TY - GEN
T1 - C-arm calibration - Is it really necessary?
AU - Jain, Ameet
AU - Kon, Ryan
AU - Zhou, Yu
AU - Fichtinger, Gabor
PY - 2005
Y1 - 2005
N2 - C-arm fluoroscopy is modelled as a perspective projection, the parameters of which are estimated through a calibration procedure. It has been universally accepted that precise intra-procedural calibration is a prerequisite for accurate quantitative C-arm fluoroscopy guidance. Calibration, however, significantly adds to system complexity, which is a major impediment to clinical practice. We challenge the status quo by questioning the assumption that precise intra-procedural calibration is really necessary. We derived theoretical bounds for the sensitivity of 3D measurements to mis-calibration. Experimental results corroborated the theory in that mis-calibration in the focal spot by as much as 50 mm still allows for tracking with an accuracy of 0.5 mm in translation and 0.65° in rotation, and such mis-calibration does not impose any additional error on the reconstruction of small objects.
AB - C-arm fluoroscopy is modelled as a perspective projection, the parameters of which are estimated through a calibration procedure. It has been universally accepted that precise intra-procedural calibration is a prerequisite for accurate quantitative C-arm fluoroscopy guidance. Calibration, however, significantly adds to system complexity, which is a major impediment to clinical practice. We challenge the status quo by questioning the assumption that precise intra-procedural calibration is really necessary. We derived theoretical bounds for the sensitivity of 3D measurements to mis-calibration. Experimental results corroborated the theory in that mis-calibration in the focal spot by as much as 50 mm still allows for tracking with an accuracy of 0.5 mm in translation and 0.65° in rotation, and such mis-calibration does not impose any additional error on the reconstruction of small objects.
UR - http://www.scopus.com/inward/record.url?scp=33744790938&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33744790938&partnerID=8YFLogxK
U2 - 10.1007/11566465_79
DO - 10.1007/11566465_79
M3 - Conference contribution
AN - SCOPUS:33744790938
SN - 3540293272
SN - 9783540293279
VL - 3749 LNCS
T3 - Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)
SP - 639
EP - 646
BT - Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)
T2 - 8th International Conference on Medical Image Computing and Computer-Assisted Intervention - MICCAI 2005
Y2 - 26 October 2005 through 29 October 2005
ER -