Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks' gestation: The AVIOx study

James I. Hagadorn, Anne M. Furey, Tuyet Hang Nghiem, Christopher H. Schmid, Dale L. Phelps, De Ann M. Pillers, Cynthia H. Cole, Pamela K. Donohue, Jennifer A. Shepard, Wally Carlo, Monica Collins, Jennifer Rylander, Stephen Bean, Francis Bednarek, Tara Loiseau, Gopal Gupta, Cassandra Horihan, Erica Burnell, David Wheeler, Sue EscoeLu Ann Papile, Conra Backstrom Lacey, Brian Darlow, David G. Sweet, Henry L. Halliday, Ajay J. Talati, Sheldon B. Korones, Carl Bose, Courtney Winston, Anna Allen, Kathleen Meyer, Karen Christianson, Bhavesh L. Shah, Anne Hansen, Alan Fujii, James Hagadorn, Cynthia Cole, Anne Furey, Skyler Greene, Ami Vora, Jennifer Cho, Prashant Shrestha, Ekua Abban, Patricia Hibberd, Christopher Schmid, Sarah Brody-Kaplan, Christine Botelho, Marcia Landa, Marina Weisburd, Caitlin Hurley

Research output: Contribution to journalArticlepeer-review

159 Scopus citations

Abstract

OBJECTIVE. The objective of this study was to document pulse oximeter saturation levels achieved in the first 4 weeks of life in infants who were born at <28 weeks' gestation, compared with the levels that were targeted by local policy, and examine factors that are associated with compliance with the target range. METHODS. Infants who were <28 weeks' gestation and ≤96 hours of age were enrolled in a prospective, multicenter cohort study. Oximetry data were collected with masked signal-extraction oximeters for a 72-hour period in each of the first 4 weeks of life. Data were compared with the pulse oximeter saturation target range prescribed by local institutional policy. Factors that were associated with intended range compliance were identified with hierarchical modeling. RESULTS. Fourteen centers from 3 countries enrolled 84 infants with mean ± SD birth weight of 863 ± 208 g and gestational age of 26 ± 1.4 weeks. Oxygen saturation policy limits ranged between 83% and 92% for lower limits and 92% and 98% for upper limits. For infants who received respiratory support, median pulse oximeter saturation level achieved was 95%. Center-specific medial levels were within the intended range at 12 centers. Centers maintained infants within their intended range 16% to 64% of the time but were above range 20% to 73% of the time. In hierarchical modeling, wider target ranges, higher target range upper limits, presence of a policy of setting oximeter alarms close to the target range limits, and lower gestational age were associated with improved target range compliance. CONCLUSIONS. Success with maintaining the intended pulse oximeter saturation range varied substantially among centers, among patients within centers, and for individual patients over time. Most noncompliance was above the intended range. Methods for improving compliance and the effect of improved compliance on neonatal outcomes require additional research.

Original languageEnglish (US)
Pages (from-to)1574-1582
Number of pages9
JournalPediatrics
Volume118
Issue number4
DOIs
StatePublished - Oct 2006

Keywords

  • Neonatal
  • Oximetry
  • Oxygen saturation
  • Policies
  • Prematurity

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health

Fingerprint

Dive into the research topics of 'Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks' gestation: The AVIOx study'. Together they form a unique fingerprint.

Cite this