C-arm tracking and reconstruction without an external tracker.

Ameet Jain, Gabor Fichtinger

Research output: Chapter in Book/Report/Conference proceedingChapter

28 Scopus citations

Abstract

For quantitative C-arm fluoroscopy, we have developed a unified mathematical framework to tackle the issues of intra-operative calibration, pose estimation, correspondence and reconstruction, without the use of optical/electromagnetic trackers or precision-made fiducial fixtures. Our method uses randomly distributed unknown points in the imaging volume, either naturally present or induced by randomly sticking beads or other simple markers in the image pace. After these points are segmented, a high dimensional non-linear optimization computes all unknown parameters for calibration, C-arm pose, correspondence and reconstruction. Preliminary phantom experiments indicate an average C-arm tracking accuracy of 0.9 degrees and a 3D reconstruction error of 0.8 mm, with an 80 region of convergence for both the AP and lateral axes. The method appears to be sufficiently accurate for many clinical applications, and appealing since it works without any external instrumentation and does not interfere with the workspace.

Original languageEnglish (US)
Title of host publicationMedical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention
Pages494-502
Number of pages9
Volume9
EditionPt 1
StatePublished - 2006

ASJC Scopus subject areas

  • General Medicine

Fingerprint

Dive into the research topics of 'C-arm tracking and reconstruction without an external tracker.'. Together they form a unique fingerprint.

Cite this