provide support to the conclusions of the studies of setup error. A spatial- frequency-dependent linear systems model is used to predict the detective quantum efficiencies (DQEs) of the two systems. Each is divided into a series of gain and spreading stages. The parameters of each stage are either measured or obtained from the literature. The model predicts the system gain to within 7% of the measured gain for the MV system and to within 10% for the kV system. The systems' noise power spectra (NPSs) and modulation transfer functions (MTFs) are measured to construct the measured DQEs. X-ray fluences are calculated using modeled polyenergetic spectra. Measured DQEs agree well with those predicted by the model. The model reveals that the MV system is well optimized, and is x-ray quantum noise limited at low spatial frequencies. The kV system is suboptimal, but for purposes of patient positioning yields images superior to those produced by the MV system. This is attributed to the kV system's higher DQE and to the inherently higher contrasts present at kV energies. (C) 2000 American Association of Physicists in Medicine.

AB - An on-line kilovoltage (kV) imaging system has been implemented on a medical linear accelerator to verify radiotherapy field placement. A kV x-ray tube is mounted on the accelerator at 90°to the megavoltage (MV) source and shares the same isocenter. Nearly identical CCD-based fluoroscopic imagers are mounted opposite the two x-ray sources. These systems are being used in a clinical study of patient setup error that examines the advantage of kV imaging for on-line localization. In the investigation reported here, the imaging performance of the kV and MV systems are characterized to provide support to the conclusions of the studies of setup error. A spatial- frequency-dependent linear systems model is used to predict the detective quantum efficiencies (DQEs) of the two systems. Each is divided into a series of gain and spreading stages. The parameters of each stage are either measured or obtained from the literature. The model predicts the system gain to within 7% of the measured gain for the MV system and to within 10% for the kV system. The systems' noise power spectra (NPSs) and modulation transfer functions (MTFs) are measured to construct the measured DQEs. X-ray fluences are calculated using modeled polyenergetic spectra. Measured DQEs agree well with those predicted by the model. The model reveals that the MV system is well optimized, and is x-ray quantum noise limited at low spatial frequencies. The kV system is suboptimal, but for purposes of patient positioning yields images superior to those produced by the MV system. This is attributed to the kV system's higher DQE and to the inherently higher contrasts present at kV energies. (C) 2000 American Association of Physicists in Medicine.

KW - Detective quantum efficiency

KW - Kilovoltage

KW - Megavoltage

KW - Portal imaging

UR - http://www.scopus.com/inward/record.url?scp=0034121931&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034121931&partnerID=8YFLogxK

U2 - 10.1118/1.598955

DO - 10.1118/1.598955

M3 - Article

C2 - 10841392

AN - SCOPUS:0034121931

VL - 27

SP - 898

EP - 905

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -