Cephalad lead migration following spinal cord stimulation implantation

Research output: Contribution to journalArticle

Abstract

Lead migration (LM) is the most common complication after spinal cord stimulation (SCS). Although multiple reports of caudad LM have been described, cephalad LM has not been reported. Here we describe a case in which a stimulator lead migrates in the cephalad direction. A 60-year-old male with failed back surgery syndrome underwent SCS lead implantation via a dual lead approach to the top of vertebral body (VB) T9. A standard strain relief loop was used for each lead in the paramedian pocket. Postoperative testing revealed 100% paresthesia coverage of the painful areas. For the first 4 days, the patient continued to have excellent coverage; however, by the seventh day, the paresthesias ascended to above the nipple line. At the 2-week follow-up, cephalad migration of the left lead to the top of VB T1 was confirmed on fluoroscopy. The patient underwent successful lead revision in which a single paramedian incision technique was used to place extra sutures and a "figure-of-eight" strain relief loop. We provide the first case report of significant cephalad LM following SCS lead implantation. This migration can occur despite the use of current standard anchoring techniques. Additional investigation into the mechanism of such LM and lead-securing techniques is warranted.

Original languageEnglish (US)
JournalPain Physician
Volume15
Issue number1
StatePublished - Jan 2012

Fingerprint

Spinal Cord Stimulation
Paresthesia
Lead
Failed Back Surgery Syndrome
Nipples
Fluoroscopy
Sutures

Keywords

  • Lead migration
  • Lead revision
  • Spinal cord stimulation
  • Strain relief loop

ASJC Scopus subject areas

  • Anesthesiology and Pain Medicine

Cite this

Cephalad lead migration following spinal cord stimulation implantation. / Mcgreevy, Kai; Williams, Kayode; Christo, Paul.

In: Pain Physician, Vol. 15, No. 1, 01.2012.

Research output: Contribution to journalArticle

@article{9b99a163ce3a4a7da50237e5fa90d434,
title = "Cephalad lead migration following spinal cord stimulation implantation",
abstract = "Lead migration (LM) is the most common complication after spinal cord stimulation (SCS). Although multiple reports of caudad LM have been described, cephalad LM has not been reported. Here we describe a case in which a stimulator lead migrates in the cephalad direction. A 60-year-old male with failed back surgery syndrome underwent SCS lead implantation via a dual lead approach to the top of vertebral body (VB) T9. A standard strain relief loop was used for each lead in the paramedian pocket. Postoperative testing revealed 100{\%} paresthesia coverage of the painful areas. For the first 4 days, the patient continued to have excellent coverage; however, by the seventh day, the paresthesias ascended to above the nipple line. At the 2-week follow-up, cephalad migration of the left lead to the top of VB T1 was confirmed on fluoroscopy. The patient underwent successful lead revision in which a single paramedian incision technique was used to place extra sutures and a {"}figure-of-eight{"} strain relief loop. We provide the first case report of significant cephalad LM following SCS lead implantation. This migration can occur despite the use of current standard anchoring techniques. Additional investigation into the mechanism of such LM and lead-securing techniques is warranted.",
keywords = "Lead migration, Lead revision, Spinal cord stimulation, Strain relief loop",
author = "Kai Mcgreevy and Kayode Williams and Paul Christo",
year = "2012",
month = "1",
language = "English (US)",
volume = "15",
journal = "Pain Physician",
issn = "1533-3159",
publisher = "Association of Pain Management Anesthesiologists",
number = "1",

}

TY - JOUR

T1 - Cephalad lead migration following spinal cord stimulation implantation

AU - Mcgreevy, Kai

AU - Williams, Kayode

AU - Christo, Paul

PY - 2012/1

Y1 - 2012/1

N2 - Lead migration (LM) is the most common complication after spinal cord stimulation (SCS). Although multiple reports of caudad LM have been described, cephalad LM has not been reported. Here we describe a case in which a stimulator lead migrates in the cephalad direction. A 60-year-old male with failed back surgery syndrome underwent SCS lead implantation via a dual lead approach to the top of vertebral body (VB) T9. A standard strain relief loop was used for each lead in the paramedian pocket. Postoperative testing revealed 100% paresthesia coverage of the painful areas. For the first 4 days, the patient continued to have excellent coverage; however, by the seventh day, the paresthesias ascended to above the nipple line. At the 2-week follow-up, cephalad migration of the left lead to the top of VB T1 was confirmed on fluoroscopy. The patient underwent successful lead revision in which a single paramedian incision technique was used to place extra sutures and a "figure-of-eight" strain relief loop. We provide the first case report of significant cephalad LM following SCS lead implantation. This migration can occur despite the use of current standard anchoring techniques. Additional investigation into the mechanism of such LM and lead-securing techniques is warranted.

AB - Lead migration (LM) is the most common complication after spinal cord stimulation (SCS). Although multiple reports of caudad LM have been described, cephalad LM has not been reported. Here we describe a case in which a stimulator lead migrates in the cephalad direction. A 60-year-old male with failed back surgery syndrome underwent SCS lead implantation via a dual lead approach to the top of vertebral body (VB) T9. A standard strain relief loop was used for each lead in the paramedian pocket. Postoperative testing revealed 100% paresthesia coverage of the painful areas. For the first 4 days, the patient continued to have excellent coverage; however, by the seventh day, the paresthesias ascended to above the nipple line. At the 2-week follow-up, cephalad migration of the left lead to the top of VB T1 was confirmed on fluoroscopy. The patient underwent successful lead revision in which a single paramedian incision technique was used to place extra sutures and a "figure-of-eight" strain relief loop. We provide the first case report of significant cephalad LM following SCS lead implantation. This migration can occur despite the use of current standard anchoring techniques. Additional investigation into the mechanism of such LM and lead-securing techniques is warranted.

KW - Lead migration

KW - Lead revision

KW - Spinal cord stimulation

KW - Strain relief loop

UR - http://www.scopus.com/inward/record.url?scp=84856426805&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84856426805&partnerID=8YFLogxK

M3 - Article

VL - 15

JO - Pain Physician

JF - Pain Physician

SN - 1533-3159

IS - 1

ER -