A simulation study to assess SVD encoding for interventional MRI: Effect of object rotation and needle insertion

Jeffrey L. Duerk, Dee H. Wu, Y. C. Chung, Zhi Pei Liang, Jonathan S. Lewin

Research output: Contribution to journalArticle

Abstract

Singular value decomposition (SVD) encoding offers great promise to provide high spatial and temporal resolution required for interventional MRI (I-MRI) (1). This study investigates its efficacy when (a) objects are rotated and (b) a small device (ie, a needle) is moved within anatomic structures. It was found that SVD-encoded MRI is biased toward the reference from which encoding vectors are derived, thus providing a potential limitation under conditions in which the object has undergone significant global change. Reference images with partial device insertion may be needed to accurately resolve the device or track the object motion. Theoretically, the differences between the reference and the object being imaged suggest that SVD encoding is suboptimal (in a minimum mean squared error sense). Other encoding/reconstruction algorithms may come closer to achieving the desired advantages in spatial and temporal fidelity.

Original languageEnglish (US)
Pages (from-to)957-960
Number of pages4
JournalJournal of Magnetic Resonance Imaging
Volume6
Issue number6
StatePublished - Nov 1996
Externally publishedYes

Fingerprint

Interventional Magnetic Resonance Imaging
Needles
Equipment and Supplies

Keywords

  • Artifacts
  • Fast imaging
  • Interventional MRI
  • SVD

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

A simulation study to assess SVD encoding for interventional MRI : Effect of object rotation and needle insertion. / Duerk, Jeffrey L.; Wu, Dee H.; Chung, Y. C.; Liang, Zhi Pei; Lewin, Jonathan S.

In: Journal of Magnetic Resonance Imaging, Vol. 6, No. 6, 11.1996, p. 957-960.

Research output: Contribution to journalArticle

Duerk, Jeffrey L. ; Wu, Dee H. ; Chung, Y. C. ; Liang, Zhi Pei ; Lewin, Jonathan S. / A simulation study to assess SVD encoding for interventional MRI : Effect of object rotation and needle insertion. In: Journal of Magnetic Resonance Imaging. 1996 ; Vol. 6, No. 6. pp. 957-960.
@article{1b6484aca58c4183a9b5ae7b1ce08f29,
title = "A simulation study to assess SVD encoding for interventional MRI: Effect of object rotation and needle insertion",
abstract = "Singular value decomposition (SVD) encoding offers great promise to provide high spatial and temporal resolution required for interventional MRI (I-MRI) (1). This study investigates its efficacy when (a) objects are rotated and (b) a small device (ie, a needle) is moved within anatomic structures. It was found that SVD-encoded MRI is biased toward the reference from which encoding vectors are derived, thus providing a potential limitation under conditions in which the object has undergone significant global change. Reference images with partial device insertion may be needed to accurately resolve the device or track the object motion. Theoretically, the differences between the reference and the object being imaged suggest that SVD encoding is suboptimal (in a minimum mean squared error sense). Other encoding/reconstruction algorithms may come closer to achieving the desired advantages in spatial and temporal fidelity.",
keywords = "Artifacts, Fast imaging, Interventional MRI, SVD",
author = "Duerk, {Jeffrey L.} and Wu, {Dee H.} and Chung, {Y. C.} and Liang, {Zhi Pei} and Lewin, {Jonathan S.}",
year = "1996",
month = "11",
language = "English (US)",
volume = "6",
pages = "957--960",
journal = "Journal of Magnetic Resonance Imaging",
issn = "1053-1807",
publisher = "John Wiley and Sons Inc.",
number = "6",

}

TY - JOUR

T1 - A simulation study to assess SVD encoding for interventional MRI

T2 - Effect of object rotation and needle insertion

AU - Duerk, Jeffrey L.

AU - Wu, Dee H.

AU - Chung, Y. C.

AU - Liang, Zhi Pei

AU - Lewin, Jonathan S.

PY - 1996/11

Y1 - 1996/11

N2 - Singular value decomposition (SVD) encoding offers great promise to provide high spatial and temporal resolution required for interventional MRI (I-MRI) (1). This study investigates its efficacy when (a) objects are rotated and (b) a small device (ie, a needle) is moved within anatomic structures. It was found that SVD-encoded MRI is biased toward the reference from which encoding vectors are derived, thus providing a potential limitation under conditions in which the object has undergone significant global change. Reference images with partial device insertion may be needed to accurately resolve the device or track the object motion. Theoretically, the differences between the reference and the object being imaged suggest that SVD encoding is suboptimal (in a minimum mean squared error sense). Other encoding/reconstruction algorithms may come closer to achieving the desired advantages in spatial and temporal fidelity.

AB - Singular value decomposition (SVD) encoding offers great promise to provide high spatial and temporal resolution required for interventional MRI (I-MRI) (1). This study investigates its efficacy when (a) objects are rotated and (b) a small device (ie, a needle) is moved within anatomic structures. It was found that SVD-encoded MRI is biased toward the reference from which encoding vectors are derived, thus providing a potential limitation under conditions in which the object has undergone significant global change. Reference images with partial device insertion may be needed to accurately resolve the device or track the object motion. Theoretically, the differences between the reference and the object being imaged suggest that SVD encoding is suboptimal (in a minimum mean squared error sense). Other encoding/reconstruction algorithms may come closer to achieving the desired advantages in spatial and temporal fidelity.

KW - Artifacts

KW - Fast imaging

KW - Interventional MRI

KW - SVD

UR - http://www.scopus.com/inward/record.url?scp=0030281093&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030281093&partnerID=8YFLogxK

M3 - Article

C2 - 8956146

AN - SCOPUS:0030281093

VL - 6

SP - 957

EP - 960

JO - Journal of Magnetic Resonance Imaging

JF - Journal of Magnetic Resonance Imaging

SN - 1053-1807

IS - 6

ER -