TY - JOUR
T1 - An assessment of Stat Tek
T2 - A reagent-strip method for glucose determination
AU - Chan, Daniel W.
AU - Derkac, Michael
PY - 1980/2
Y1 - 1980/2
N2 - Stat Tek, a reagent - strip method for glucose determination, has been evaluated. Aqueous solution of glucose standard is not recommended. The linearity of serum glucose is between 630 to 3,450 mg/l. The within run precision is between 0.7% to 3.0%. The day-to-day precision is between 2.8% to 5.6%. The recovery of glucose added to serum is between 95 to 101%. The correlation of glucose by Stat-Tek and SMA 6/60 show small proportional and constant errors. Stat-Tek is simple and easy to use; however, proper quality control and training of the user are essential for the correct measurement of glucose.
AB - Stat Tek, a reagent - strip method for glucose determination, has been evaluated. Aqueous solution of glucose standard is not recommended. The linearity of serum glucose is between 630 to 3,450 mg/l. The within run precision is between 0.7% to 3.0%. The day-to-day precision is between 2.8% to 5.6%. The recovery of glucose added to serum is between 95 to 101%. The correlation of glucose by Stat-Tek and SMA 6/60 show small proportional and constant errors. Stat-Tek is simple and easy to use; however, proper quality control and training of the user are essential for the correct measurement of glucose.
UR - http://www.scopus.com/inward/record.url?scp=0019142249&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0019142249&partnerID=8YFLogxK
U2 - 10.1016/S0009-9120(80)90642-6
DO - 10.1016/S0009-9120(80)90642-6
M3 - Article
C2 - 7363454
AN - SCOPUS:0019142249
SN - 0009-9120
VL - 13
SP - 60
EP - 61
JO - Clinical Biochemistry
JF - Clinical Biochemistry
IS - 1
ER -