T1 - Biomechanical comparison of three methods of sacral fracture fixation in osteoporotic bone
AU - Mears, Simon C.
AU - Sutter, Edward G.
AU - Wall, Simon J.
AU - Rose, David M.
AU - Belkoff, Stephen M.
PY - 2010/5/1
Y1 - 2010/5/1
N2 - Study Design.: Biomechanical cadaveric bench study. Objective.: To measure motion at the fracture site in an osteoporotic cadaveric sacral insufficiency fracture model before and after fracture creation, after fixation (via 1 of 3 fixation techniques), and after cyclic loading and to compare those values with motion of the intact pelvis. Summary of Background Data.: Sacral insufficiency fractures occur frequently in the elderly and pose treatment challenges. Screw fixation and sacroplasty have been proposed as possible treatments. There is little information about the stabilization provided by these treatments. Methods.: We potted 18 osteoporotic cadaveric pelves, mounted them on a materials testing machine, measured sacroiliac (SI) joint motion with a vertical load applied to the lumbar spine, and created simulated sacral insufficiency fractures. Then, we measured fracture site motion under load and repaired the fracture using 1 of 3 techniques: a unilateral SI screw, a bilateral SI screw, or sacroplasty. A vertical compressive load (10-350 N) was applied cyclically at 0.5 Hz to the lumbar spine of the repaired specimens for 5000 cycles. Kinematic analysis was conducted prefracture, postfracture, postrepair, and after cyclic loading. Results.: Postfracture, there was a significant increase in motion relative to the intact SI joint. After fixation, the average motion in all 3 groups was similar to that of the intact pelvis. After cyclical loading, motion increased in all groups. No significant differences were found between treatments. Conclusion.: All 3 fixation methods resulted acutely in motion similar to that of the intact pelvis. Although motion increased as a function of cyclical loading, no significant differences were found between fixation methods. All 3 repair methods reduced fracture site motion, but clinical studies are needed to determine if each method relieves pain and provides sufficient fixation for fracture healing.
AB - Study Design.: Biomechanical cadaveric bench study. Objective.: To measure motion at the fracture site in an osteoporotic cadaveric sacral insufficiency fracture model before and after fracture creation, after fixation (via 1 of 3 fixation techniques), and after cyclic loading and to compare those values with motion of the intact pelvis. Summary of Background Data.: Sacral insufficiency fractures occur frequently in the elderly and pose treatment challenges. Screw fixation and sacroplasty have been proposed as possible treatments. There is little information about the stabilization provided by these treatments. Methods.: We potted 18 osteoporotic cadaveric pelves, mounted them on a materials testing machine, measured sacroiliac (SI) joint motion with a vertical load applied to the lumbar spine, and created simulated sacral insufficiency fractures. Then, we measured fracture site motion under load and repaired the fracture using 1 of 3 techniques: a unilateral SI screw, a bilateral SI screw, or sacroplasty. A vertical compressive load (10-350 N) was applied cyclically at 0.5 Hz to the lumbar spine of the repaired specimens for 5000 cycles. Kinematic analysis was conducted prefracture, postfracture, postrepair, and after cyclic loading. Results.: Postfracture, there was a significant increase in motion relative to the intact SI joint. After fixation, the average motion in all 3 groups was similar to that of the intact pelvis. After cyclical loading, motion increased in all groups. No significant differences were found between treatments. Conclusion.: All 3 fixation methods resulted acutely in motion similar to that of the intact pelvis. Although motion increased as a function of cyclical loading, no significant differences were found between fixation methods. All 3 repair methods reduced fracture site motion, but clinical studies are needed to determine if each method relieves pain and provides sufficient fixation for fracture healing.
KW - Insufficiency fracture
KW - Mechanical testing
KW - Osteoporosis
KW - Sacrum
UR - http://www.scopus.com/inward/record.url?scp=77952010760&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77952010760&partnerID=8YFLogxK
U2 - 10.1097/BRS.0b013e3181cb4fcd
DO - 10.1097/BRS.0b013e3181cb4fcd
M3 - Article
C2 - 20393387
AN - SCOPUS:77952010760
SN - 0362-2436
VL - 35
SP - E392-E395
JO - Spine
JF - Spine
IS - 10
ER -