Automated detection of foveal fixation by use of retinal birefringence scanning

David G. Hunter, Saurabh N. Patel, David L. Guyton

Research output: Contribution to journalArticle

Abstract

Foveal fixation was monitored in normal subjects remotely and continuously by use of a noninvasive retinal scan. Polarized infrared light was imaged onto the retina and scanned in a 3° annulus at 44 Hz. Reflections were analyzed by differential polarization detection. In all 32 eyes studied, the detected signal was predominantly 88 Hz during central fixation (within ±1°) and 44 Hz during paracentral fixation. Phase shift at 44 Hz correlated with the direction of eye displacement. Potential applications of this technique include screening for eye disease, eye position monitoring during clinical procedures, and use of eye fixation to operate devices.

Original languageEnglish (US)
Pages (from-to)1273-1279
Number of pages7
JournalApplied Optics
Volume38
Issue number7
DOIs
StatePublished - Jan 1 1999

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Engineering (miscellaneous)
  • Electrical and Electronic Engineering

Fingerprint Dive into the research topics of 'Automated detection of foveal fixation by use of retinal birefringence scanning'. Together they form a unique fingerprint.

  • Cite this