Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT

Eric Frey, Karen L. Gilland, Benjamin Tsui

Research output: Contribution to journalArticle

Abstract

In this paper, we apply the channelized Hotelling observer (CHO) using a defect detection task to the optimization and evaluation of three-dimensional iterative reconstruction-based compensation methods for myocardial perfusion single-photon emission computed tomography (SPECT). We used a population of 24 mathematical cardiac-torso phantoms that realistically model the activity and attenuation distribution in three classes of patients: females, and males with flat diaphragms and raised diaphragms. Projection data were generated and subsequently reconstructed using methods based on the ordered subsets-expectation maximization (OSEM) algorithm. The methods evaluated included compensation for attenuation, detector response blurring, and scatter in various combinations. We applied the CHO to optimize the number of iterations for OSEM and the cutoff frequency and order of a three-dimensional postreconstruction Butterworth filter. Using the optimal parameters, we then compared the compensation methods. The index of comparison in these studies was the area under the receiver operating characteristics curve (AUC) for the CHO. We found that attenuation compensation with either detector response or scatter compensation gave statistically significant increases in the AUC compared to attenuation compensation alone. The greatest increase in the AUC occurred when all three compensations were applied. These results indicate that compensation for detector response and scatter, in addition to attenuation compensation, will improve defect detectability in myocardial SPECT images.

Original languageEnglish (US)
Pages (from-to)1040-1050
Number of pages11
JournalIEEE Transactions on Medical Imaging
Volume21
Issue number9
DOIs
StatePublished - Sep 2002

Fingerprint

Single photon emission computed tomography
Single-Photon Emission-Computed Tomography
Image quality
Perfusion
Area Under Curve
Diaphragm
Torso
ROC Curve
Diaphragms
Detectors
Set theory
Butterworth filters
Compensation and Redress
Cutoff frequency
Population

Keywords

  • Attenuation compensation
  • Mathematical observers
  • Scatter compensation
  • SPECT

ASJC Scopus subject areas

  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Electrical and Electronic Engineering
  • Computer Science Applications
  • Computational Theory and Mathematics

Cite this

@article{cb0b9295e0b54789a150992e780bb4f0,
title = "Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT",
abstract = "In this paper, we apply the channelized Hotelling observer (CHO) using a defect detection task to the optimization and evaluation of three-dimensional iterative reconstruction-based compensation methods for myocardial perfusion single-photon emission computed tomography (SPECT). We used a population of 24 mathematical cardiac-torso phantoms that realistically model the activity and attenuation distribution in three classes of patients: females, and males with flat diaphragms and raised diaphragms. Projection data were generated and subsequently reconstructed using methods based on the ordered subsets-expectation maximization (OSEM) algorithm. The methods evaluated included compensation for attenuation, detector response blurring, and scatter in various combinations. We applied the CHO to optimize the number of iterations for OSEM and the cutoff frequency and order of a three-dimensional postreconstruction Butterworth filter. Using the optimal parameters, we then compared the compensation methods. The index of comparison in these studies was the area under the receiver operating characteristics curve (AUC) for the CHO. We found that attenuation compensation with either detector response or scatter compensation gave statistically significant increases in the AUC compared to attenuation compensation alone. The greatest increase in the AUC occurred when all three compensations were applied. These results indicate that compensation for detector response and scatter, in addition to attenuation compensation, will improve defect detectability in myocardial SPECT images.",
keywords = "Attenuation compensation, Mathematical observers, Scatter compensation, SPECT",
author = "Eric Frey and Gilland, {Karen L.} and Benjamin Tsui",
year = "2002",
month = "9",
doi = "10.1109/TMI.2002.804437",
language = "English (US)",
volume = "21",
pages = "1040--1050",
journal = "IEEE Transactions on Medical Imaging",
issn = "0278-0062",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "9",

}

TY - JOUR

T1 - Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT

AU - Frey, Eric

AU - Gilland, Karen L.

AU - Tsui, Benjamin

PY - 2002/9

Y1 - 2002/9

N2 - In this paper, we apply the channelized Hotelling observer (CHO) using a defect detection task to the optimization and evaluation of three-dimensional iterative reconstruction-based compensation methods for myocardial perfusion single-photon emission computed tomography (SPECT). We used a population of 24 mathematical cardiac-torso phantoms that realistically model the activity and attenuation distribution in three classes of patients: females, and males with flat diaphragms and raised diaphragms. Projection data were generated and subsequently reconstructed using methods based on the ordered subsets-expectation maximization (OSEM) algorithm. The methods evaluated included compensation for attenuation, detector response blurring, and scatter in various combinations. We applied the CHO to optimize the number of iterations for OSEM and the cutoff frequency and order of a three-dimensional postreconstruction Butterworth filter. Using the optimal parameters, we then compared the compensation methods. The index of comparison in these studies was the area under the receiver operating characteristics curve (AUC) for the CHO. We found that attenuation compensation with either detector response or scatter compensation gave statistically significant increases in the AUC compared to attenuation compensation alone. The greatest increase in the AUC occurred when all three compensations were applied. These results indicate that compensation for detector response and scatter, in addition to attenuation compensation, will improve defect detectability in myocardial SPECT images.

AB - In this paper, we apply the channelized Hotelling observer (CHO) using a defect detection task to the optimization and evaluation of three-dimensional iterative reconstruction-based compensation methods for myocardial perfusion single-photon emission computed tomography (SPECT). We used a population of 24 mathematical cardiac-torso phantoms that realistically model the activity and attenuation distribution in three classes of patients: females, and males with flat diaphragms and raised diaphragms. Projection data were generated and subsequently reconstructed using methods based on the ordered subsets-expectation maximization (OSEM) algorithm. The methods evaluated included compensation for attenuation, detector response blurring, and scatter in various combinations. We applied the CHO to optimize the number of iterations for OSEM and the cutoff frequency and order of a three-dimensional postreconstruction Butterworth filter. Using the optimal parameters, we then compared the compensation methods. The index of comparison in these studies was the area under the receiver operating characteristics curve (AUC) for the CHO. We found that attenuation compensation with either detector response or scatter compensation gave statistically significant increases in the AUC compared to attenuation compensation alone. The greatest increase in the AUC occurred when all three compensations were applied. These results indicate that compensation for detector response and scatter, in addition to attenuation compensation, will improve defect detectability in myocardial SPECT images.

KW - Attenuation compensation

KW - Mathematical observers

KW - Scatter compensation

KW - SPECT

UR - http://www.scopus.com/inward/record.url?scp=0036770471&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036770471&partnerID=8YFLogxK

U2 - 10.1109/TMI.2002.804437

DO - 10.1109/TMI.2002.804437

M3 - Article

C2 - 12564872

AN - SCOPUS:0036770471

VL - 21

SP - 1040

EP - 1050

JO - IEEE Transactions on Medical Imaging

JF - IEEE Transactions on Medical Imaging

SN - 0278-0062

IS - 9

ER -