A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range

Ran Zeimer, D. K. Gieser, J. T. Wilensky, J. M. Noth, M. M. Mori, E. E. Odunukwe

Research output: Contribution to journalArticle

Abstract

Glaucomatous damage develops in certain patients because of elevated episcleral venous pressure (EVP). To measure the EVP in typical clinical settings, a practical and reliable instrument is needed. We have developed such an instrument, the venomanometer, and present tests of its reproducibility. The intraobserver reproducibility was 0.7 mm Hg, and the mean difference between the readings of two observers was 0.7 ± 1.2 mm Hg. We then used the venomanometer to study the EVP in 122 eyes of 68 normal subjects distributed in seven age groups between 10 and 80 years. The EVP did not vary with age (mean, 7.6 ± 1.3 mm Hg). This value is compared with those obtained with other instruments.

Original languageEnglish (US)
Pages (from-to)1447-1449
Number of pages3
JournalArchives of Ophthalmology
Volume101
Issue number9
StatePublished - 1983
Externally publishedYes

Fingerprint

Venous Pressure
Reference Values
Reading
Age Groups

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Zeimer, R., Gieser, D. K., Wilensky, J. T., Noth, J. M., Mori, M. M., & Odunukwe, E. E. (1983). A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range. Archives of Ophthalmology, 101(9), 1447-1449.

A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range. / Zeimer, Ran; Gieser, D. K.; Wilensky, J. T.; Noth, J. M.; Mori, M. M.; Odunukwe, E. E.

In: Archives of Ophthalmology, Vol. 101, No. 9, 1983, p. 1447-1449.

Research output: Contribution to journalArticle

Zeimer, R, Gieser, DK, Wilensky, JT, Noth, JM, Mori, MM & Odunukwe, EE 1983, 'A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range', Archives of Ophthalmology, vol. 101, no. 9, pp. 1447-1449.
Zeimer, Ran ; Gieser, D. K. ; Wilensky, J. T. ; Noth, J. M. ; Mori, M. M. ; Odunukwe, E. E. / A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range. In: Archives of Ophthalmology. 1983 ; Vol. 101, No. 9. pp. 1447-1449.
@article{cec27c2adbed4971a058da61cc46db55,
title = "A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range",
abstract = "Glaucomatous damage develops in certain patients because of elevated episcleral venous pressure (EVP). To measure the EVP in typical clinical settings, a practical and reliable instrument is needed. We have developed such an instrument, the venomanometer, and present tests of its reproducibility. The intraobserver reproducibility was 0.7 mm Hg, and the mean difference between the readings of two observers was 0.7 ± 1.2 mm Hg. We then used the venomanometer to study the EVP in 122 eyes of 68 normal subjects distributed in seven age groups between 10 and 80 years. The EVP did not vary with age (mean, 7.6 ± 1.3 mm Hg). This value is compared with those obtained with other instruments.",
author = "Ran Zeimer and Gieser, {D. K.} and Wilensky, {J. T.} and Noth, {J. M.} and Mori, {M. M.} and Odunukwe, {E. E.}",
year = "1983",
language = "English (US)",
volume = "101",
pages = "1447--1449",
journal = "JAMA Ophthalmology",
issn = "2168-6165",
publisher = "American Medical Association",
number = "9",

}

TY - JOUR

T1 - A practical venomanometer. Measurement of episcleral venous pressure and assessment of the normal range

AU - Zeimer, Ran

AU - Gieser, D. K.

AU - Wilensky, J. T.

AU - Noth, J. M.

AU - Mori, M. M.

AU - Odunukwe, E. E.

PY - 1983

Y1 - 1983

N2 - Glaucomatous damage develops in certain patients because of elevated episcleral venous pressure (EVP). To measure the EVP in typical clinical settings, a practical and reliable instrument is needed. We have developed such an instrument, the venomanometer, and present tests of its reproducibility. The intraobserver reproducibility was 0.7 mm Hg, and the mean difference between the readings of two observers was 0.7 ± 1.2 mm Hg. We then used the venomanometer to study the EVP in 122 eyes of 68 normal subjects distributed in seven age groups between 10 and 80 years. The EVP did not vary with age (mean, 7.6 ± 1.3 mm Hg). This value is compared with those obtained with other instruments.

AB - Glaucomatous damage develops in certain patients because of elevated episcleral venous pressure (EVP). To measure the EVP in typical clinical settings, a practical and reliable instrument is needed. We have developed such an instrument, the venomanometer, and present tests of its reproducibility. The intraobserver reproducibility was 0.7 mm Hg, and the mean difference between the readings of two observers was 0.7 ± 1.2 mm Hg. We then used the venomanometer to study the EVP in 122 eyes of 68 normal subjects distributed in seven age groups between 10 and 80 years. The EVP did not vary with age (mean, 7.6 ± 1.3 mm Hg). This value is compared with those obtained with other instruments.

UR - http://www.scopus.com/inward/record.url?scp=0020551059&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020551059&partnerID=8YFLogxK

M3 - Article

C2 - 6615314

AN - SCOPUS:0020551059

VL - 101

SP - 1447

EP - 1449

JO - JAMA Ophthalmology

JF - JAMA Ophthalmology

SN - 2168-6165

IS - 9

ER -