An analytical model for the design of RF resonators for MR body imaging

T. K F Foo, C. E. Hayes, Y. W. Kang

Research output: Contribution to journalArticle

Abstract

A closed form, analytical solution describing the RF fields generated by an RF body coil resonator for MR imaging at 1.5 and 4.0 T is presented. This solution extends the results of earlier studies of RF penetration in the body by explicitly including the RF coil, the RF shield, and the field variation along the z axis for high-pass birdcage coils. A salient feature of this treatment is the calculation of the axial propagation constant, k(z), which determines the z dependence of the RF field. We have determined the relative power deposition in the body, the B̄1 field homogeneity, and coil losses, which are functions of the coil-to-shield separation and body size. The relative power deposition in the body has been calculated to vary as the 1.58 power of the body radius. The calculations have also predicted that the field homogeneity in the z direction exhibits greater degradation at higher frequencies in a high-pass coil than in a low-pass coil. The model predicts an increase in coil losses by a factor of 2.8 as the coil-to-shield separation is reduced from 5 to 2 cm in a standard body resonator. Although the results for only a homogeneous cylindrical object or body are presented, the theory can be extended to a multilayered heterogeneous object of varying permittivity and conductivity.

Original languageEnglish (US)
Pages (from-to)165-177
Number of pages13
JournalMagnetic Resonance in Medicine
Volume21
Issue number2
DOIs
StatePublished - 1991
Externally publishedYes

Fingerprint

Body Size
Direction compound

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

An analytical model for the design of RF resonators for MR body imaging. / Foo, T. K F; Hayes, C. E.; Kang, Y. W.

In: Magnetic Resonance in Medicine, Vol. 21, No. 2, 1991, p. 165-177.

Research output: Contribution to journalArticle

Foo, T. K F ; Hayes, C. E. ; Kang, Y. W. / An analytical model for the design of RF resonators for MR body imaging. In: Magnetic Resonance in Medicine. 1991 ; Vol. 21, No. 2. pp. 165-177.
@article{d39c15c41c56465b988b49b4b740576c,
title = "An analytical model for the design of RF resonators for MR body imaging",
abstract = "A closed form, analytical solution describing the RF fields generated by an RF body coil resonator for MR imaging at 1.5 and 4.0 T is presented. This solution extends the results of earlier studies of RF penetration in the body by explicitly including the RF coil, the RF shield, and the field variation along the z axis for high-pass birdcage coils. A salient feature of this treatment is the calculation of the axial propagation constant, k(z), which determines the z dependence of the RF field. We have determined the relative power deposition in the body, the B̄1 field homogeneity, and coil losses, which are functions of the coil-to-shield separation and body size. The relative power deposition in the body has been calculated to vary as the 1.58 power of the body radius. The calculations have also predicted that the field homogeneity in the z direction exhibits greater degradation at higher frequencies in a high-pass coil than in a low-pass coil. The model predicts an increase in coil losses by a factor of 2.8 as the coil-to-shield separation is reduced from 5 to 2 cm in a standard body resonator. Although the results for only a homogeneous cylindrical object or body are presented, the theory can be extended to a multilayered heterogeneous object of varying permittivity and conductivity.",
author = "Foo, {T. K F} and Hayes, {C. E.} and Kang, {Y. W.}",
year = "1991",
doi = "10.1002/mrm.1910210202",
language = "English (US)",
volume = "21",
pages = "165--177",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - An analytical model for the design of RF resonators for MR body imaging

AU - Foo, T. K F

AU - Hayes, C. E.

AU - Kang, Y. W.

PY - 1991

Y1 - 1991

N2 - A closed form, analytical solution describing the RF fields generated by an RF body coil resonator for MR imaging at 1.5 and 4.0 T is presented. This solution extends the results of earlier studies of RF penetration in the body by explicitly including the RF coil, the RF shield, and the field variation along the z axis for high-pass birdcage coils. A salient feature of this treatment is the calculation of the axial propagation constant, k(z), which determines the z dependence of the RF field. We have determined the relative power deposition in the body, the B̄1 field homogeneity, and coil losses, which are functions of the coil-to-shield separation and body size. The relative power deposition in the body has been calculated to vary as the 1.58 power of the body radius. The calculations have also predicted that the field homogeneity in the z direction exhibits greater degradation at higher frequencies in a high-pass coil than in a low-pass coil. The model predicts an increase in coil losses by a factor of 2.8 as the coil-to-shield separation is reduced from 5 to 2 cm in a standard body resonator. Although the results for only a homogeneous cylindrical object or body are presented, the theory can be extended to a multilayered heterogeneous object of varying permittivity and conductivity.

AB - A closed form, analytical solution describing the RF fields generated by an RF body coil resonator for MR imaging at 1.5 and 4.0 T is presented. This solution extends the results of earlier studies of RF penetration in the body by explicitly including the RF coil, the RF shield, and the field variation along the z axis for high-pass birdcage coils. A salient feature of this treatment is the calculation of the axial propagation constant, k(z), which determines the z dependence of the RF field. We have determined the relative power deposition in the body, the B̄1 field homogeneity, and coil losses, which are functions of the coil-to-shield separation and body size. The relative power deposition in the body has been calculated to vary as the 1.58 power of the body radius. The calculations have also predicted that the field homogeneity in the z direction exhibits greater degradation at higher frequencies in a high-pass coil than in a low-pass coil. The model predicts an increase in coil losses by a factor of 2.8 as the coil-to-shield separation is reduced from 5 to 2 cm in a standard body resonator. Although the results for only a homogeneous cylindrical object or body are presented, the theory can be extended to a multilayered heterogeneous object of varying permittivity and conductivity.

UR - http://www.scopus.com/inward/record.url?scp=0025934166&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025934166&partnerID=8YFLogxK

U2 - 10.1002/mrm.1910210202

DO - 10.1002/mrm.1910210202

M3 - Article

C2 - 1745117

AN - SCOPUS:0025934166

VL - 21

SP - 165

EP - 177

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 2

ER -